Something wrong with this record ?
Practical method transfer from high performance liquid chromatography to ultra-high performance liquid chromatography: the importance of frictional heating
L. Nováková, JL. Veuthey, D. Guillarme,
Language English Country Netherlands
Document type Journal Article, Research Support, Non-U.S. Gov't
- MeSH
- Hydrogen-Ion Concentration MeSH
- Pharmaceutical Preparations chemistry isolation & purification MeSH
- Buffers MeSH
- Pressure MeSH
- Friction MeSH
- Hot Temperature MeSH
- Chromatography, High Pressure Liquid instrumentation methods MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
In theory, with identical stationary phase chemistry, the transfer of an HPLC method to UHPLC conditions is straightforward and necessitates the calculation of new conditions based on column and instrument geometries. Occasionally, undesirable changes in selectivity, retention or efficiency have been reported and have been attributed to a frictional heating phenomenon that is due to the elevated generated pressure drop. In the present study, the frictional heating in a UHPLC system was evaluated experimentally under gradient elution conditions (acetonitrile/buffer at pH 3 and 9) with generated pressure drops in the range of 100-1000 bar on both 1.0mm and 2.1mm I.D. columns using a mixture of 10 representative basic, acidic and neutral pharmaceutical compounds. Under adiabatic conditions (i.e., still-air oven), the longitudinal temperature gradient was estimated at +4 °C, +8 °C and +16 °C at 300, 600 and 1000 bar, respectively, on a 2.1mm I.D. column using an empirical measurement procedure. With the 1.0mm I.D. column, these values were reduced to +3 °C, +6 °C and +12 °C, respectively. Finally, various approaches to eliminate or at least to reduce the effect of frictional heating are briefly discussed.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12022358
- 003
- CZ-PrNML
- 005
- 20170410095250.0
- 007
- ta
- 008
- 120806s2011 ne f 000 0#eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.chroma.2011.08.096 $2 doi
- 035 __
- $a (PubMed)21944847
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Nováková, Lucie $u Department of Analytical Chemistry, Faculty of Pharmacy, Charles University, Heyrovského 1203, 500 05 Hradec Králové, Czech Republic.
- 245 10
- $a Practical method transfer from high performance liquid chromatography to ultra-high performance liquid chromatography: the importance of frictional heating / $c L. Nováková, JL. Veuthey, D. Guillarme,
- 520 9_
- $a In theory, with identical stationary phase chemistry, the transfer of an HPLC method to UHPLC conditions is straightforward and necessitates the calculation of new conditions based on column and instrument geometries. Occasionally, undesirable changes in selectivity, retention or efficiency have been reported and have been attributed to a frictional heating phenomenon that is due to the elevated generated pressure drop. In the present study, the frictional heating in a UHPLC system was evaluated experimentally under gradient elution conditions (acetonitrile/buffer at pH 3 and 9) with generated pressure drops in the range of 100-1000 bar on both 1.0mm and 2.1mm I.D. columns using a mixture of 10 representative basic, acidic and neutral pharmaceutical compounds. Under adiabatic conditions (i.e., still-air oven), the longitudinal temperature gradient was estimated at +4 °C, +8 °C and +16 °C at 300, 600 and 1000 bar, respectively, on a 2.1mm I.D. column using an empirical measurement procedure. With the 1.0mm I.D. column, these values were reduced to +3 °C, +6 °C and +12 °C, respectively. Finally, various approaches to eliminate or at least to reduce the effect of frictional heating are briefly discussed.
- 650 _2
- $a pufry $7 D002021
- 650 _2
- $a vysokoúčinná kapalinová chromatografie $x přístrojové vybavení $x metody $7 D002851
- 650 _2
- $a tření $7 D017276
- 650 _2
- $a vysoká teplota $7 D006358
- 650 _2
- $a koncentrace vodíkových iontů $7 D006863
- 650 _2
- $a léčivé přípravky $x chemie $x izolace a purifikace $7 D004364
- 650 _2
- $a tlak $7 D011312
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Veuthey, Jean Luc
- 700 1_
- $a Guillarme, Davy
- 773 0_
- $w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1218, č. 44 (2011), s. 7971-7981
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/21944847 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y m $z 0
- 990 __
- $a 20120806 $b ABA008
- 991 __
- $a 20170410095548 $b ABA008
- 999 __
- $a ok $b bmc $g 944271 $s 779655
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2011 $b 1218 $c 44 $d 7971-7981 $e 20110909 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
- LZP __
- $a Pubmed-20120806/12/01