Something wrong with this record ?
Organ and effective doses from verification techniques in image-guided radiotherapy
V. Dufek, I. Horakova, L. Novak,
Language English Country Great Britain
Document type Journal Article, Research Support, Non-U.S. Gov't
PubMed
21816726
DOI
10.1093/rpd/ncr335
Knihovny.cz E-resources
- MeSH
- Radiotherapy Dosage MeSH
- Phantoms, Imaging MeSH
- Head radiography MeSH
- Neck radiography MeSH
- Humans MeSH
- Prostatic Neoplasms radiography radiotherapy MeSH
- Pelvis radiography MeSH
- Cone-Beam Computed Tomography MeSH
- Radiotherapy, Image-Guided MeSH
- Thermoluminescent Dosimetry MeSH
- Check Tag
- Humans MeSH
- Male MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
The purpose of this work was an evaluation of organ doses and effective doses from three verification techniques in Image-Guided Radiotherapy: from kilovoltage (kV) cone beam computed tomography (CBCT) scans, from two orthogonal kV images and from two orthogonal megavoltage (MV) images for two different treatment sites: pelvis and head and neck (H&N). For comparison reasons, organ doses and effective doses from prostate and H&N radiotherapy were also evaluated. Measurements of organ doses were performed in a male anthropomorphic Rando phantom by means of thermoluminescent dosemeters. In this investigation, measured organ doses from one CBCT scan, from two MV images and from two kV images of pelvis represent typically 1-6, 1-10 and 0.05-1 %, respectively, of organ doses resulting from one fraction of prostate radiotherapy. The maximum effective doses from CBCT scans, kV images and MV images of pelvis are 5.6, 0.8 and 11.9 mSv, respectively.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12022570
- 003
- CZ-PrNML
- 005
- 20170410094313.0
- 007
- ta
- 008
- 120806s2011 xxk f 000 0#eng||
- 009
- AR
- 024 7_
- $a 10.1093/rpd/ncr335 $2 doi
- 035 __
- $a (PubMed)21816726
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxk
- 100 1_
- $a Dufek, V $u National Radiation Protection Institute, Bartoskova 28, 14000 Prague, Czech Republic. vladimir.dufek@suro.cz
- 245 10
- $a Organ and effective doses from verification techniques in image-guided radiotherapy / $c V. Dufek, I. Horakova, L. Novak,
- 520 9_
- $a The purpose of this work was an evaluation of organ doses and effective doses from three verification techniques in Image-Guided Radiotherapy: from kilovoltage (kV) cone beam computed tomography (CBCT) scans, from two orthogonal kV images and from two orthogonal megavoltage (MV) images for two different treatment sites: pelvis and head and neck (H&N). For comparison reasons, organ doses and effective doses from prostate and H&N radiotherapy were also evaluated. Measurements of organ doses were performed in a male anthropomorphic Rando phantom by means of thermoluminescent dosemeters. In this investigation, measured organ doses from one CBCT scan, from two MV images and from two kV images of pelvis represent typically 1-6, 1-10 and 0.05-1 %, respectively, of organ doses resulting from one fraction of prostate radiotherapy. The maximum effective doses from CBCT scans, kV images and MV images of pelvis are 5.6, 0.8 and 11.9 mSv, respectively.
- 650 _2
- $a počítačová tomografie s kuželovým svazkem $7 D054893
- 650 _2
- $a hlava $x radiografie $7 D006257
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 _2
- $a krk $x radiografie $7 D009333
- 650 _2
- $a pánev $x radiografie $7 D010388
- 650 _2
- $a fantomy radiodiagnostické $7 D019047
- 650 _2
- $a nádory prostaty $x radiografie $x radioterapie $7 D011471
- 650 _2
- $a celková dávka radioterapie $7 D011879
- 650 _2
- $a radioterapie řízená obrazem $7 D061089
- 650 _2
- $a termoluminiscenční dozimetrie $7 D013819
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Horakova, I
- 700 1_
- $a Novak, L
- 773 0_
- $w MED00004029 $t Radiation protection dosimetry $x 1742-3406 $g Roč. 147, č. 1-2 (2011), s. 277-280
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/21816726 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y m $z 0
- 990 __
- $a 20120806 $b ABA008
- 991 __
- $a 20170410094611 $b ABA008
- 999 __
- $a ok $b bmc $g 944483 $s 779867
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2011 $b 147 $c 1-2 $d 277-280 $e 20110804 $i 1742-3406 $m Radiation protection dosimetry $n Radiat Prot Dosimetry $x MED00004029
- LZP __
- $a Pubmed-20120806/12/01