Narrow-bore core-shell particles and monolithic columns in the analysis of silybin diastereoisomers
P. Marhol, R. Gažák, P. Bednář, V. Křen,
Jazyk angličtina Země Německo
Typ dokumentu hodnotící studie, časopisecké články, práce podpořená grantem
PubMed
21710694
DOI
10.1002/jssc.201100309
Knihovny.cz E-zdroje
- MeSH
- silymarin chemie izolace a purifikace MeSH
- stereoizomerie MeSH
- syntetické pryskyřice chemie MeSH
- vysokoúčinná kapalinová chromatografie přístrojové vybavení metody MeSH
- Publikační typ
- časopisecké články MeSH
- hodnotící studie MeSH
- práce podpořená grantem MeSH
Two chromatographic narrow-bore columns, a novel 2.6 μm particle-packed Kinetex™ C18 core-shell (50×2.1 mm id) and monolithic Chromolith(®) FastGradient RP-18e (50×2 mm id), were evaluated for the analysis of diastereoisomers of the flavonolignans silybin and 23-O-acetylsilybin under isocratic conditions. The main advantages of the core-shell column are markedly higher efficiency (hmin =2.8 versus 5.6 for silybin A) and better peak symmetry. The Kinetex column exhibits only a slight change in the height equivalent of the theoretical plate with a higher linear velocity of the mobile phase. The monolithic column shows notably higher selectivity in terms of selectivity factor (1.21 versus 1.12) in the analysis of critical-pair of diastereoisomers (silybin A and silybin B) and enables shorter run duration (approx. twofold) together with lower backpressure. The resolution power was found to be comparable, but the Kinetex column required a higher pressure of the mobile phase that, together with the higher chance of clogging, can be a disadvantage in the separation of biological samples. Successful baseline separation of silybin diastereoisomers in real pharmaceutical sample on monolithic column was accomplished.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc16000831
- 003
- CZ-PrNML
- 005
- 20160126104038.0
- 007
- ta
- 008
- 160108s2011 gw f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1002/jssc.201100309 $2 doi
- 035 __
- $a (PubMed)21710694
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a gw
- 100 1_
- $a Marhol, Petr $u Institute of Microbiology, Centre of Biocatalysis and Biotransformation, Academy of Sciences of the Czech Republic, Prague, Czech Republic. marholp@gmail.com.
- 245 10
- $a Narrow-bore core-shell particles and monolithic columns in the analysis of silybin diastereoisomers / $c P. Marhol, R. Gažák, P. Bednář, V. Křen,
- 520 9_
- $a Two chromatographic narrow-bore columns, a novel 2.6 μm particle-packed Kinetex™ C18 core-shell (50×2.1 mm id) and monolithic Chromolith(®) FastGradient RP-18e (50×2 mm id), were evaluated for the analysis of diastereoisomers of the flavonolignans silybin and 23-O-acetylsilybin under isocratic conditions. The main advantages of the core-shell column are markedly higher efficiency (hmin =2.8 versus 5.6 for silybin A) and better peak symmetry. The Kinetex column exhibits only a slight change in the height equivalent of the theoretical plate with a higher linear velocity of the mobile phase. The monolithic column shows notably higher selectivity in terms of selectivity factor (1.21 versus 1.12) in the analysis of critical-pair of diastereoisomers (silybin A and silybin B) and enables shorter run duration (approx. twofold) together with lower backpressure. The resolution power was found to be comparable, but the Kinetex column required a higher pressure of the mobile phase that, together with the higher chance of clogging, can be a disadvantage in the separation of biological samples. Successful baseline separation of silybin diastereoisomers in real pharmaceutical sample on monolithic column was accomplished.
- 650 _2
- $a vysokoúčinná kapalinová chromatografie $x přístrojové vybavení $x metody $7 D002851
- 650 _2
- $a syntetické pryskyřice $x chemie $7 D012117
- 650 _2
- $a silymarin $x chemie $x izolace a purifikace $7 D012838
- 650 _2
- $a stereoizomerie $7 D013237
- 655 _2
- $a hodnotící studie $7 D023362
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Gažák, Radek
- 700 1_
- $a Bednář, Petr
- 700 1_
- $a Křen, Vladimír
- 773 0_
- $w MED00006463 $t Journal of separation science $x 1615-9314 $g Roč. 34, č. 16-17 (2011), s. 2206-13
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/21710694 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20160108 $b ABA008
- 991 __
- $a 20160126104201 $b ABA008
- 999 __
- $a ok $b bmc $g 1103112 $s 925037
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2011 $b 34 $c 16-17 $d 2206-13 $e 20110627 $i 1615-9314 $m Journal of separation science $n J Sep Sci $x MED00006463
- LZP __
- $a Pubmed-20160108