Stanovení součinitele tření hlavic a jamek TEP kyčle [Determination of a friction coefficient for THA bearing couples]
M. Vrbka, D. Nečas, J. Bartošík, M. Hartl, I. Křupka, A. Galandáková, J. Gallo
Language Czech Country Czech Republic
Document type English Abstract, Journal Article
Grant support
NT14267
MZ0
CEP Register
PubMed
26516952
- MeSH
- Ceramics MeSH
- Hip Prosthesis * MeSH
- Humans MeSH
- Arthroplasty, Replacement, Hip instrumentation MeSH
- Surface Properties MeSH
- Prosthesis Design MeSH
- Prosthesis Failure MeSH
- Materials Testing methods MeSH
- Friction MeSH
- Check Tag
- Humans MeSH
- Publication type
- English Abstract MeSH
- Journal Article MeSH
PURPOSE OF THE STUDY: The wear of articular surfaces is considered one of the most important factors limiting the life of total hip arthroplasty (THA). It is assumed that the particles released from the surface of a softer material induce a complex inflammatory response, which will eventually result in osteolysis and aseptic loosening. Implant wear is related to a friction coefficient which depends on combination of the materials used, roughness of the articulating surfaces, internal clearance, and dimensions of the prosthesis. MATERIAL AND METHODS: The selected parameters of the bearing couples tested were studied using an experimental device based on the principle of a pendulum. Bovine serum was used as a lubricant and the load corresponded to a human body mass of 75 kg. The friction coefficient was derived from a curve of slowdown of pendulum oscillations. Roughness was measured with a device working on the principle of interferometry. Clearance was assessed by measuring diameters of the acetabular and femoral heads with a 3D optical scanner. The specimens tested included unused metal-on-highly cross-linked polyethylene, ceramic-on-highly cross-linked polyethylene and ceramic-on-ceramic bearing couples with the diameters of 28 mm and 36 mm. For each measured parameter, an arithmetic mean was calculated from 10 measurements. RESULTS: 1) The roughness of polyethylene surfaces was higher by about one order of magnitude than the roughness of metal and ceramic components. The Protasul metal head had the least rough surface (0.003 μm). 2) The ceramic-on-ceramic couples had the lowest clearance. Bearing couples with polyethylene acetabular liners had markedly higher clearances ranging from 150 μm to 545 μm. A clearance increased with large femoral heads (up to 4-fold in one of the couple tested). 3) The friction coefficient was related to the combination of materials; it was lowest in ceramic-on-ceramic surfaces (0.11 to 0.12) and then in ceramic-on-polyethylene implants (0.13 to 0.14). The friction coefficient is supposed to increase with a decreasing femoral head diameter. However, in the bearing couples with polyethylene liners manufactured by one company, paradoxically, the friction coefficient slightly increased with an increase in femoral head size from 28 mm to 36 mm. 4) The lowest friction moment (< 3.5 Nm) was found for ceramic-on-ceramic implants 28 mm in diameter; the highest values were recorded in metal-on-polyethylene bearing couples 36 mm in diameter (> 7 Nm). DISCUSSION: Although our study confirmed that the bearing couples produced by different manufacturers varied to some extent in the parameters studied, in our opinion, this variability was not significant because it was not within an order of magnitude in any of the tests. CONCLUSIONS: The study showed that both the friction coefficient and the friction moment are affected more by the combination of materials than by the diameter of a femoral head. The best results were achieved in ceramic-on-ceramic implants.
Ortopedická klinika LF UP a FN Olomouc Olomouc
Ústav konstruování Fakulta strojního inženýrství VUT v Brně Brno
Determination of a friction coefficient for THA bearing couples
- 000
- 00000naa a2200000 a 4500
- 001
- bmc16022576
- 003
- CZ-PrNML
- 005
- 20190708091009.0
- 007
- ta
- 008
- 160811s2015 xr f 000 0|cze||
- 009
- AR
- 024 7_
- $2 doi $a 10.55095/achot2015/057
- 035 __
- $a (PubMed)26516952
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a cze
- 044 __
- $a xr
- 100 1_
- $a Vrbka, Miloš $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno $7 xx0117094
- 245 10
- $a Stanovení součinitele tření hlavic a jamek TEP kyčle / $c M. Vrbka, D. Nečas, J. Bartošík, M. Hartl, I. Křupka, A. Galandáková, J. Gallo
- 246 31
- $a Determination of a friction coefficient for THA bearing couples
- 520 9_
- $a PURPOSE OF THE STUDY: The wear of articular surfaces is considered one of the most important factors limiting the life of total hip arthroplasty (THA). It is assumed that the particles released from the surface of a softer material induce a complex inflammatory response, which will eventually result in osteolysis and aseptic loosening. Implant wear is related to a friction coefficient which depends on combination of the materials used, roughness of the articulating surfaces, internal clearance, and dimensions of the prosthesis. MATERIAL AND METHODS: The selected parameters of the bearing couples tested were studied using an experimental device based on the principle of a pendulum. Bovine serum was used as a lubricant and the load corresponded to a human body mass of 75 kg. The friction coefficient was derived from a curve of slowdown of pendulum oscillations. Roughness was measured with a device working on the principle of interferometry. Clearance was assessed by measuring diameters of the acetabular and femoral heads with a 3D optical scanner. The specimens tested included unused metal-on-highly cross-linked polyethylene, ceramic-on-highly cross-linked polyethylene and ceramic-on-ceramic bearing couples with the diameters of 28 mm and 36 mm. For each measured parameter, an arithmetic mean was calculated from 10 measurements. RESULTS: 1) The roughness of polyethylene surfaces was higher by about one order of magnitude than the roughness of metal and ceramic components. The Protasul metal head had the least rough surface (0.003 μm). 2) The ceramic-on-ceramic couples had the lowest clearance. Bearing couples with polyethylene acetabular liners had markedly higher clearances ranging from 150 μm to 545 μm. A clearance increased with large femoral heads (up to 4-fold in one of the couple tested). 3) The friction coefficient was related to the combination of materials; it was lowest in ceramic-on-ceramic surfaces (0.11 to 0.12) and then in ceramic-on-polyethylene implants (0.13 to 0.14). The friction coefficient is supposed to increase with a decreasing femoral head diameter. However, in the bearing couples with polyethylene liners manufactured by one company, paradoxically, the friction coefficient slightly increased with an increase in femoral head size from 28 mm to 36 mm. 4) The lowest friction moment (< 3.5 Nm) was found for ceramic-on-ceramic implants 28 mm in diameter; the highest values were recorded in metal-on-polyethylene bearing couples 36 mm in diameter (> 7 Nm). DISCUSSION: Although our study confirmed that the bearing couples produced by different manufacturers varied to some extent in the parameters studied, in our opinion, this variability was not significant because it was not within an order of magnitude in any of the tests. CONCLUSIONS: The study showed that both the friction coefficient and the friction moment are affected more by the combination of materials than by the diameter of a femoral head. The best results were achieved in ceramic-on-ceramic implants.
- 650 _2
- $a náhrada kyčelního kloubu $x přístrojové vybavení $7 D019644
- 650 _2
- $a keramika $7 D002516
- 650 _2
- $a tření $7 D017276
- 650 12
- $a kyčelní protézy $7 D006622
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a testování materiálů $x metody $7 D008422
- 650 _2
- $a protézy - design $7 D011474
- 650 _2
- $a selhání protézy $7 D011475
- 650 _2
- $a povrchové vlastnosti $7 D013499
- 655 _2
- $a anglický abstrakt $7 D004740
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Nečas, David $7 xx0209867 $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno
- 700 1_
- $a Bartošík, J. $7 _AN088222 $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno
- 700 1_
- $a Hartl, Martin, $7 mzk2002157052 $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno $d 1967-
- 700 1_
- $a Křupka, Ivan, $7 mzk2002157050 $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno $d 1967-
- 700 1_
- $a Galandáková, Adéla, $7 stk2008428623 $u Ústav lékařské chemie a biochemie LF UP v Olomouci, Olomouc $d 1979-
- 700 1_
- $a Gallo, Jiří $7 xx0019005 $u Ortopedická klinika LF UP a FN Olomouc, Olomouc
- 773 0_
- $w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $x 0001-5415 $g Roč. 82, č. 5 (2015), s. 341-347
- 910 __
- $a ABA008 $b A 8 $c 507 $y 4 $z 0
- 990 __
- $a 20160811 $b ABA008
- 991 __
- $a 20190708091206 $b ABA008
- 999 __
- $a ok $b bmc $g 1158903 $s 947121
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2015 $b 82 $c 5 $d 341-347 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
- GRA __
- $a NT14267 $p MZ0
- LZP __
- $b NLK118 $a Pubmed-20160811