A Multicenter Evaluation of a Nongel Mechanical Separator Plasma Blood Collection Tube for Testing of Selected Therapeutic Drugs
Jazyk angličtina Země Velká Británie, Anglie Médium print
Typ dokumentu časopisecké články, multicentrická studie, pozorovací studie, práce podpořená grantem
PubMed
32603441
DOI
10.1093/jalm/jfaa033
PII: 5827479
Knihovny.cz E-zdroje
- Klíčová slova
- Barricor, TDM, gel adsorption, therapeutic drug monitoring, therapeutic drug stability,
- MeSH
- lidé MeSH
- monitorování léčiv přístrojové vybavení MeSH
- odběr vzorku krve přístrojové vybavení MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- pozorovací studie MeSH
- práce podpořená grantem MeSH
BACKGROUND: Some therapeutic drugs are unstable during sample storage in gel tubes. BD Vacutainer® Barricor™ Plasma Blood Collection Tube with nongel separator was compared with plasma gel tubes, BD Vacutainer PST™, PST II, and BD Vacutainer Serum Tube for acetaminophen, salicylate, digoxin, carbamazepine, phenytoin, valproic acid, and vancomycin during sample storage for up to 7 days. METHODS: Seven hospital sites enrolled 705 participants who were taking at least one selected drug. The study tubes were collected and tested at initial time (0 h), after 48 h of storage at room temperature and on day 7 (after additional 5 days of refrigerated storage). The performance of BD Barricor tube was evaluated for each drug by comparing BD Barricor samples with samples from the other tubes at 0 h from the same participant; stability was evaluated by comparing test results from the same tube at 0 h, 48 h, and 7 days. RESULTS: At 0 h, BD Barricor showed clinically equivalent results for selected therapeutic drugs compared with the other tubes, except phenytoin in BD PST. Phenytoin samples ≥20 µg/mL in BD PST had 10-12% lower values than samples in BD Barricor. During sample storage, all selected drugs remained stable for 7 days in BD Barricor and in serum aliquots. In BD PST, all drugs remained stable except phenytoin and carbamazepine and in BD PST II except for phenytoin. CONCLUSION: The BD Barricor Tube is effective for the collection and storage of plasma blood samples for therapeutic drug monitoring without sample aliquoting.
General University Hospital Prague Czech Republic
John T Mather Memorial Hospital Port Jefferson NY
School of Medicine University of Maryland MD USA
University Hospital Hradec Krάlόve Czech Republic
Citace poskytuje Crossref.org