Electrical Impedance Tomography Can Be Used to Quantify Lung Hyperinflation during HFOV: The Pilot Study in Pigs
V. Ort, K. Roubik
Language English Country Switzerland
Document type Journal Article
Grant support
SGS19/203/OHK4/3T/17
Grant Agency of the Czech Technical University in Prague
SGS20/202/OHK4/3T/17
Grant Agency of the Czech Technical University in Prague
NLK
Directory of Open Access Journals
from 2011
PubMed Central
from 2011
Europe PubMed Central
from 2011
ProQuest Central
from 2011-01-01
Open Access Digital Library
from 2011-01-01
Open Access Digital Library
from 2011-01-01
ROAD: Directory of Open Access Scholarly Resources
from 2011
- Publication type
- Journal Article MeSH
Dynamic hyperinflation is reported as a potential risk during high-frequency oscillatory ventilation (HFOV), and its existence has been documented both by physical models and by CT. The aim of this study is to determine the suitability of electrical impendence tomography (EIT) for the measurement of dynamic lung hyperinflation and hypoinflation during HFOV. Eleven healthy pigs were anaesthetized and ventilated using HFOV. The difference between the airway pressure at the airway opening and alveolar space was measured by EIT and esophageal balloons at three mean airway pressures (12, 18 and 24 cm H2O) and two inspiratory to expiratory time ratios (1:1, 1:2). The I:E ratio was the primary parameter associated with differences between airway and alveolar pressures. All animals showed hyperinflation at a 1:1 ratio (median 1.9 cm H2O) and hypoinflation at a 1:2 (median -4.0 cm H2O) as measured by EIT. EIT measurements had a linear correlation to esophageal balloon measurements (r2 = -0.915, p = 0.0085). EIT measurements were slightly higher than that of the esophageal balloon transducer with the mean difference of 0.57 cm H2O. Presence of a hyperinflation or hypoinflation was also confirmed independently by chest X-ray. We found that dynamic hyperinflation developed during HFOV may be detected and characterized noninvasively by EIT.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22023425
- 003
- CZ-PrNML
- 005
- 20221031095226.0
- 007
- ta
- 008
- 221010s2022 sz f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.3390/diagnostics12092081 $2 doi
- 035 __
- $a (PubMed)36140483
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a sz
- 100 1_
- $a Ort, Vaclav $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, nam. Sitna 3105, 272 01 Kladno, Czech Republic $1 0000000316772446
- 245 10
- $a Electrical Impedance Tomography Can Be Used to Quantify Lung Hyperinflation during HFOV: The Pilot Study in Pigs / $c V. Ort, K. Roubik
- 520 9_
- $a Dynamic hyperinflation is reported as a potential risk during high-frequency oscillatory ventilation (HFOV), and its existence has been documented both by physical models and by CT. The aim of this study is to determine the suitability of electrical impendence tomography (EIT) for the measurement of dynamic lung hyperinflation and hypoinflation during HFOV. Eleven healthy pigs were anaesthetized and ventilated using HFOV. The difference between the airway pressure at the airway opening and alveolar space was measured by EIT and esophageal balloons at three mean airway pressures (12, 18 and 24 cm H2O) and two inspiratory to expiratory time ratios (1:1, 1:2). The I:E ratio was the primary parameter associated with differences between airway and alveolar pressures. All animals showed hyperinflation at a 1:1 ratio (median 1.9 cm H2O) and hypoinflation at a 1:2 (median -4.0 cm H2O) as measured by EIT. EIT measurements had a linear correlation to esophageal balloon measurements (r2 = -0.915, p = 0.0085). EIT measurements were slightly higher than that of the esophageal balloon transducer with the mean difference of 0.57 cm H2O. Presence of a hyperinflation or hypoinflation was also confirmed independently by chest X-ray. We found that dynamic hyperinflation developed during HFOV may be detected and characterized noninvasively by EIT.
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Roubik, Karel $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, nam. Sitna 3105, 272 01 Kladno, Czech Republic $1 0000000158136355
- 773 0_
- $w MED00195450 $t Diagnostics (Basel, Switzerland) $x 2075-4418 $g Roč. 12, č. 9 (2022)
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/36140483 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20221010 $b ABA008
- 991 __
- $a 20221031095224 $b ABA008
- 999 __
- $a ind $b bmc $g 1853863 $s 1174713
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2022 $b 12 $c 9 $e 20220828 $i 2075-4418 $m Diagnostics $n Diagnostics $x MED00195450
- GRA __
- $a SGS19/203/OHK4/3T/17 $p Grant Agency of the Czech Technical University in Prague
- GRA __
- $a SGS20/202/OHK4/3T/17 $p Grant Agency of the Czech Technical University in Prague
- LZP __
- $a Pubmed-20221010