Thermomechanical properties of nickel-titanium closed-coil springs and their implications for clinical practice

A. Bezrouk, L. Balsky, M. Smutny, I. Selke Krulichova, J. Zahora, J. Hanus, TR. Meling,

. 2014 ; 146 (3) : 319-27.

Jazyk angličtina Země Spojené státy americké

Typ dokumentu srovnávací studie, časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/bmc15007942

INTRODUCTION: The aim was to study nickel-titanium closed-coil springs in a clinically relevant test setting with respect to the accuracy of the "preactivation" for nickel-titanium closed-coil springs application and whether it is possible to keep activation forces constant during the whole time of treatment. METHODS: We tested 10 types of springs from 5 manufacturers under clinically relevant conditions, allowing us to study the interactions between load and temperature over time. Hystereses were compared using t tests. RESULTS: Springs with a large mechanical hysteresis also showed a large thermal hysteresis. After heating shock, these springs showed intensive force spikes and persistent high loads. Some springs showed negligible thermal and mechanical hysteresis. Such springs never showed any clinically significant persistent high loads. CONCLUSIONS: Springs with a large hysteresis were unable to keep activation forces constant during the whole time of treatment even after any preactivation, and they might cause persistently high loads and possibly overloading. Only springs with minor hysteresis, low temperature dependence of force, and a clinically useful plateau have the following clinical advantages: reduced chair time, optimal rates of tooth movement, reproducible clinical results, and conservation of anchorage.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc15007942
003      
CZ-PrNML
005      
20241120091956.0
007      
ta
008      
150306s2014 xxu f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.ajodo.2014.05.025 $2 doi
035    __
$a (PubMed)25172254
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxu
100    1_
$a Bezrouk, Ales $u Assistant professor, Department of Medical Biophysics, Charles University, Prague, Faculty of Medicine, Hradec Kralove, Czech Republic. Electronic address: bezrouka@lfhk.cuni.cz.
245    10
$a Thermomechanical properties of nickel-titanium closed-coil springs and their implications for clinical practice / $c A. Bezrouk, L. Balsky, M. Smutny, I. Selke Krulichova, J. Zahora, J. Hanus, TR. Meling,
520    9_
$a INTRODUCTION: The aim was to study nickel-titanium closed-coil springs in a clinically relevant test setting with respect to the accuracy of the "preactivation" for nickel-titanium closed-coil springs application and whether it is possible to keep activation forces constant during the whole time of treatment. METHODS: We tested 10 types of springs from 5 manufacturers under clinically relevant conditions, allowing us to study the interactions between load and temperature over time. Hystereses were compared using t tests. RESULTS: Springs with a large mechanical hysteresis also showed a large thermal hysteresis. After heating shock, these springs showed intensive force spikes and persistent high loads. Some springs showed negligible thermal and mechanical hysteresis. Such springs never showed any clinically significant persistent high loads. CONCLUSIONS: Springs with a large hysteresis were unable to keep activation forces constant during the whole time of treatment even after any preactivation, and they might cause persistently high loads and possibly overloading. Only springs with minor hysteresis, low temperature dependence of force, and a clinically useful plateau have the following clinical advantages: reduced chair time, optimal rates of tooth movement, reproducible clinical results, and conservation of anchorage.
650    _2
$a nízká teplota $7 D003080
650    _2
$a zubní slitiny $x chemie $7 D003722
650    _2
$a analýza zatížení zubů $x přístrojové vybavení $7 D003799
650    _2
$a pružnost $7 D004548
650    _2
$a vysoká teplota $7 D006358
650    _2
$a lidé $7 D006801
650    _2
$a testování materiálů $7 D008422
650    _2
$a mechanické jevy $7 D055595
650    _2
$a nikl $x chemie $7 D009532
650    12
$a ortodontické dráty $7 D009969
650    _2
$a mechanický stres $7 D013314
650    _2
$a termodynamika $7 D013816
650    _2
$a titan $x chemie $7 D014025
650    _2
$a torze mechanická $7 D054159
655    _2
$a srovnávací studie $7 D003160
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Balský, Libor $u Private practice, Hradec Kralove, Czech Republic. $7 xx0324717
700    1_
$a Smutný, Martin, $u Postgraduate student, Department of Medical Biophysics, Charles University, Prague, Faculty of Medicine, Hradec Kralove, Czech Republic. $d 1989- $7 xx0300942
700    1_
$a Selke Krulichova, Iva $u Assistant professor, Department of Medical Biophysics, Charles University, Prague, Faculty of Medicine, Hradec Kralove, Czech Republic.
700    1_
$a Zahora, Jiri $u Assistant professor, Department of Medical Biophysics, Charles University, Prague, Faculty of Medicine, Hradec Kralove, Czech Republic.
700    1_
$a Hanus, Josef $u Associate professor and chair, Department of Medical Biophysics, Charles University, Prague, Faculty of Medicine, Hradec Kralove, Czech Republic.
700    1_
$a Meling, Torstein R $u Head, Department of Neurosurgery, National Hospital, Oslo, Norway.
773    0_
$w MED00000274 $t American journal of orthodontics and dentofacial orthopedics $x 0889-5406 $g Roč. 146, č. 3 (2014), s. 319-27
856    41
$u https://pubmed.ncbi.nlm.nih.gov/25172254 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20150306 $b ABA008
991    __
$a 20241120091951 $b ABA008
999    __
$a ok $b bmc $g 1065215 $s 890742
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2014 $b 146 $c 3 $d 319-27 $i 0889-5406 $m American journal of orthodontics and dentofacial orthopedics $n Am J Orthod Dentofacial Orthop $x MED00000274
LZP    __
$a Pubmed-20150306

Najít záznam

Citační ukazatele

Nahrávání dat ...

    Možnosti archivace