Je něco špatně v tomto záznamu ?
Upgrade of the Langendorff apparatus using the infrared thermo-control system and an intelligent heater
Josef Skopalík, Jiří Sekora, Martin Pešl, Markéta Bébarová, Olga Švecová, Tomáš Parák, Vratislav Čmiel, Ivo Provaznik, Edita Jeklová, Josef Mašek
Jazyk angličtina Země Česko
Typ dokumentu práce podpořená grantem
Odkazy
DOI
10.14311/ctj.2020.4.03
Knihovny.cz E-zdroje
- MeSH
- infračervené záření MeSH
- lidé MeSH
- preparace izolovaného srdce * metody MeSH
- termoregulace MeSH
- zahřívání metody přístrojové vybavení MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- práce podpořená grantem MeSH
Biological experiments involving isolated organs and tissues demand precise temperature monitoring and regulation. An automatic temperature control system was proposed and optimised on real isolated swine hearts and the prototype is described in this work. The traditional Langendorff apparatus consists of a heart holder, a reservoir of perfusion solution flowing to aortic cannula and a heating bath allowing passive heat transfer to the reservoir of perfusion solution. The commercial infrared camera FLIR T62101 was added to this basic set-up and used for very precise monitoring of the temperature kinetic of the organ and connected with an electronic feedback loop, which allowed real-time and precise regulation of heat transfer from the heating bath to the perfusion solution and in turn indirectly to the heart tissue. This provides real time control and active regulation of the myocardial tissue temperature. The infrared camera was tested in several modes and several variants of detection were optimised for ideal measurement of the region of interest of the ex vivo organ. The kinetics of the temperature changes and temperature stability of the tissue were recorded and calibrated by external electronic thermometers (type Pt100, inserted in tissue). The time lapse from the hang-up of the hypo termed organ (30 °C) until optimal warming (37 °C) was less than eight minutes in the final instrument prototype. The final stability of the 37 °C tissue temperature was approved; the temperature fluctuation of left ventricle tissue was characterised as 36.8 ± 0.5 °C. This upgraded traditional instrument could be used in specific preclinical and clinical transplantation and analytical projects in future.
Department of Biology Faculty of Medicine Masaryk University Brno Czech Republic
Department of Pharmacology and Toxicology Faculty of Pharmacy Masaryk University Brno Czech Republic
Department of Physiology Faculty of Medicine Masaryk University Brno Czech Republic
International Clinical Research Center St Anne's University Hospital Brno Czech Republic
Literatura
- 000
- 00000naa a2200000 a 4500
- 001
- bmc23006601
- 003
- CZ-PrNML
- 005
- 20231006164659.0
- 007
- ta
- 008
- 230517s2020 xr a f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.14311/CTJ.2020.4.03 $2 doi
- 040 __
- $a ABA008 $d ABA008 $e AACR2 $b cze
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Skopalík, Josef $7 xx0108995 $u Department of Biomedical Engineering, Faculty of Electrical Engineering and Communication, Brno University of Technology, Brno, Czech Republic $u Department of Pharmacology and Toxicology, Faculty of Pharmacy, Masaryk University, Brno, Czech Republic
- 245 10
- $a Upgrade of the Langendorff apparatus using the infrared thermo-control system and an intelligent heater / $c Josef Skopalík, Jiří Sekora, Martin Pešl, Markéta Bébarová, Olga Švecová, Tomáš Parák, Vratislav Čmiel, Ivo Provaznik, Edita Jeklová, Josef Mašek
- 504 __
- $a Literatura
- 520 9_
- $a Biological experiments involving isolated organs and tissues demand precise temperature monitoring and regulation. An automatic temperature control system was proposed and optimised on real isolated swine hearts and the prototype is described in this work. The traditional Langendorff apparatus consists of a heart holder, a reservoir of perfusion solution flowing to aortic cannula and a heating bath allowing passive heat transfer to the reservoir of perfusion solution. The commercial infrared camera FLIR T62101 was added to this basic set-up and used for very precise monitoring of the temperature kinetic of the organ and connected with an electronic feedback loop, which allowed real-time and precise regulation of heat transfer from the heating bath to the perfusion solution and in turn indirectly to the heart tissue. This provides real time control and active regulation of the myocardial tissue temperature. The infrared camera was tested in several modes and several variants of detection were optimised for ideal measurement of the region of interest of the ex vivo organ. The kinetics of the temperature changes and temperature stability of the tissue were recorded and calibrated by external electronic thermometers (type Pt100, inserted in tissue). The time lapse from the hang-up of the hypo termed organ (30 °C) until optimal warming (37 °C) was less than eight minutes in the final instrument prototype. The final stability of the 37 °C tissue temperature was approved; the temperature fluctuation of left ventricle tissue was characterised as 36.8 ± 0.5 °C. This upgraded traditional instrument could be used in specific preclinical and clinical transplantation and analytical projects in future.
- 650 17
- $a preparace izolovaného srdce $x metody $7 D000067268 $2 czmesh
- 650 _7
- $a termoregulace $7 D001833 $2 czmesh
- 650 _7
- $a zahřívání $x metody $x přístrojové vybavení $7 D017694 $2 czmesh
- 650 _7
- $a infračervené záření $7 D007259 $2 czmesh
- 650 _7
- $a lidé $7 D006801 $2 czmesh
- 655 _7
- $a práce podpořená grantem $7 D013485 $2 czmesh
- 700 1_
- $a Sekora, Jiří $7 _AN082478 $u Department of Biomedical Engineering, Faculty of Electrical Engineering and Communication, Brno University of Technology, Brno, Czech Republic
- 700 1_
- $a Pešl, Martin $7 xx0143722 $u Department of Biology, Faculty of Medicine, Masaryk University, Brno, Czech Republic $u International Clinical Research Center, St. Anne's University Hospital, Brno, Czech Republic $u First Department of Internal Medicine - Cardioangiology Faculty of Medicine, Masaryk University, and St. Anne's University Hospital, Brno, Czech Republic
- 700 1_
- $a Bébarová, Markéta, $d 1976- $7 xx0037780 $u Department of Physiology, Faculty of Medicine, Masaryk University, Brno, Czech Republic
- 700 1_
- $a Švecová, Olga $7 xx0301708 $u Department of Physiology, Faculty of Medicine, Masaryk University, Brno, Czech Republic
- 700 1_
- $a Parák, Tomáš $7 jx20090302030 $u Department of Pharmacology and Toxicology, Faculty of Pharmacy, Masaryk University, Brno, Czech Republic
- 700 1_
- $a Čmiel, Vratislav $7 xx0260455 $u Department of Biomedical Engineering, Faculty of Electrical Engineering and Communication, Brno University of Technology, Brno, Czech Republic
- 700 1_
- $a Provazník, Ivo, $d 1968- $7 mzk2002106297 $u Department of Biomedical Engineering, Faculty of Electrical Engineering and Communication, Brno University of Technology, Brno, Czech Republic
- 700 1_
- $a Jeklová, Edita $7 xx0106583 $u Veterinary Research Institute, Brno, Czech Republic
- 700 1_
- $a Mašek, Josef $7 xx0171644 $u Veterinary Research Institute, Brno, Czech Republic
- 773 0_
- $t Lékař a technika $x 0301-5491 $g Roč. 50, č. 4 (2020), s. 137-141 $w MED00011033
- 910 __
- $a ABA008 $b B 1367 $c 1071 b $y p $z 0
- 990 __
- $a 20230517141648 $b ABA008
- 991 __
- $a 20231006164653 $b ABA008
- 999 __
- $a ok $b bmc $g 1933660 $s 1192823
- BAS __
- $a 3
- BMC __
- $a 2020 $b 50 $c 4 $d 137-141 $i 0301-5491 $m Lékař a technika $n Lék. tech. $x MED00011033
- LZP __
- $c NLK197 $d 20231006 $a NLK 2023-16/kv