The relationship of polyethylene wear to particle size, distribution, and number: A possible factor explaining the risk of osteolysis after hip arthroplasty
J. Gallo, M. Slouf, SB. Goodman,
Jazyk angličtina Země Spojené státy americké
Typ dokumentu časopisecké články, práce podpořená grantem
PubMed
20524192
DOI
10.1002/jbm.b.31638
Knihovny.cz E-zdroje
- MeSH
- algoritmy MeSH
- lidé MeSH
- náhrada kyčelního kloubu škodlivé účinky MeSH
- osteolýza etiologie MeSH
- polyethyleny chemie MeSH
- selhání protézy MeSH
- velikost částic MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
The most critical factor in the development of periprosthetic osteolysis (OL) in total hip arthroplasty (THA) is the biological reaction to wear debris. This reaction is dependent, in part, on the size and concentration of particles, which are determined predominantly by the polyethylene (PE) wear rate. This implies that the risk for developing OL and prosthesis failure can be estimated from wear measurements. We developed a computational algorithm for calculating the total number of PE particles for volumetric wear when particle size and distribution are known. We found that: (i) total number of PE wear particles decreases up to 5 orders of magnitude if the average size of particles increases and the total volumetric wear remains constant; (ii) total amount of PE wear particles decreases up to 4 orders of magnitude if the width of the distribution increases and total volumetric wear remains constant; (iii) for the same volumetric wear, the number of particles significantly decreases/increases with the increase/decrease in their average size and range. These findings suggest that the risk for the development of OL in THA cannot be simply estimated from the volumetric wear alone.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12025669
- 003
- CZ-PrNML
- 005
- 20121207114300.0
- 007
- ta
- 008
- 120817s2010 xxu f 000 0#eng||
- 009
- AR
- 024 7_
- $a 10.1002/jbm.b.31638 $2 doi
- 035 __
- $a (PubMed)20524192
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxu
- 100 1_
- $a Gallo, Jiri $u Department of Orthopaedics, Teaching Hospital, Palacký University, Olomouc 77520, Czech Republic. jiri.gallo@volny.cz
- 245 14
- $a The relationship of polyethylene wear to particle size, distribution, and number: A possible factor explaining the risk of osteolysis after hip arthroplasty / $c J. Gallo, M. Slouf, SB. Goodman,
- 520 9_
- $a The most critical factor in the development of periprosthetic osteolysis (OL) in total hip arthroplasty (THA) is the biological reaction to wear debris. This reaction is dependent, in part, on the size and concentration of particles, which are determined predominantly by the polyethylene (PE) wear rate. This implies that the risk for developing OL and prosthesis failure can be estimated from wear measurements. We developed a computational algorithm for calculating the total number of PE particles for volumetric wear when particle size and distribution are known. We found that: (i) total number of PE wear particles decreases up to 5 orders of magnitude if the average size of particles increases and the total volumetric wear remains constant; (ii) total amount of PE wear particles decreases up to 4 orders of magnitude if the width of the distribution increases and total volumetric wear remains constant; (iii) for the same volumetric wear, the number of particles significantly decreases/increases with the increase/decrease in their average size and range. These findings suggest that the risk for the development of OL in THA cannot be simply estimated from the volumetric wear alone.
- 650 _2
- $a algoritmy $7 D000465
- 650 _2
- $a náhrada kyčelního kloubu $x škodlivé účinky $7 D019644
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a osteolýza $x etiologie $7 D010014
- 650 _2
- $a velikost částic $7 D010316
- 650 _2
- $a polyethyleny $x chemie $7 D011095
- 650 _2
- $a selhání protézy $7 D011475
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Slouf, Miroslav
- 700 1_
- $a Goodman, Stuart B
- 773 0_
- $w MED00007497 $t Journal of biomedical materials research. Part B, Applied biomaterials $x 1552-4981 $g Roč. 94, č. 1 (2010), s. 171-7
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/20524192 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y m
- 990 __
- $a 20120817 $b ABA008
- 991 __
- $a 20121207114334 $b ABA008
- 999 __
- $a ok $b bmc $g 947711 $s 783015
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2010 $b 94 $c 1 $d 171-7 $i 1552-4981 $m Journal of biomedical materials research. Part B, Applied biomaterials $n J Biomed Mater Res $x MED00007497
- LZP __
- $a Pubmed-20120817/10/03