Je něco špatně v tomto záznamu ?
Prediction of wastewater quality using amperometric bioelectronic tongues
I. Czolkos, E. Dock, E. Tønning, J. Christensen, M. Winther-Nielsen, C. Carlsson, R. Mojzíková, P. Skládal, U. Wollenberger, L. Nørgaard, T. Ruzgas, J. Emnéus,
Jazyk angličtina Země Anglie, Velká Británie
Typ dokumentu časopisecké články, práce podpořená grantem
- MeSH
- biosenzitivní techniky * MeSH
- chemické látky znečišťující vodu chemie izolace a purifikace MeSH
- monitorování životního prostředí * MeSH
- odpadní voda chemie MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
Wastewater samples from a Swedish chemi-thermo-mechanical pulp (CTMP) mill collected at different purification stages in a wastewater treatment plant (WWTP) were analyzed with an amperometric enzyme-based biosensor array in a flow-injection system. In order to resolve the complex composition of the wastewater, the array consists of several sensing elements which yield a multidimensional response. We used principal component analysis (PCA) to decompose the array's responses, and found that wastewater with different degrees of pollution can be differentiated. With the help of partial least squares regression (PLS-R), we could link the sensor responses to the Microtox® toxicity parameter, as well as to global organic pollution parameters (COD, BOD, and TOC). From investigating the influences of individual sensors in the array, it was found that the best models were in most cases obtained when all sensors in the array were included in the PLS-R model. We find that fast simultaneous determination of several global environmental parameters characterizing wastewaters is possible with this kind of biosensor array, in particular because of the link between the sensor responses and the biological effect onto the ecosystem into which the wastewater would be released. In conjunction with multivariate data analysis tools, there is strong potential to reduce the total time until a result is yielded from days to a few minutes.
AkzoNobel Industrial Coatings AB Staffanstorpsvägen 50 205 17 Malmö Sweden
AnoxKaldnes AB Klosterängsvägen 11A 226 47 Lund Sweden
Fiberline Composites A S Barmstedt Allé 5 5500 Middelfart Denmark
Institute of Biochemistry Masaryk University Kotlářská 2 611 37 Brno Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc16020319
- 003
- CZ-PrNML
- 005
- 20160726092833.0
- 007
- ta
- 008
- 160722s2016 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.bios.2015.08.055 $2 doi
- 024 7_
- $a 10.1016/j.bios.2015.08.055 $2 doi
- 035 __
- $a (PubMed)26342573
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Czolkos, Ilja $u DTU Nanotech - Department of Micro and Nanotechnology, Technical University of Denmark, Ørsteds Plads, 2800 Kgs. Lyngby, Denmark. Electronic address: ilcz@nanotech.dtu.dk.
- 245 10
- $a Prediction of wastewater quality using amperometric bioelectronic tongues / $c I. Czolkos, E. Dock, E. Tønning, J. Christensen, M. Winther-Nielsen, C. Carlsson, R. Mojzíková, P. Skládal, U. Wollenberger, L. Nørgaard, T. Ruzgas, J. Emnéus,
- 520 9_
- $a Wastewater samples from a Swedish chemi-thermo-mechanical pulp (CTMP) mill collected at different purification stages in a wastewater treatment plant (WWTP) were analyzed with an amperometric enzyme-based biosensor array in a flow-injection system. In order to resolve the complex composition of the wastewater, the array consists of several sensing elements which yield a multidimensional response. We used principal component analysis (PCA) to decompose the array's responses, and found that wastewater with different degrees of pollution can be differentiated. With the help of partial least squares regression (PLS-R), we could link the sensor responses to the Microtox® toxicity parameter, as well as to global organic pollution parameters (COD, BOD, and TOC). From investigating the influences of individual sensors in the array, it was found that the best models were in most cases obtained when all sensors in the array were included in the PLS-R model. We find that fast simultaneous determination of several global environmental parameters characterizing wastewaters is possible with this kind of biosensor array, in particular because of the link between the sensor responses and the biological effect onto the ecosystem into which the wastewater would be released. In conjunction with multivariate data analysis tools, there is strong potential to reduce the total time until a result is yielded from days to a few minutes.
- 650 12
- $a biosenzitivní techniky $7 D015374
- 650 12
- $a monitorování životního prostředí $7 D004784
- 650 _2
- $a odpadní voda $x chemie $7 D062065
- 650 _2
- $a chemické látky znečišťující vodu $x chemie $x izolace a purifikace $7 D014874
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Dock, Eva $u AkzoNobel Industrial Coatings AB, Staffanstorpsvägen 50, 205 17 Malmö, Sweden.
- 700 1_
- $a Tønning, Erik $u Fiberline Composites A/S, Barmstedt Allé 5, 5500 Middelfart, Denmark.
- 700 1_
- $a Christensen, Jakob $u Umetrics AB, Stortorget 21, 211 34 Malmö, Sweden.
- 700 1_
- $a Winther-Nielsen, Margrethe $u Department of Environmental Risk Assessment, DHI Water and Environment, Agern Allé 5, 2970 Hørsholm, Denmark.
- 700 1_
- $a Carlsson, Charlotte $u AnoxKaldnes AB, Klosterängsvägen 11A, 226 47 Lund, Sweden.
- 700 1_
- $a Mojzíková, Renata $u Department of Biology, Faculty of Dentistry and Medicine, Palacký University Olumouc, Hněvotínská 3, 775 15 Olomouc, Czech Republic.
- 700 1_
- $a Skládal, Petr $u Institute of Biochemistry, Masaryk University, Kotlářská 2, 611 37 Brno, Czech Republic.
- 700 1_
- $a Wollenberger, Ulla $u Analytical Biochemistry, Institute of Biochemistry and Biology, University of Potsdam, 14476 Golm, Germany.
- 700 1_
- $a Nørgaard, Lars $u Department of Food Science, Faculty of Science, University of Copenhagen, Rolighedsvej 30, 1958 Frederiksberg, Denmark.
- 700 1_
- $a Ruzgas, Tautgirdas $u Department of Biomedical Science, Faculty of Health and Society, Malmö University, Jan Waldenströms gata 25, 205 06 Malmö, Sweden.
- 700 1_
- $a Emnéus, Jenny $u DTU Nanotech - Department of Micro and Nanotechnology, Technical University of Denmark, Ørsteds Plads, 2800 Kgs. Lyngby, Denmark. Electronic address: ilcz@nanotech.dtu.dk.
- 773 0_
- $w MED00006627 $t Biosensors & bioelectronics $x 1873-4235 $g Roč. 75, č. - (2016), s. 375-82
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/26342573 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20160722 $b ABA008
- 991 __
- $a 20160726093053 $b ABA008
- 999 __
- $a ok $b bmc $g 1154989 $s 944847
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2016 $b 75 $c - $d 375-82 $i 1873-4235 $m Biosensors & bioelectronics $n Biosens Bioelectron $x MED00006627
- LZP __
- $a Pubmed-20160722