• This record comes from PubMed

Chiral separation of benzothiazole derivatives of amino acids using capillary zone electrophoresis

. 2017 Feb ; 40 (3) : 798-803. [epub] 20161223

Language English Country Germany Media print-electronic

Document type Journal Article

A method for the separation of enantiomers of leucine and phenylalanine benzothiazole derivatives as potential antimicrobial agents was developed using capillary zone electrophoresis with a dual cyclodextrin (CD) system. The best resolution of enantiomers was achieved in 100 mmol/L phosphate background electrolyte (pH 3.5) with the dual CD system consisting of 10 mmol/L of β-CD with 10 mmol/L of 2-hydroxypropyl-β-cyclodextrin for leucine derivative and 10 mmol/L of 2-hydroxypropyl-γ-cyclodextrin for phenylalanine derivative, respectively. Under the optimal conditions, the highest enantioresolution of 1.25 was achieved in a noncoated-fused silica capillary at 17°C and 24 kV applied voltage.

References provided by Crossref.org