COMPARISON OF DIFFERENT TECHNIQUES FOR EVALUATION OF DOSE DISTRIBUTIONS IN RADIOTHERAPY USING RADIOCHROMIC EBT3 FILMS
Language English Country England, Great Britain Media print
Document type Comparative Study, Evaluation Study, Journal Article
PubMed
31711200
DOI
10.1093/rpd/ncz231
PII: 5618731
Knihovny.cz E-resources
- MeSH
- Algorithms MeSH
- Radiotherapy Dosage MeSH
- Phantoms, Imaging * MeSH
- Film Dosimetry instrumentation methods MeSH
- Calibration MeSH
- Humans MeSH
- Prostatic Neoplasms radiotherapy MeSH
- Radiotherapy Planning, Computer-Assisted methods MeSH
- Radiotherapy, Intensity-Modulated methods MeSH
- Gamma Rays MeSH
- Check Tag
- Humans MeSH
- Male MeSH
- Publication type
- Journal Article MeSH
- Evaluation Study MeSH
- Comparative Study MeSH
In radiotherapy, radiochromic films can be used for verification of delivery of dose distributions calculated by treatment planning systems. The main objective of this work was to compare three different techniques for evaluation of dose distributions for prostate cancer treatment plans using radiochromic EBT3 films. These techniques are: red channel evaluation taking into account only a response of irradiated film (R), red channel evaluation taking into account a response of unirradiated and irradiated film (Rcor) and multichannel evaluation in FilmQA software (RGB). Also comparison between film and MatriXX measurement was performed. Comparison showed that gamma analysis passing rates strongly depend on evaluation technique and on a model of scanner for digitizing films. The highest gamma passing rates were obtained with red channel evaluation taking into account a response of unirradiated and irradiated film using Epson V750 scanner (Rcor) and multichannel evaluation in FilmQA using Epson 11000XL scanner.
Department of Radiological Physics Na Bulovce Hospital Prague Czech Republic
Medical Exposure Section National Radiation Protection Institute Prague Czech Republic
References provided by Crossref.org