Detail
Article
Online article
FT
Medvik - BMC
  • Something wrong with this record ?

Determination of 1-methylhistidine and 3-methylhistidine by capillary and chip electrophoresis with contactless conductivity detection

Tůma P, Samcová E, Opekar F, Jurka V, Stulík K

. 2007 ; 28 (13) : 2174-2180.

Language English Country Germany

E-resources Online

NLK Wiley Online Library (archiv) from 1999-01-01 to 2012-12-31

CE with capacitively coupled contactless detection (C4D) was used to determine 3-methylhistidine (3-MH) and 1-methylhistidine (1-MH). The C4D response to 3-MH was studied in a BGE consisting of 500 mM acetic acid and ammonia at varying concentration and the results were compared with the theory. Complete separation of a model mixture of 3-MH, 1-MH, and histidine (His) was attained in two optimized BGEs, one containing 500 mM HAc, 20 mM NH4OH, and 0.1 % m/v hydroxyethylcellulose (HEC), pH 3.4 (I) and the other consisting of 100 mM morpholinoethanesulfonic acid (MES), 25 mM LiOH, and 0.1 % m/v HEC, pH 5.5 (II). These optimized BGEs were tested in CE/C4D analyses of urine. Promising results were obtained for separation and determination of 3-MH, 1-MH, and His on a silicon microchip, using aluminum strips as the C4D electrodes; the three analytes were baseline-separated within less than 30 s with a separation channel effective length of 38 mm. The LOD were satisfactory and amounted to 26.4 microM for 3-MH and 18.3 microM for 1-MH.

000      
00000naa 2200000 a 4500
001      
bmc10001014
003      
CZ-PrNML
005      
20111210154844.0
008      
100118s2007 gw e eng||
009      
AR
040    __
$a ABA008 $b cze $c ABA008 $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a gw
100    1_
$a Tůma, Petr $7 xx0105786
245    10
$a Determination of 1-methylhistidine and 3-methylhistidine by capillary and chip electrophoresis with contactless conductivity detection / $c Tůma P, Samcová E, Opekar F, Jurka V, Stulík K
314    __
$a Institute of Biochemistry, Cell and Molecular Biology, 3rd Faculty of Medicine, Charles University, Prague, Czech Republic. petr.tuma@lf3.cuni.cz
520    9_
$a CE with capacitively coupled contactless detection (C4D) was used to determine 3-methylhistidine (3-MH) and 1-methylhistidine (1-MH). The C4D response to 3-MH was studied in a BGE consisting of 500 mM acetic acid and ammonia at varying concentration and the results were compared with the theory. Complete separation of a model mixture of 3-MH, 1-MH, and histidine (His) was attained in two optimized BGEs, one containing 500 mM HAc, 20 mM NH4OH, and 0.1 % m/v hydroxyethylcellulose (HEC), pH 3.4 (I) and the other consisting of 100 mM morpholinoethanesulfonic acid (MES), 25 mM LiOH, and 0.1 % m/v HEC, pH 5.5 (II). These optimized BGEs were tested in CE/C4D analyses of urine. Promising results were obtained for separation and determination of 3-MH, 1-MH, and His on a silicon microchip, using aluminum strips as the C4D electrodes; the three analytes were baseline-separated within less than 30 s with a separation channel effective length of 38 mm. The LOD were satisfactory and amounted to 26.4 microM for 3-MH and 18.3 microM for 1-MH.
650    _2
$a financování organizované $7 D005381
650    _2
$a elektrická vodivost $7 D004553
650    _2
$a elektroforéza kapilární $x metody $7 D019075
650    _2
$a elektroforéza mikročipová $x metody $x přístrojové vybavení $7 D046209
650    _2
$a lidé $7 D006801
650    _2
$a methylhistidiny $x analýza $x moč $7 D008762
700    1_
$a Samcová, Eva, $d 1947- $7 nlk19990073817
700    1_
$a Opekar, František. $7 _BN004560
700    1_
$a Jurka, Vlastimil. $7 xx0303001
700    1_
$a Štulík, Karel, $d 1941-2013 $7 jk01131560
773    0_
$w MED00001508 $t Electrophoresis $g Roč. 28, č. 13 (2007), s. 2174-2180 $x 0173-0835
910    __
$a ABA008 $b x $y 8
990    __
$a 20100114083259 $b ABA008
991    __
$a 20100215154206 $b ABA008
999    __
$a ok $b bmc $g 703742 $s 566184
BAS    __
$a 3
BMC    __
$a 2007 $b 28 $c 13 $d 2174-2180 $i 0173-0835 $m Electrophoresis $x MED00001508
LZP    __
$a 2010-b1/vtme

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...