Accurate normalization factor for wire myography of rat femoral artery
P. Slezák, I. Waczulíková, P. Bališ, A. Púzserová
Language English Country Czech Republic
NLK
Directory of Open Access Journals
from 1991
Free Medical Journals
from 1998
ProQuest Central
from 2005-01-01
Medline Complete (EBSCOhost)
from 2006-01-01
Nursing & Allied Health Database (ProQuest)
from 2005-01-01
Health & Medicine (ProQuest)
from 2005-01-01
ROAD: Directory of Open Access Scholarly Resources
from 1998
- MeSH
- Femoral Artery physiology MeSH
- Mesenteric Arteries physiology MeSH
- Financing, Organized MeSH
- Rats MeSH
- Myography methods instrumentation MeSH
- Animals MeSH
- Check Tag
- Rats MeSH
- Animals MeSH
Wire myograph is a device for the in vitro investigation of both, active and passive properties of arteries. Arteries from a variety of animal species, pathological states, and vascular beds were investigated using this method. We focus on the normalization procedure which is aimed to standardize experimental settings and, in part, to simulate physiological conditions. During normalization, it is determined the internal circumference of a vessel stretched to a tension that corresponds to the transmural pressure of 100 mm Hg (IC100). Once it is determined, the internal circumference is traditionally set to (0.9·IC100). However, this constant 0.9, called also the normalization factor (NF), was experimentally determined for rat small mesenteric arteries only. Therefore, the aim of our work was to show the influence of different NFs on the passive tension and reactivity of both, rat femoral arteries (FA) and the first branches of superior mesenteric arteries (MA). We found out that the maximal active wall tension of the FA was achieved at the NF value of 1.1, and that of the MA at 0.9. Considering the values of the active wall tension we suggest that higher reactivity and better signal-tonoise ratio in FA can be achieved when the NF is set at least to 1.0.
References provided by Crossref.org
Lit.: 13
- 000
- 00000naa 2200000 a 4500
- 001
- bmc11004158
- 003
- CZ-PrNML
- 005
- 20120918081526.0
- 008
- 110306s2010 xr e eng||
- 009
- AR
- 024 7_
- $a 10.33549/physiolres.932043 $2 doi
- 040 __
- $a ABA008 $b cze $c ABA008 $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Slezák, Peter $7 xx0139239
- 245 10
- $a Accurate normalization factor for wire myography of rat femoral artery / $c P. Slezák, I. Waczulíková, P. Bališ, A. Púzserová
- 314 __
- $a Institute of Normal and Pathological Physiology, Slovak Academy of Sciences, Bratislava
- 504 __
- $a Lit.: 13
- 520 9_
- $a Wire myograph is a device for the in vitro investigation of both, active and passive properties of arteries. Arteries from a variety of animal species, pathological states, and vascular beds were investigated using this method. We focus on the normalization procedure which is aimed to standardize experimental settings and, in part, to simulate physiological conditions. During normalization, it is determined the internal circumference of a vessel stretched to a tension that corresponds to the transmural pressure of 100 mm Hg (IC100). Once it is determined, the internal circumference is traditionally set to (0.9·IC100). However, this constant 0.9, called also the normalization factor (NF), was experimentally determined for rat small mesenteric arteries only. Therefore, the aim of our work was to show the influence of different NFs on the passive tension and reactivity of both, rat femoral arteries (FA) and the first branches of superior mesenteric arteries (MA). We found out that the maximal active wall tension of the FA was achieved at the NF value of 1.1, and that of the MA at 0.9. Considering the values of the active wall tension we suggest that higher reactivity and better signal-tonoise ratio in FA can be achieved when the NF is set at least to 1.0.
- 650 _2
- $a financování organizované $7 D005381
- 650 _2
- $a zvířata $7 D000818
- 650 _2
- $a arteria femoralis $x fyziologie $7 D005263
- 650 _2
- $a arteriae mesentericae $x fyziologie $7 D008638
- 650 _2
- $a myografie $x metody $x přístrojové vybavení $7 D009213
- 650 _2
- $a krysa rodu Rattus $7 D051381
- 700 1_
- $a Waczulíková, Iveta. $7 xx0238686
- 700 1#
- $a Bališ, Peter. $7 xx0198090
- 700 1_
- $a Púzserová, Angelika $7 xx0091370
- 773 0_
- $w MED00003824 $t Physiological research $g Roč. 59, č. 6 (2010), s. 1033-1036 $x 0862-8408
- 856 41
- $u http://www.biomed.cas.cz/physiolres/pdf/59/59_1033.pdf $y plný text volně přístupný
- 910 __
- $a ABA008 $b A 4120 $c 266 $y 1
- 990 __
- $a 20110303103805 $b ABA008
- 991 __
- $a 20120918081711 $b ABA008
- 999 __
- $a ok $b bmc $g 831498 $s 696187
- BAS __
- $a 3
- BMC __
- $a 2010 $b 59 $c 6 $d 1033-1036 $m Physiological research $x MED00003824
- LZP __
- $a 2011-14/ipme