True lab-in-a-syringe technology for bioassays
Language English Country Netherlands Media print-electronic
Document type Journal Article
PubMed
28738580
DOI
10.1016/j.talanta.2017.06.017
PII: S0039-9140(17)30649-5
Knihovny.cz E-resources
- Keywords
- Lab-in-a-syringe, Lateral-flow assay, Low-cost microfluidics, Sensing,
- MeSH
- Biological Assay instrumentation MeSH
- Equipment Design MeSH
- Syringes * MeSH
- Publication type
- Journal Article MeSH
In our work, we introduced a novel concept of the lab-in-a-syringe tests. We solved the problem of detection in already published LIS tests by putting all the reaction and detection pads directly into the syringe barrel. We also used more layers to make the results visible for users. Two detection layouts: (i) with using rounded pads-based detection, and (ii) with using rectangular detection pads, were studied. As the proof of concept, we studied the determination of Ni(II) using dimethylglyoxime as the reagent and blocking of the interference of Fe(II). The calibrations for Ni(II) at the optimal conditions has excellent R2 of 0.998 with production costs of 0.2 USD per one test.
References provided by Crossref.org