Activity standardisation of 45Ca and 204Tl using the new TDCR system at CMI

. 2012 Sep ; 70 (9) : 2200-3. [epub] 20120308

Jazyk angličtina Země Anglie, Velká Británie Médium print-electronic

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid22476016
Odkazy

PubMed 22476016
DOI 10.1016/j.apradiso.2012.02.098
PII: S0969-8043(12)00146-7
Knihovny.cz E-zdroje

A new triple-to-double coincidence ratio (TDCR) system has been established at CMI as an additional technique for primary activity standardisation. Details of the detectors and the electronics are given. Construction of the optical chamber was validated using measurement of a (3)H standard-the efficiency achieved with Ultima Gold was 58%. Several methods of efficiency variation (coloured bands painted on vial, positioning and photomultipliers defocussing) are compared. Activity concentrations of (204)Tl and (45)Ca were determined with the TDCR method and with the efficiency tracer technique. Advantages of the TDCR method compared to the efficiency tracing method and agreement between results are shown.