PSEUDO-3D IMRT VERIFICATION WITH EBT3 RADIOCHROMIC FILM
Language English Country Great Britain, England Media print
Document type Journal Article
PubMed
31943097
DOI
10.1093/rpd/ncz232
PII: 5706874
Knihovny.cz E-resources
- MeSH
- Radiotherapy Dosage MeSH
- Phantoms, Imaging * MeSH
- Film Dosimetry instrumentation methods MeSH
- Humans MeSH
- Prostatic Neoplasms radiotherapy MeSH
- Radiotherapy Planning, Computer-Assisted instrumentation methods MeSH
- Radiotherapy, Intensity-Modulated instrumentation methods MeSH
- Gamma Rays MeSH
- Check Tag
- Humans MeSH
- Male MeSH
- Publication type
- Journal Article MeSH
This work proposes a new method for pseudo-3D verification of intensity-modulated radiation therapy (IMRT) dose distributions. Unlike commercial solutions, it uses measured doses only for gamma evaluation. Its resolution is far better than with electronic detectors within the measured plane and comparable in other directions. It is readily available at clinics because it uses existing resources-a slab phantom and EBT3 films. The method was tested on six IMRT clinical cases. An in-house code for 2D and pseudo-3D gamma analysis was written in MATLAB and compared to OmniPro I'mRT.