Surgical aspects of the implantation of a left ventricle assist device
Language English Country Czech Republic Media print
Document type Journal Article
PubMed
3133186
Knihovny.cz E-resources
- MeSH
- Assisted Circulation * MeSH
- Heart-Assist Devices * MeSH
- Prostheses and Implants * MeSH
- Cattle MeSH
- Animals MeSH
- Check Tag
- Cattle MeSH
- Animals MeSH
- Publication type
- Journal Article MeSH
The authors have summarised their experience with short-term use of a left ventricle assist device (LVAD). Nine experiments were carried out in healthy calves. A polyurethane pump of their own design, placed extracorporeally, was used. Its in-flow cannula was placed in the left atrium, the outflow one in the initial segment of the descending aorta. The LVAD was operated for a period of 5 days. Fourteen days after its disconnection, the animal was sacrificed and both the bypass system and parenchymal organs were examined. No surgical complications were observed during the experiments. Macro- and microscopical pictures showed minor infarctions in the renal tissue which, however, did not impair renal function. The findings in other organs were related to anaesthesia, the operative procedure and way of animal sacrificing. The experiments have proved the safety of the bypass system used over a short period and, consequently, its potential for future clinical application.