Can ultrafast single-particle analysis using ICP-MS affect the detection limit? Case study: Silver nanoparticles

. 2020 Apr 01 ; 210 () : 120665. [epub] 20191224

Status PubMed-not-MEDLINE Jazyk angličtina Země Nizozemsko Médium print-electronic

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid31987196
Odkazy

PubMed 31987196
DOI 10.1016/j.talanta.2019.120665
PII: S0039-9140(19)31298-6
Knihovny.cz E-zdroje

Ultrafast measurement using dwell times below 100 μs down to 10 μs is a relatively new feature of single particle analysis using ICP-MS. In this study, we tested the effect of shorter dwell times on the particle size detection limit (Dd.l.). Decreasing dwell times below 100 μs did not lead to a statistically significant decrease in Dd.l. The particle size detection limit (quadrupole ICP-MS) of silver nanoparticles (NP) was estimated to be approx. 10-11 nm. Ag NPs close to Dd.l. were analysed. The 14-nm NPs showed low detection yield; only 5% of number of NPs estimated from transport efficiency was detected. The 20-nm NPs showed 44% detection yield; only in the case of 30-nm NPs did the number of detected NPs correspond to transport efficiency. It is obvious that near Dd.l. estimates of NP concentrations should be made with great caution.