Something wrong with this record ?
Semi-automatic motion compensation of contrast-enhanced ultrasound images from abdominal organs for perfusion analysis
S. Schäfer, K. Nylund, F. Sævik, T. Engjom, M. Mézl, R. Jiřík, G. Dimcevski, OH. Gilja, K. Tönnies,
Language English Country United States
Document type Journal Article, Research Support, Non-U.S. Gov't
NLK
ProQuest Central
from 2003-01-01 to 2023-12-31
Medline Complete (EBSCOhost)
from 2012-09-01 to 2015-07-31
Nursing & Allied Health Database (ProQuest)
from 2003-01-01 to 2023-12-31
Health & Medicine (ProQuest)
from 2003-01-01 to 2023-12-31
- MeSH
- Abdomen ultrasonography MeSH
- Crohn Disease ultrasonography MeSH
- Cystic Fibrosis ultrasonography MeSH
- Databases, Factual * MeSH
- Image Interpretation, Computer-Assisted methods MeSH
- Contrast Media administration & dosage MeSH
- Humans MeSH
- Motion MeSH
- Check Tag
- Humans MeSH
- Male MeSH
- Female MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
This paper presents a system for correcting motion influences in time-dependent 2D contrast-enhanced ultrasound (CEUS) images to assess tissue perfusion characteristics. The system consists of a semi-automatic frame selection method to find images with out-of-plane motion as well as a method for automatic motion compensation. Translational and non-rigid motion compensation is applied by introducing a temporal continuity assumption. A study consisting of 40 clinical datasets was conducted to compare the perfusion with simulated perfusion using pharmacokinetic modeling. Overall, the proposed approach decreased the mean average difference between the measured perfusion and the pharmacokinetic model estimation. It was non-inferior for three out of four patient cohorts to a manual approach and reduced the analysis time by 41% compared to manual processing.
Department of Biomedical Engineering Brno University of Technology Czech Republic
Department of Clinical Medicine University of Bergen Norway
National Centre for Ultrasound in Gastroenterology Haukeland University Hospital Bergen Norway
Section of Gastroenterology Department of Medicine Haukeland University Hospital Bergen Norway
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc16021029
- 003
- CZ-PrNML
- 005
- 20250507155257.0
- 007
- ta
- 008
- 160722e20141017xxu f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.compbiomed.2014.09.014 $2 doi
- 024 7_
- $a 10.1016/j.compbiomed.2014.09.014 $2 doi
- 035 __
- $a (PubMed)25453358
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxu
- 100 1_
- $a Schäfer, Sebastian $u Department of Simulation and Graphics, University of Magdeburg, Universitätsplatz 2, 39106 Magdeburg, Germany. Electronic address: sebastian.schaefer@ovgu.de.
- 245 10
- $a Semi-automatic motion compensation of contrast-enhanced ultrasound images from abdominal organs for perfusion analysis / $c S. Schäfer, K. Nylund, F. Sævik, T. Engjom, M. Mézl, R. Jiřík, G. Dimcevski, OH. Gilja, K. Tönnies,
- 520 9_
- $a This paper presents a system for correcting motion influences in time-dependent 2D contrast-enhanced ultrasound (CEUS) images to assess tissue perfusion characteristics. The system consists of a semi-automatic frame selection method to find images with out-of-plane motion as well as a method for automatic motion compensation. Translational and non-rigid motion compensation is applied by introducing a temporal continuity assumption. A study consisting of 40 clinical datasets was conducted to compare the perfusion with simulated perfusion using pharmacokinetic modeling. Overall, the proposed approach decreased the mean average difference between the measured perfusion and the pharmacokinetic model estimation. It was non-inferior for three out of four patient cohorts to a manual approach and reduced the analysis time by 41% compared to manual processing.
- 650 _2
- $a břicho $x ultrasonografie $7 D000005
- 650 _2
- $a kontrastní látky $x aplikace a dávkování $7 D003287
- 650 _2
- $a Crohnova nemoc $x ultrasonografie $7 D003424
- 650 _2
- $a cystická fibróza $x ultrasonografie $7 D003550
- 650 12
- $a databáze faktografické $7 D016208
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a interpretace obrazu počítačem $x metody $7 D007090
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 _2
- $a pohyb těles $7 D009038
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Nylund, Kim $u National Centre for Ultrasound in Gastroenterology, Haukeland University Hospital, Bergen, Norway.
- 700 1_
- $a Sævik, Fredrik $u Department of Clinical Medicine, University of Bergen, Norway; National Centre for Ultrasound in Gastroenterology, Haukeland University Hospital, Bergen, Norway.
- 700 1_
- $a Engjom, Trond $u Section of Gastroenterology, Department of Medicine, Haukeland University Hospital, Bergen, Norway; National Centre for Ultrasound in Gastroenterology, Haukeland University Hospital, Bergen, Norway.
- 700 1_
- $a Mézl, Martin $u Department of Biomedical Engineering, Brno University of Technology, Czech Republic. $7 xx0331830
- 700 1_
- $a Jiřík, Radovan $u Institute of Scientific Instruments of the Academy of Sciences of the Czech Republic, Brno, Czech Republic; International Clinical Research Center - Center of Biomedical Engineering, St. Anne׳s University Hospital Brno, Czech Republic.
- 700 1_
- $a Dimcevski, Georg $u Section of Gastroenterology, Department of Medicine, Haukeland University Hospital, Bergen, Norway; National Centre for Ultrasound in Gastroenterology, Haukeland University Hospital, Bergen, Norway.
- 700 1_
- $a Gilja, Odd Helge $u National Centre for Ultrasound in Gastroenterology, Haukeland University Hospital, Bergen, Norway; Department of Clinical Medicine, University of Bergen, Norway.
- 700 1_
- $a Tönnies, Klaus $u Department of Simulation and Graphics, University of Magdeburg, Universitätsplatz 2, 39106 Magdeburg, Germany.
- 773 0_
- $w MED00001218 $t Computers in biology and medicine $x 1879-0534 $g Roč. 63 (20141017), s. 229-37
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/25453358 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20160722 $b ABA008
- 991 __
- $a 20250507155255 $b ABA008
- 999 __
- $a ok $b bmc $g 1155699 $s 945557
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2015 $b 63 $c - $d 229-37 $e 20141017 $i 1879-0534 $m Computers in biology and medicine $n Comput Biol Med $x MED00001218
- LZP __
- $a Pubmed-20160722