Practical method transfer from high performance liquid chromatography to ultra-high performance liquid chromatography: the importance of frictional heating

. 2011 Nov 04 ; 1218 (44) : 7971-81. [epub] 20110909

Jazyk angličtina Země Nizozemsko Médium print-electronic

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/pmid21944847
Odkazy

PubMed 21944847
DOI 10.1016/j.chroma.2011.08.096
PII: S0021-9673(11)01357-4
Knihovny.cz E-zdroje

In theory, with identical stationary phase chemistry, the transfer of an HPLC method to UHPLC conditions is straightforward and necessitates the calculation of new conditions based on column and instrument geometries. Occasionally, undesirable changes in selectivity, retention or efficiency have been reported and have been attributed to a frictional heating phenomenon that is due to the elevated generated pressure drop. In the present study, the frictional heating in a UHPLC system was evaluated experimentally under gradient elution conditions (acetonitrile/buffer at pH 3 and 9) with generated pressure drops in the range of 100-1000 bar on both 1.0mm and 2.1mm I.D. columns using a mixture of 10 representative basic, acidic and neutral pharmaceutical compounds. Under adiabatic conditions (i.e., still-air oven), the longitudinal temperature gradient was estimated at +4 °C, +8 °C and +16 °C at 300, 600 and 1000 bar, respectively, on a 2.1mm I.D. column using an empirical measurement procedure. With the 1.0mm I.D. column, these values were reduced to +3 °C, +6 °C and +12 °C, respectively. Finally, various approaches to eliminate or at least to reduce the effect of frictional heating are briefly discussed.

Citace poskytuje Crossref.org

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...