Je něco špatně v tomto záznamu ?
Focusing of alkali earth metals in ligand step gradient
E. Sišperová, E. Glovinová, J. Budilová, J. Pospíchal,
Jazyk angličtina Země Nizozemsko
Typ dokumentu časopisecké články, práce podpořená grantem
- MeSH
- chelátory chemie MeSH
- elektrolyty chemie MeSH
- izotachoforéza metody MeSH
- koncentrace vodíkových iontů MeSH
- kovy alkalických zemin analýza chemie MeSH
- senzitivita a specificita MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
A capillary electroseparation technique for focusing and selective pre-concentration of metal chelates with subsequent on-line isotachophoresis (ITP) analysis was developed and verified. The ions of alkali earth metals (Mg, Ca, Sr, and Ba) were pre-concentrated from the mixture and analyzed. The focusing of the metals was carried out in a ligand step gradient, which was created by the addition of a convenient ligand agent to the regular stationary pH step gradient. The analytical procedure consisted of three steps. During the first step, the metal ions were electrokinetically continuously dosed into the column where they were selectively trapped on the stationary ligand step gradient in the form of unmoving zones of chelate complexes with effectively zero charge. After a detectable amount of analyte was accumulated, the dosing was stopped. The accumulated zones were mobilized to the analytical column, where they were analyzed by the ITP method with conductivity or photometric detection. The proper electrolyte systems for dosing, mobilizing, and analyzing in isoelectric focusing (IEF), moving boundary electrophoresis (MBE), and ITP modes were consequently developed and put into practice. The trapping selectivity can be regulated by the choice of pH and convenient complexing agents. A mixture of alkali earth metals were used as model analytes. Using a 3000 s dosing time, the proposed method improved the detection limit by 5-29 times in comparison to analysis by ITP with classical injection.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12027637
- 003
- CZ-PrNML
- 005
- 20121207125337.0
- 007
- ta
- 008
- 120817e20110321ne f 000 0#eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.chroma.2011.03.023 $2 doi
- 035 __
- $a (PubMed)21477804
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Sišperová, Eliška $u Department of Chemistry and Biochemistry, Mendel University in Brno, Brno, Czech Republic.
- 245 10
- $a Focusing of alkali earth metals in ligand step gradient / $c E. Sišperová, E. Glovinová, J. Budilová, J. Pospíchal,
- 520 9_
- $a A capillary electroseparation technique for focusing and selective pre-concentration of metal chelates with subsequent on-line isotachophoresis (ITP) analysis was developed and verified. The ions of alkali earth metals (Mg, Ca, Sr, and Ba) were pre-concentrated from the mixture and analyzed. The focusing of the metals was carried out in a ligand step gradient, which was created by the addition of a convenient ligand agent to the regular stationary pH step gradient. The analytical procedure consisted of three steps. During the first step, the metal ions were electrokinetically continuously dosed into the column where they were selectively trapped on the stationary ligand step gradient in the form of unmoving zones of chelate complexes with effectively zero charge. After a detectable amount of analyte was accumulated, the dosing was stopped. The accumulated zones were mobilized to the analytical column, where they were analyzed by the ITP method with conductivity or photometric detection. The proper electrolyte systems for dosing, mobilizing, and analyzing in isoelectric focusing (IEF), moving boundary electrophoresis (MBE), and ITP modes were consequently developed and put into practice. The trapping selectivity can be regulated by the choice of pH and convenient complexing agents. A mixture of alkali earth metals were used as model analytes. Using a 3000 s dosing time, the proposed method improved the detection limit by 5-29 times in comparison to analysis by ITP with classical injection.
- 650 _2
- $a chelátory $x chemie $7 D002614
- 650 _2
- $a elektrolyty $x chemie $7 D004573
- 650 _2
- $a koncentrace vodíkových iontů $7 D006863
- 650 _2
- $a izotachoforéza $x metody $7 D057625
- 650 _2
- $a kovy alkalických zemin $x analýza $x chemie $7 D008673
- 650 _2
- $a senzitivita a specificita $7 D012680
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Glovinová, Eliška
- 700 1_
- $a Budilová, Jana
- 700 1_
- $a Pospíchal, Jan
- 773 0_
- $w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1218, č. 20 (20110321), s. 3105-10
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/21477804 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y m
- 990 __
- $a 20120817 $b ABA008
- 991 __
- $a 20121207125411 $b ABA008
- 999 __
- $a ok $b bmc $g 949679 $s 784983
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2011 $b 1218 $c 20 $d 3105-10 $e 20110321 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
- LZP __
- $a Pubmed-20120817/11/03