Corrosion behaviour of TiN and ZrN in the environment containing fluoride ions
L. Joska, J. Fojt, M. Hradilova, F. Hnilica, L. Cvrcek,
Language English Country England, Great Britain
Document type Comparative Study, Journal Article, Research Support, Non-U.S. Gov't
- MeSH
- Electrochemical Techniques MeSH
- Fluorides chemistry MeSH
- Hydrogen-Ion Concentration MeSH
- Corrosion MeSH
- Humans MeSH
- Saliva, Artificial chemistry MeSH
- Buffers MeSH
- Saliva chemistry MeSH
- Titanium chemistry MeSH
- Mouth chemistry MeSH
- Zirconium chemistry MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
- Comparative Study MeSH
Nowadays, a wide range of materials for human implants is used. To reach the required properties of implants, coatings are applied in some cases. This contribution is focused on the corrosion properties of TiN and ZrN layers on cp-titanium (commercially pure titanium) under environment modelling conditions in an oral cavity. Measurements were done in artificial saliva and a physiological solution unbuffered and buffered to a pH value of 4.2 with the addition of fluoride ions up to 4000 ppm. Standard corrosion electrochemical techniques were applied. Both types of layers were stable in both model saliva and physiological solution with non-adjusted pH. The decrease in pH to 4.2 resulted in a minor decrease of corrosion resistance in all cases, but polarization resistance was still in the order of 10(5) Ω cm². An important change in a specimens' behaviour was noticed in the presence of fluoride ions. TiN was stable in the highest concentration of fluorides used. The ZrN layers were destabilized in an environment containing a few hundred ppm of fluoride ions. As for TiN, the decisive factor is the influence of porosity; the corrosion resistance of ZrN is limited. From the corrosion point of view, the application of the TiN-based barrier layers in dental implantology is more advisable than the use of ZrN, provided that the application of a barrier is inevitable.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12026772
- 003
- CZ-PrNML
- 005
- 20160307145003.0
- 007
- ta
- 008
- 120816s2010 enk f 000 0#eng||
- 009
- AR
- 024 7_
- $a 10.1088/1748-6041/5/5/054108 $2 doi
- 035 __
- $a (PubMed)20876958
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Joska, Luděk $u Department of Metals and Corrosion Engineering, Institute of Chemical Technology, Technicka 5, 16628 Prague 6, Czech Republic. joskal@vscht.cz $7 xx0103993
- 245 10
- $a Corrosion behaviour of TiN and ZrN in the environment containing fluoride ions / $c L. Joska, J. Fojt, M. Hradilova, F. Hnilica, L. Cvrcek,
- 520 9_
- $a Nowadays, a wide range of materials for human implants is used. To reach the required properties of implants, coatings are applied in some cases. This contribution is focused on the corrosion properties of TiN and ZrN layers on cp-titanium (commercially pure titanium) under environment modelling conditions in an oral cavity. Measurements were done in artificial saliva and a physiological solution unbuffered and buffered to a pH value of 4.2 with the addition of fluoride ions up to 4000 ppm. Standard corrosion electrochemical techniques were applied. Both types of layers were stable in both model saliva and physiological solution with non-adjusted pH. The decrease in pH to 4.2 resulted in a minor decrease of corrosion resistance in all cases, but polarization resistance was still in the order of 10(5) Ω cm². An important change in a specimens' behaviour was noticed in the presence of fluoride ions. TiN was stable in the highest concentration of fluorides used. The ZrN layers were destabilized in an environment containing a few hundred ppm of fluoride ions. As for TiN, the decisive factor is the influence of porosity; the corrosion resistance of ZrN is limited. From the corrosion point of view, the application of the TiN-based barrier layers in dental implantology is more advisable than the use of ZrN, provided that the application of a barrier is inevitable.
- 650 _2
- $a pufry $7 D002021
- 650 _2
- $a koroze $7 D003343
- 650 _2
- $a elektrochemické techniky $7 D055664
- 650 _2
- $a fluoridy $x chemie $7 D005459
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a koncentrace vodíkových iontů $7 D006863
- 650 _2
- $a ústa $x chemie $7 D009055
- 650 _2
- $a sliny $x chemie $7 D012463
- 650 _2
- $a náhrada slin $x chemie $7 D012464
- 650 _2
- $a titan $x chemie $7 D014025
- 650 _2
- $a zirkonium $x chemie $7 D015040
- 655 _2
- $a srovnávací studie $7 D003160
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Fojt, Jaroslav $7 _AN077081
- 700 1_
- $a Hradilova, Monika
- 700 1_
- $a Hnilica, František $7 ntka172570
- 700 1_
- $a Cvrček, Ladislav $7 xx0101233
- 773 0_
- $w MED00180190 $t Biomedical materials (Bristol, England) $x 1748-605X $g Roč. 5, č. 5 (2010), s. 054108
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/20876958 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y m $z 0
- 990 __
- $a 20120816 $b ABA008
- 991 __
- $a 20160307145017 $b ABA008
- 999 __
- $a ok $b bmc $g 948814 $s 784118
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2010 $b 5 $c 5 $d 054108 $e 20100928 $i 1748-605X $m Biomedical materials $n Biomed Mater $x MED00180190
- LZP __
- $b NLK113 $a Pubmed-20120816/11/01