Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine
M. Kelbl, J. Kocis, R. Vesely, Z. Florian, T. Navrat, P. Vosynek,
Jazyk angličtina Země Jižní Korea
Typ dokumentu časopisecké články
NLK
Directory of Open Access Journals
od 2007
Free Medical Journals
od 2007
PubMed Central
od 2007
Europe PubMed Central
od 2007
ProQuest Central
od 2007-06-01 do 2022-08-31
Open Access Digital Library
od 2007-01-01
Health & Medicine (ProQuest)
od 2007-06-01 do 2022-08-31
ROAD: Directory of Open Access Scholarly Resources
od 2007
PubMed
26240706
DOI
10.4184/asj.2015.9.4.503
Knihovny.cz E-zdroje
- Publikační typ
- časopisecké články MeSH
STUDY DESIGN: An in vitro biomechanical study. PURPOSE: To evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant. OVERVIEW OF LITERATURE: Several types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomechanical properties of these implants have not been sufficiently studied yet. METHODS: A total of ten porcine specimens (levels Th12-L3) were prepared. The tests were performed for intact, injured, and implanted specimens. In each state, the specimen was subjected to a tension load of a prescribed force, and subsequently, twisted by a given angle. The force load was 200 N. The torsion load had a deformation character, i.e., the control variable was the twisting angle and the measured variable was the moment of a couple. The amplitude of the load alternating cycle was 3°. Another parameter that was evaluated was the area of the hysteresis loop. The area corresponds to the deformation energy which is dissipated during the cycle. RESULTS: A statistically significant difference was found between the intact and injured states as well as between the injured and implanted specimens. The statistical evaluation also showed a statistically different value of the hysteresis loop area. In the case of instability, the area decreased to 33% of the physiological value. For the implanted sample, the area increased to 170% of the physiological value. CONCLUSIONS: The Arcofix implant with its parameters appears to be suitable and sufficiently stable for the treatment of the anterior column of the spine.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc15031259
- 003
- CZ-PrNML
- 005
- 20210121145344.0
- 007
- ta
- 008
- 151005s2015 ko f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.4184/asj.2015.9.4.503 $2 doi
- 035 __
- $a (PubMed)26240706
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ko
- 100 1_
- $a Kelbl, Martin $u Department of Traumatology, Faculty of Medicine, Masaryk University, Brno, Czech Republic.
- 245 10
- $a Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine / $c M. Kelbl, J. Kocis, R. Vesely, Z. Florian, T. Navrat, P. Vosynek,
- 520 9_
- $a STUDY DESIGN: An in vitro biomechanical study. PURPOSE: To evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant. OVERVIEW OF LITERATURE: Several types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomechanical properties of these implants have not been sufficiently studied yet. METHODS: A total of ten porcine specimens (levels Th12-L3) were prepared. The tests were performed for intact, injured, and implanted specimens. In each state, the specimen was subjected to a tension load of a prescribed force, and subsequently, twisted by a given angle. The force load was 200 N. The torsion load had a deformation character, i.e., the control variable was the twisting angle and the measured variable was the moment of a couple. The amplitude of the load alternating cycle was 3°. Another parameter that was evaluated was the area of the hysteresis loop. The area corresponds to the deformation energy which is dissipated during the cycle. RESULTS: A statistically significant difference was found between the intact and injured states as well as between the injured and implanted specimens. The statistical evaluation also showed a statistically different value of the hysteresis loop area. In the case of instability, the area decreased to 33% of the physiological value. For the implanted sample, the area increased to 170% of the physiological value. CONCLUSIONS: The Arcofix implant with its parameters appears to be suitable and sufficiently stable for the treatment of the anterior column of the spine.
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Kocis, Jan $u Department of Traumatology, Faculty of Medicine, Masaryk University, Brno, Czech Republic.
- 700 1_
- $a Vesely, Radek $u Department of Traumatology, Faculty of Medicine, Masaryk University, Brno, Czech Republic.
- 700 1_
- $a Florian, Zdenek $u Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology, Brno, Czech Republic.
- 700 1_
- $a Navrat, Tomas $u Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology, Brno, Czech Republic.
- 700 1_
- $a Vosynek, Petr $u Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology, Brno, Czech Republic.
- 773 0_
- $w MED00172759 $t Asian spine journal $x 1976-1902 $g Roč. 9, č. 4 (2015), s. 503-10
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/26240706 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20151005 $b ABA008
- 991 __
- $a 20210121145343 $b ABA008
- 999 __
- $a ind $b bmc $g 1092140 $s 914383
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2015 $b 9 $c 4 $d 503-10 $e 20150728 $i 1976-1902 $m Asian spine journal $n Asian Spine J $x MED00172759
- LZP __
- $a Pubmed-20151005