Possibilities of retention prediction in fast gradient liquid chromatography. Part 1: Comparison of separation on packed fully porous, nonporous and monolithic columns
K. Vyňuchalová, P. Jandera,
Jazyk angličtina Země Nizozemsko
Typ dokumentu srovnávací studie, časopisecké články, práce podpořená grantem
- MeSH
- chromatografie kapalinová přístrojové vybavení metody MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
- srovnávací studie MeSH
Due to ever increasing importance of short analysis times in modern HPLC practice, fast gradient elution is becoming widely used both in one-dimensional and two-dimensional separations, especially in the second dimension where the speed of separation is of primary importance. For method development and optimization, prediction of retention in gradient elution from the isocratic data is very important. This is principally possible using the well-established theory of gradient elution, especially for reversed-phase separations. Feasibility of these approaches has been proven in conventional liquid chromatography and commercial prediction software is available for this purpose. However, fast gradient chromatography employing short columns and gradient times in between 1 and 2 min or even less, impose additional stringent demands on the accuracy of prediction, due to rapid changes in mobile phase composition and increased role of extra-column contributions and instrumental dwell volume on the retention. In the present work, possibilities of prediction of the gradient elution times from the isocratic data measured only at two mobile phase concentrations were investigated in fast chromatography of alkylbenzenes, phenylurea and triazine pesticides on five columns differing in lengths: two monolithic ones and three packed with particles of different size. In the second dimension, gradient times are strongly restricted by the short cycle frequency. The effects of the flow-rate on the elution times, on the accuracy of their prediction and on peak capacity were studied under time constraint conditions. Gradient theory can be useful for the optimization of fast gradients.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc13024012
- 003
- CZ-PrNML
- 005
- 20130710114758.0
- 007
- ta
- 008
- 130703s2013 ne f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.chroma.2012.12.058 $2 doi
- 035 __
- $a (PubMed)23336942
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Vyňuchalová, Kateřina $u Department of Analytical Chemistry, Faculty of Chemical Technology, University of Pardubice, Studentská 573, CZ-53210 Pardubice, Czech Republic.
- 245 10
- $a Possibilities of retention prediction in fast gradient liquid chromatography. Part 1: Comparison of separation on packed fully porous, nonporous and monolithic columns / $c K. Vyňuchalová, P. Jandera,
- 520 9_
- $a Due to ever increasing importance of short analysis times in modern HPLC practice, fast gradient elution is becoming widely used both in one-dimensional and two-dimensional separations, especially in the second dimension where the speed of separation is of primary importance. For method development and optimization, prediction of retention in gradient elution from the isocratic data is very important. This is principally possible using the well-established theory of gradient elution, especially for reversed-phase separations. Feasibility of these approaches has been proven in conventional liquid chromatography and commercial prediction software is available for this purpose. However, fast gradient chromatography employing short columns and gradient times in between 1 and 2 min or even less, impose additional stringent demands on the accuracy of prediction, due to rapid changes in mobile phase composition and increased role of extra-column contributions and instrumental dwell volume on the retention. In the present work, possibilities of prediction of the gradient elution times from the isocratic data measured only at two mobile phase concentrations were investigated in fast chromatography of alkylbenzenes, phenylurea and triazine pesticides on five columns differing in lengths: two monolithic ones and three packed with particles of different size. In the second dimension, gradient times are strongly restricted by the short cycle frequency. The effects of the flow-rate on the elution times, on the accuracy of their prediction and on peak capacity were studied under time constraint conditions. Gradient theory can be useful for the optimization of fast gradients.
- 650 _2
- $a chromatografie kapalinová $x přístrojové vybavení $x metody $7 D002853
- 655 _2
- $a srovnávací studie $7 D003160
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Jandera, Pavel $u -
- 773 0_
- $w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1278(2013), s. 37-45
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/23336942 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20130703 $b ABA008
- 991 __
- $a 20130710115221 $b ABA008
- 999 __
- $a ok $b bmc $g 987692 $s 822392
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2013 $b 1278 $d 37-45 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
- LZP __
- $a Pubmed-20130703