Spectral fluence of neutrons generated by radiotherapeutic linacs
M. Králík, J. Šolc, V. Vondráček, J. Šmoldasová, E. Farkašová, I. Tichá,
Language English Country England, Great Britain
Document type Comparative Study, Evaluation Study, Journal Article, Research Support, Non-U.S. Gov't
PubMed
24925899
DOI
10.1093/rpd/ncu192
Knihovny.cz E-resources
- MeSH
- Equipment Failure Analysis MeSH
- Particle Accelerators instrumentation MeSH
- Radiation Dosage MeSH
- Computer-Aided Design MeSH
- Equipment Design MeSH
- Monte Carlo Method MeSH
- Neutrons * MeSH
- Computer Simulation MeSH
- Scattering, Radiation MeSH
- Radiosurgery instrumentation MeSH
- Radiometry instrumentation methods MeSH
- Models, Statistical * MeSH
- Publication type
- Journal Article MeSH
- Evaluation Study MeSH
- Research Support, Non-U.S. Gov't MeSH
- Comparative Study MeSH
Spectral fluences of neutrons generated in the heads of the radiotherapeutic linacs Varian Clinac 2100 C/D and Siemens ARTISTE were measured by means of the Bonner spheres spectrometer whose active detector of thermal neutrons was replaced by an activation detector, i.e. a tablet made of pure manganese. Measurements with different collimator settings reveal an interesting dependence of neutron fluence on the area defined by the collimator jaws. The determined neutron spectral fluences were used to derive ambient dose equivalent rate along the treatment coach. To clarify at which components of the linac neutrons are mainly created, the measurements were complemented with MCNPX calculations based on a realistic model of the Varian Clinac.
Czech Metrology Institute Prague 10 CZ 102 00 Czech Republic
Proton Therapy Center Budínova 2 Prague 8 CZ 180 00 Czech Republic
Thomayer's Hospital Vídeňská 80 Prague 4 CZ 140 59 Czech Republic
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc16000732
- 003
- CZ-PrNML
- 005
- 20160127105112.0
- 007
- ta
- 008
- 160108s2015 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1093/rpd/ncu192 $2 doi
- 035 __
- $a (PubMed)24925899
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Králík, Miloslav $u Czech Metrology Institute, Prague 10 CZ-102 00, Czech Republic mkralik@cmi.cz.
- 245 10
- $a Spectral fluence of neutrons generated by radiotherapeutic linacs / $c M. Králík, J. Šolc, V. Vondráček, J. Šmoldasová, E. Farkašová, I. Tichá,
- 520 9_
- $a Spectral fluences of neutrons generated in the heads of the radiotherapeutic linacs Varian Clinac 2100 C/D and Siemens ARTISTE were measured by means of the Bonner spheres spectrometer whose active detector of thermal neutrons was replaced by an activation detector, i.e. a tablet made of pure manganese. Measurements with different collimator settings reveal an interesting dependence of neutron fluence on the area defined by the collimator jaws. The determined neutron spectral fluences were used to derive ambient dose equivalent rate along the treatment coach. To clarify at which components of the linac neutrons are mainly created, the measurements were complemented with MCNPX calculations based on a realistic model of the Varian Clinac.
- 650 _2
- $a počítačová simulace $7 D003198
- 650 _2
- $a design s pomocí počítače $7 D017076
- 650 _2
- $a design vybavení $7 D004867
- 650 _2
- $a analýza selhání vybavení $7 D019544
- 650 12
- $a statistické modely $7 D015233
- 650 _2
- $a metoda Monte Carlo $7 D009010
- 650 12
- $a neutrony $7 D009502
- 650 _2
- $a částice - urychlovače $x přístrojové vybavení $7 D010315
- 650 _2
- $a dávka záření $7 D011829
- 650 _2
- $a radiometrie $x přístrojové vybavení $x metody $7 D011874
- 650 _2
- $a radiochirurgie $x přístrojové vybavení $7 D016634
- 650 _2
- $a radiační rozptyl $7 D012542
- 655 _2
- $a srovnávací studie $7 D003160
- 655 _2
- $a hodnotící studie $7 D023362
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Šolc, Jaroslav $u Czech Metrology Institute, Prague 10 CZ-102 00, Czech Republic.
- 700 1_
- $a Vondráček, Vladimir $u Proton Therapy Center, Budínova 2, Prague 8 CZ-180 00, Czech Republic.
- 700 1_
- $a Šmoldasová, Jana $u Czech Metrology Institute, Prague 10 CZ-102 00, Czech Republic.
- 700 1_
- $a Farkašová, Estera $u Thomayer's Hospital, Vídeňská 80, Prague 4 CZ-140 59, Czech Republic.
- 700 1_
- $a Tichá, Ivana $u Thomayer's Hospital, Vídeňská 80, Prague 4 CZ-140 59, Czech Republic.
- 773 0_
- $w MED00004029 $t Radiation protection dosimetry $x 1742-3406 $g Roč. 163, č. 3 (2015), s. 373-80
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/24925899 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20160108 $b ABA008
- 991 __
- $a 20160127105236 $b ABA008
- 999 __
- $a ok $b bmc $g 1103013 $s 924938
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2015 $b 163 $c 3 $d 373-80 $e 20140612 $i 1742-3406 $m Radiation protection dosimetry $n Radiat Prot Dosimetry $x MED00004029
- LZP __
- $a Pubmed-20160108