Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine

M. Kelbl, J. Kocis, R. Vesely, Z. Florian, T. Navrat, P. Vosynek,

. 2015 ; 9 (4) : 503-10. [pub] 20150728

Jazyk angličtina Země Jižní Korea

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc15031259

STUDY DESIGN: An in vitro biomechanical study. PURPOSE: To evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant. OVERVIEW OF LITERATURE: Several types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomechanical properties of these implants have not been sufficiently studied yet. METHODS: A total of ten porcine specimens (levels Th12-L3) were prepared. The tests were performed for intact, injured, and implanted specimens. In each state, the specimen was subjected to a tension load of a prescribed force, and subsequently, twisted by a given angle. The force load was 200 N. The torsion load had a deformation character, i.e., the control variable was the twisting angle and the measured variable was the moment of a couple. The amplitude of the load alternating cycle was 3°. Another parameter that was evaluated was the area of the hysteresis loop. The area corresponds to the deformation energy which is dissipated during the cycle. RESULTS: A statistically significant difference was found between the intact and injured states as well as between the injured and implanted specimens. The statistical evaluation also showed a statistically different value of the hysteresis loop area. In the case of instability, the area decreased to 33% of the physiological value. For the implanted sample, the area increased to 170% of the physiological value. CONCLUSIONS: The Arcofix implant with its parameters appears to be suitable and sufficiently stable for the treatment of the anterior column of the spine.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc15031259
003      
CZ-PrNML
005      
20210121145344.0
007      
ta
008      
151005s2015 ko f 000 0|eng||
009      
AR
024    7_
$a 10.4184/asj.2015.9.4.503 $2 doi
035    __
$a (PubMed)26240706
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a ko
100    1_
$a Kelbl, Martin $u Department of Traumatology, Faculty of Medicine, Masaryk University, Brno, Czech Republic.
245    10
$a Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine / $c M. Kelbl, J. Kocis, R. Vesely, Z. Florian, T. Navrat, P. Vosynek,
520    9_
$a STUDY DESIGN: An in vitro biomechanical study. PURPOSE: To evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant. OVERVIEW OF LITERATURE: Several types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomechanical properties of these implants have not been sufficiently studied yet. METHODS: A total of ten porcine specimens (levels Th12-L3) were prepared. The tests were performed for intact, injured, and implanted specimens. In each state, the specimen was subjected to a tension load of a prescribed force, and subsequently, twisted by a given angle. The force load was 200 N. The torsion load had a deformation character, i.e., the control variable was the twisting angle and the measured variable was the moment of a couple. The amplitude of the load alternating cycle was 3°. Another parameter that was evaluated was the area of the hysteresis loop. The area corresponds to the deformation energy which is dissipated during the cycle. RESULTS: A statistically significant difference was found between the intact and injured states as well as between the injured and implanted specimens. The statistical evaluation also showed a statistically different value of the hysteresis loop area. In the case of instability, the area decreased to 33% of the physiological value. For the implanted sample, the area increased to 170% of the physiological value. CONCLUSIONS: The Arcofix implant with its parameters appears to be suitable and sufficiently stable for the treatment of the anterior column of the spine.
655    _2
$a časopisecké články $7 D016428
700    1_
$a Kocis, Jan $u Department of Traumatology, Faculty of Medicine, Masaryk University, Brno, Czech Republic.
700    1_
$a Vesely, Radek $u Department of Traumatology, Faculty of Medicine, Masaryk University, Brno, Czech Republic.
700    1_
$a Florian, Zdenek $u Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology, Brno, Czech Republic.
700    1_
$a Navrat, Tomas $u Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology, Brno, Czech Republic.
700    1_
$a Vosynek, Petr $u Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology, Brno, Czech Republic.
773    0_
$w MED00172759 $t Asian spine journal $x 1976-1902 $g Roč. 9, č. 4 (2015), s. 503-10
856    41
$u https://pubmed.ncbi.nlm.nih.gov/26240706 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20151005 $b ABA008
991    __
$a 20210121145343 $b ABA008
999    __
$a ind $b bmc $g 1092140 $s 914383
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2015 $b 9 $c 4 $d 503-10 $e 20150728 $i 1976-1902 $m Asian spine journal $n Asian Spine J $x MED00172759
LZP    __
$a Pubmed-20151005

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...