BACKGROUND: Aveir DR (Abbott, Abbott Park, IL) is a dual-chamber leadless pacemaker (LP) system with distinct atrial and ventricular LPs (ALP, VLP) that communicate wirelessly to deliver atrioventricular synchronous pacing. Success rates of these implant-to-implant (i2i) transmissions have not been systematically evaluated. OBJECTIVE: This study aims to evaluate chronic i2i success rates in a clinical setting. METHODS: Patients meeting standard dual-chamber pacing indications were enrolled and implanted with dual-chamber LP systems as part of a prospective international clinical trial (Aveir DR i2i Study). The percent of successful i2i transmissions from ALP-to-VLP (A-to-V) and VLP-to-ALP (V-to-A) were interrogated from LPs in de novo patients using the device programmer at implant, discharge, and at 1, 3, and 6 months postimplant (1M, 3M, 6M). RESULTS: A total of 399 patients completed device implant and i2i diagnostic interrogation (62% male; age 69 years; 65% sinus node dysfunction, 32% atrioventricular [AV] block). Median A-to-V and V-to-A i2i success rates exceeded 90% of beats at all time-points from implant to 6M. The minority of patients with A-to-V or V-to-A i2i success in <70% of beats at implant (A-to-V: 19% of patients, V-to-A: 31% of patients) showed roughly 40% improvement by 1M, with this minority dropping to roughly 5% of patients by 6M. Improvement in i2i communication success may be attributed to reprogramming of i2i setting levels, natural changes in dominant posture, and device stabilization. CONCLUSION: Wireless implant-to-implant communication in the new dual-chamber leadless pacemaker system demonstrated successful transmissions in >90% of beats throughout the 6-month evaluation period. Communication success improved significantly over time postimplant for specific subgroups. CLINICAL TRIAL REGISTRATION: Aveir DR i2i Study, ClinicalTrials.gov ID NCT05252702.
- MeSH
- bezdrátová technologie MeSH
- design vybavení * MeSH
- kardiostimulace umělá metody MeSH
- kardiostimulátor * MeSH
- lidé středního věku MeSH
- lidé MeSH
- následné studie MeSH
- prospektivní studie MeSH
- senioři MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
In a pivotal trial (NCT05252702), the AVEIR DR (Abbott) leadless pacemaker system was found to be safe and effective in delivering DDDR synchronous atrial and ventricular pacing. This dual-chamber system employs 2 leadless pacemakers with implant-to-implant communication. Although implantation of the ventricular device as a single-chamber pacemaker has been well described, there are additional considerations surrounding the dual-chamber implantation procedure. Herein, we review the dual-chamber leadless pacemaker implantation workflow while providing guidance to optimize safe and effective implantation procedures.
- MeSH
- design vybavení * MeSH
- implantace protézy metody přístrojové vybavení MeSH
- kardiostimulace umělá metody MeSH
- kardiostimulátor * MeSH
- lidé MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- přehledy MeSH
BACKGROUND: The first dual-chamber leadless pacemaker (DC-LP) system consists of 2 separate atrial and ventricular devices that communicate to maintain synchronous atrioventricular pacing and sensing. The initial safety and efficacy were previously reported. OBJECTIVE: The purpose of this study was to evaluate the chronic electrical performance of the DC-LP system. METHODS: Patients meeting standard dual-chamber pacing indications were enrolled and implanted with the DC-LP system (Aveir DR, Abbott), including right atrial and ventricular helix-fixation LPs (atrial leadless pacemaker [ALP], ventricular leadless pacemaker [VLP]). Pacing capture threshold, sensed amplitude, and pacing impedance were collected using the device programmer at prespecified timepoints from 0-6 months postimplant. RESULTS: De novo devices were successfully implanted in 381 patients with complete 6-month data (62% male; age 69 ± 14 years; weight 82 ± 20 kg; 65% sinus nodal dysfunction, 30% atrioventricular block). ALPs were implanted predominantly in the right atrial appendage anterior base and VLPs primarily at the mid-to-apical right ventricular septum. From implant to 1 month, pacing capture thresholds (0.4-ms pulse width) improved in both ALPs (2.4 ± 1.5 V to 0.8 ± 0.8 V; P <.001) and VLPs (0.8 ± 0.6 V to 0.6 ± 0.4 V; P <.001). Sensed amplitudes improved in both ALPs (1.8 ± 1.3 mV to 3.4 ± 1.9 mV; P <.001) and VLPs (8.8 ± 4.0 mV to 11.7 ± 4.2 mV; P <.001). Impedances were stable in ALPs (334 ± 68 Ω to 329 ± 52 Ω; P = .17) and reduced in VLPs (789 ± 351 Ω to 646 ± 190 Ω; P <.001). Electrical measurements remained relatively stable from 1-6 months postimplant. No differences in electrical metrics were observed among ALP or VLP implant locations. CONCLUSION: This first in-human evaluation of the new dual-chamber leadless pacemaker system demonstrated reliable electrical performance throughout the initial 6-month evaluation period.
- MeSH
- atrioventrikulární blokáda terapie patofyziologie MeSH
- časové faktory MeSH
- design vybavení * MeSH
- kardiostimulace umělá * metody MeSH
- kardiostimulátor * MeSH
- lidé MeSH
- následné studie MeSH
- senioři MeSH
- srdeční komory patofyziologie MeSH
- syndrom chorého sinu terapie patofyziologie MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH