2014 ICHLNRRA intercomparison of radon/thoron gas and radon short-lived decay products measuring instruments in the NRPI Prague

K. Jílek, J. Timková,

. 2015 ; 164 (4) : 556-62. [pub] 20150519

Jazyk angličtina Země Anglie, Velká Británie

Typ dokumentu srovnávací studie, hodnotící studie, časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/bmc16020776

During the Eighth International Conference on High Levels of Natural Radiation and Radon Areas held in autumn 2014 at Prague, the third intercomparison of radon/thoron gas and radon short-lived decay products measurement instruments was organised by and held at the Natural Radiation Division of the National Radiation Protection Institute (NRPI; SÚRO v.v.i.) in Prague. The intercomparison was newly focussed also on continuous monitors with active sampling adapters capable to distinguish radon/thoron gas in their mix field.The results of radon gas measurements carried out in the big NRPI radon chamber indicated very well an average deviation of up to 5 % from the reference NRPI value for 80 % of all the exposed instruments. The results of equilibrium equivalent concentration continuous monitors indicated an average deviation of up to 5 % from the reference NRPI value for 40 % of all the exposed instruments and their ~8-10 % shift compared with the NRPI. The results of investigated ambient conditions upon response of exposed continuous monitors indicated influence of aerosol changes upon response of radon monitors with an active air sampling adapters through the filter, only. The exposures of both radon/thoron gas discriminative continuous monitors and passive detectors have been indicated inconsistent results: on one hand, their excellent agreement up to several per cent for both the gases, and on the other hand, systematic unsatisfactory differences up to 40 %. Additional radon/thoron exercises are recommended to improve both the instruments themselves and quality of their operators.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc16020776
003      
CZ-PrNML
005      
20160722121411.0
007      
ta
008      
160722s2015 enk f 000 0|eng||
009      
AR
024    7_
$a 10.1093/rpd/ncv311 $2 doi
024    7_
$a 10.1093/rpd/ncv311 $2 doi
035    __
$a (PubMed)25990114
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Jílek, K $u National Radiation Protection Institute, Prague, the Czech Republic karel.jilek@suro.cz.
245    10
$a 2014 ICHLNRRA intercomparison of radon/thoron gas and radon short-lived decay products measuring instruments in the NRPI Prague / $c K. Jílek, J. Timková,
520    9_
$a During the Eighth International Conference on High Levels of Natural Radiation and Radon Areas held in autumn 2014 at Prague, the third intercomparison of radon/thoron gas and radon short-lived decay products measurement instruments was organised by and held at the Natural Radiation Division of the National Radiation Protection Institute (NRPI; SÚRO v.v.i.) in Prague. The intercomparison was newly focussed also on continuous monitors with active sampling adapters capable to distinguish radon/thoron gas in their mix field.The results of radon gas measurements carried out in the big NRPI radon chamber indicated very well an average deviation of up to 5 % from the reference NRPI value for 80 % of all the exposed instruments. The results of equilibrium equivalent concentration continuous monitors indicated an average deviation of up to 5 % from the reference NRPI value for 40 % of all the exposed instruments and their ~8-10 % shift compared with the NRPI. The results of investigated ambient conditions upon response of exposed continuous monitors indicated influence of aerosol changes upon response of radon monitors with an active air sampling adapters through the filter, only. The exposures of both radon/thoron gas discriminative continuous monitors and passive detectors have been indicated inconsistent results: on one hand, their excellent agreement up to several per cent for both the gases, and on the other hand, systematic unsatisfactory differences up to 40 %. Additional radon/thoron exercises are recommended to improve both the instruments themselves and quality of their operators.
650    _2
$a radioaktivní látky znečišťující vzduch $x analýza $7 D000396
650    12
$a radiační pozadí $7 D001417
650    _2
$a design vybavení $7 D004867
650    _2
$a plyny $x analýza $7 D005740
650    12
$a testování odbornosti laboratoří $7 D059021
650    _2
$a řízení kvality $7 D011786
650    _2
$a monitorování radiace $x přístrojové vybavení $x metody $x normy $7 D011834
650    _2
$a radiometrie $x přístrojové vybavení $x metody $7 D011874
650    _2
$a radon $x analýza $7 D011886
650    _2
$a dceřiné produkty radonu $x analýza $7 D017687
650    _2
$a reprodukovatelnost výsledků $7 D015203
651    _2
$a Alžírsko $7 D000462
651    _2
$a Evropa $7 D005060
651    _2
$a Japonsko $7 D007564
651    _2
$a Spojené státy americké $7 D014481
655    _2
$a srovnávací studie $7 D003160
655    _2
$a hodnotící studie $7 D023362
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Timková, J $u National Radiation Protection Institute, Prague, the Czech Republic.
773    0_
$w MED00004029 $t Radiation protection dosimetry $x 1742-3406 $g Roč. 164, č. 4 (2015), s. 556-62
856    41
$u https://pubmed.ncbi.nlm.nih.gov/25990114 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20160722 $b ABA008
991    __
$a 20160722121625 $b ABA008
999    __
$a ok $b bmc $g 1155446 $s 945304
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2015 $b 164 $c 4 $d 556-62 $e 20150519 $i 1742-3406 $m Radiation protection dosimetry $n Radiat Prot Dosimetry $x MED00004029
LZP    __
$a Pubmed-20160722

Najít záznam

Citační ukazatele

Pouze přihlášení uživatelé

Možnosti archivace

Nahrávání dat ...