• Something wrong with this record ?

Finite element analysis for the evaluation of the structural behaviour, of a prosthesis for trans-tibial amputees

M. Omasta, D. Paloušek, T. Návrat, J. Rosický,

. 2012 ; 34 (1) : 38-45.

Language English Country England, Great Britain

Document type Journal Article, Research Support, Non-U.S. Gov't

The finite element analysis (FEA) has been identified as a useful tool for the stress and strain behaviour determination in lower limb prosthetics. The residual limb and prosthetic socket interface was the main subject of interest in previous studies. This paper focuses on the finite element analysis for the evaluation of structural behaviour of the Sure-flex™ prosthetic foot and other load-bearing components. A prosthetic socket was not included in the FEA. An approach for the finite element modelling including foot analysis, reverse engineering and material property testing was used. The foot analysis incorporated ground reaction forces measurement, motion analysis and strain gauge analysis. For the material model determination, non-destructive laboratory testing and its FE simulation was used. A new, realistic way of load application is presented along with a detailed investigation of stress distribution in the load-bearing components of the prosthesis. A novel approach for numerical and experimental agreement determination was introduced. This showed differences in the strain on the pylon between the experimental and the numerical model within 30% for the anteroposterior bending and up to 25% for the compression. The highest von Mises stresses were found on the foot-pylon connecting component at toe off. Peak stress of 216MPa occurred on the posterior adjusting screw and maximum stress of 156MPa was found at the neck of the male pyramid.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc12024413
003      
CZ-PrNML
005      
20121207122902.0
007      
ta
008      
120815s2012 enk f 000 0#eng||
009      
AR
024    7_
$a 10.1016/j.medengphy.2011.06.014 $2 doi
035    __
$a (PubMed)21764351
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Omasta, Milan $u Institute of Machine and Industrial Design, Faculty of Mechanical Engineering, Brno University of Technology, Czech Republic. yomast00@stud.fme.vutbr.cz
245    10
$a Finite element analysis for the evaluation of the structural behaviour, of a prosthesis for trans-tibial amputees / $c M. Omasta, D. Paloušek, T. Návrat, J. Rosický,
520    9_
$a The finite element analysis (FEA) has been identified as a useful tool for the stress and strain behaviour determination in lower limb prosthetics. The residual limb and prosthetic socket interface was the main subject of interest in previous studies. This paper focuses on the finite element analysis for the evaluation of structural behaviour of the Sure-flex™ prosthetic foot and other load-bearing components. A prosthetic socket was not included in the FEA. An approach for the finite element modelling including foot analysis, reverse engineering and material property testing was used. The foot analysis incorporated ground reaction forces measurement, motion analysis and strain gauge analysis. For the material model determination, non-destructive laboratory testing and its FE simulation was used. A new, realistic way of load application is presented along with a detailed investigation of stress distribution in the load-bearing components of the prosthesis. A novel approach for numerical and experimental agreement determination was introduced. This showed differences in the strain on the pylon between the experimental and the numerical model within 30% for the anteroposterior bending and up to 25% for the compression. The highest von Mises stresses were found on the foot-pylon connecting component at toe off. Peak stress of 216MPa occurred on the posterior adjusting screw and maximum stress of 156MPa was found at the neck of the male pyramid.
650    _2
$a dospělí $7 D000328
650    _2
$a amputovaní $7 D000674
650    _2
$a analýza metodou konečných prvků $7 D020342
650    _2
$a lidé $7 D006801
650    _2
$a mužské pohlaví $7 D008297
650    _2
$a protézy a implantáty $7 D019736
650    _2
$a protézy - design $7 D011474
650    _2
$a mechanický stres $7 D013314
650    _2
$a tibie $7 D013977
650    _2
$a zatížení muskuloskeletálního systému $7 D016474
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Paloušek, David
700    1_
$a Návrat, Tomáš
700    1_
$a Rosický, Jiří
773    0_
$w MED00008431 $t Medical engineering & physics $x 1873-4030 $g Roč. 34, č. 1 (2012), s. 38-45
856    41
$u https://pubmed.ncbi.nlm.nih.gov/21764351 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y m
990    __
$a 20120815 $b ABA008
991    __
$a 20121207122937 $b ABA008
999    __
$a ok $b bmc $g 946561 $s 781741
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2012 $b 34 $c 1 $d 38-45 $i 1873-4030 $m Medical engineering & physics $n Med Eng Phys $x MED00008431
LZP    __
$a Pubmed-20120815/12/02

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...