Acute effects of right ventricular pacing on cardiac haemodynamics and transvalvular impedance
Jazyk angličtina Země Česko Médium print-electronic
Typ dokumentu časopisecké články
PubMed
23446212
DOI
10.5507/bp.2013.009
Knihovny.cz E-zdroje
- MeSH
- echokardiografie MeSH
- elektrická impedance škodlivé účinky terapeutické užití MeSH
- elektrokardiografie MeSH
- hemodynamika fyziologie MeSH
- kardiostimulace umělá škodlivé účinky metody MeSH
- lidé MeSH
- srdce - funkce komor fyziologie MeSH
- srdeční komory patofyziologie MeSH
- tepový objem MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
AIMS: To assess the acute side-effects of right ventricular (RV) stimulation applied in apex and mid-septum, in order to establish the optimal lead location in clinical practice. METHODS: During pacemaker implantation, the ventricular lead was temporarily fixed in the apex and then moved to mid-septum. In both positions, surface and endocardial electrograms and transvalvular impedance (32 cases), left ventricular (LV) pressure (23), and transthoracic echocardiography (10) were acquired with intrinsic activity and VDD pacing. RESULTS: A larger increase in QRS duration was noticed with apical than septal pacing (65±25 vs. 45±29 ms; P<10(-4)). The proportion of cases where RV stimulation affected the transvalvular impedance waveform was higher with apical lead location (56% vs. 20%; P<0.02). VDD pacing at either site reduced the maximum dP/dt by 6% with respect to intrinsic AV conduction (IAVC; P<0.005). The maximum pressure drop taking place in 100 ms was reduced by 6 and 8%, respectively, with apical and septal pacing (P<0.01 vs. IAVC). Apical VDD decreased mitral annulus velocity in early diastole (E') from 7.5±1.4 to 5.9±0.9 cm/s (P<0.02) and prolonged the E-wave deceleration time (DT) from 156±33 to 199±54 ms (P<0.02), while septal pacing induced non-significant modifications in E' and DT. CONCLUSION: Ventricular stimulation acutely impairs LV systolic and diastolic performance, independent of the pacing site. Septal lead location preserves RV contraction mechanics and reduces the electrical interventricular delay.
Citace poskytuje Crossref.org