Stanovení součinitele tření hlavic a jamek TEP kyčle [Determination of a friction coefficient for THA bearing couples]

M. Vrbka, D. Nečas, J. Bartošík, M. Hartl, I. Křupka, A. Galandáková, J. Gallo

. 2015 ; 82 (5) : 341-347.

Language Czech Country Czech Republic

Document type English Abstract, Journal Article

Grant support
NT14267 MZ0 CEP Register

Digital library NLK
Source

E-resources Online

NLK Free Medical Journals from 2006

Links

PubMed 26516952

PURPOSE OF THE STUDY: The wear of articular surfaces is considered one of the most important factors limiting the life of total hip arthroplasty (THA). It is assumed that the particles released from the surface of a softer material induce a complex inflammatory response, which will eventually result in osteolysis and aseptic loosening. Implant wear is related to a friction coefficient which depends on combination of the materials used, roughness of the articulating surfaces, internal clearance, and dimensions of the prosthesis. MATERIAL AND METHODS: The selected parameters of the bearing couples tested were studied using an experimental device based on the principle of a pendulum. Bovine serum was used as a lubricant and the load corresponded to a human body mass of 75 kg. The friction coefficient was derived from a curve of slowdown of pendulum oscillations. Roughness was measured with a device working on the principle of interferometry. Clearance was assessed by measuring diameters of the acetabular and femoral heads with a 3D optical scanner. The specimens tested included unused metal-on-highly cross-linked polyethylene, ceramic-on-highly cross-linked polyethylene and ceramic-on-ceramic bearing couples with the diameters of 28 mm and 36 mm. For each measured parameter, an arithmetic mean was calculated from 10 measurements. RESULTS: 1) The roughness of polyethylene surfaces was higher by about one order of magnitude than the roughness of metal and ceramic components. The Protasul metal head had the least rough surface (0.003 μm). 2) The ceramic-on-ceramic couples had the lowest clearance. Bearing couples with polyethylene acetabular liners had markedly higher clearances ranging from 150 μm to 545 μm. A clearance increased with large femoral heads (up to 4-fold in one of the couple tested). 3) The friction coefficient was related to the combination of materials; it was lowest in ceramic-on-ceramic surfaces (0.11 to 0.12) and then in ceramic-on-polyethylene implants (0.13 to 0.14). The friction coefficient is supposed to increase with a decreasing femoral head diameter. However, in the bearing couples with polyethylene liners manufactured by one company, paradoxically, the friction coefficient slightly increased with an increase in femoral head size from 28 mm to 36 mm. 4) The lowest friction moment (< 3.5 Nm) was found for ceramic-on-ceramic implants 28 mm in diameter; the highest values were recorded in metal-on-polyethylene bearing couples 36 mm in diameter (> 7 Nm). DISCUSSION: Although our study confirmed that the bearing couples produced by different manufacturers varied to some extent in the parameters studied, in our opinion, this variability was not significant because it was not within an order of magnitude in any of the tests. CONCLUSIONS: The study showed that both the friction coefficient and the friction moment are affected more by the combination of materials than by the diameter of a femoral head. The best results were achieved in ceramic-on-ceramic implants.

Determination of a friction coefficient for THA bearing couples

000      
00000naa a2200000 a 4500
001      
bmc16022576
003      
CZ-PrNML
005      
20190708091009.0
007      
ta
008      
160811s2015 xr f 000 0|cze||
009      
AR
024    7_
$2 doi $a 10.55095/achot2015/057
035    __
$a (PubMed)26516952
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a cze
044    __
$a xr
100    1_
$a Vrbka, Miloš $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno $7 xx0117094
245    10
$a Stanovení součinitele tření hlavic a jamek TEP kyčle / $c M. Vrbka, D. Nečas, J. Bartošík, M. Hartl, I. Křupka, A. Galandáková, J. Gallo
246    31
$a Determination of a friction coefficient for THA bearing couples
520    9_
$a PURPOSE OF THE STUDY: The wear of articular surfaces is considered one of the most important factors limiting the life of total hip arthroplasty (THA). It is assumed that the particles released from the surface of a softer material induce a complex inflammatory response, which will eventually result in osteolysis and aseptic loosening. Implant wear is related to a friction coefficient which depends on combination of the materials used, roughness of the articulating surfaces, internal clearance, and dimensions of the prosthesis. MATERIAL AND METHODS: The selected parameters of the bearing couples tested were studied using an experimental device based on the principle of a pendulum. Bovine serum was used as a lubricant and the load corresponded to a human body mass of 75 kg. The friction coefficient was derived from a curve of slowdown of pendulum oscillations. Roughness was measured with a device working on the principle of interferometry. Clearance was assessed by measuring diameters of the acetabular and femoral heads with a 3D optical scanner. The specimens tested included unused metal-on-highly cross-linked polyethylene, ceramic-on-highly cross-linked polyethylene and ceramic-on-ceramic bearing couples with the diameters of 28 mm and 36 mm. For each measured parameter, an arithmetic mean was calculated from 10 measurements. RESULTS: 1) The roughness of polyethylene surfaces was higher by about one order of magnitude than the roughness of metal and ceramic components. The Protasul metal head had the least rough surface (0.003 μm). 2) The ceramic-on-ceramic couples had the lowest clearance. Bearing couples with polyethylene acetabular liners had markedly higher clearances ranging from 150 μm to 545 μm. A clearance increased with large femoral heads (up to 4-fold in one of the couple tested). 3) The friction coefficient was related to the combination of materials; it was lowest in ceramic-on-ceramic surfaces (0.11 to 0.12) and then in ceramic-on-polyethylene implants (0.13 to 0.14). The friction coefficient is supposed to increase with a decreasing femoral head diameter. However, in the bearing couples with polyethylene liners manufactured by one company, paradoxically, the friction coefficient slightly increased with an increase in femoral head size from 28 mm to 36 mm. 4) The lowest friction moment (< 3.5 Nm) was found for ceramic-on-ceramic implants 28 mm in diameter; the highest values were recorded in metal-on-polyethylene bearing couples 36 mm in diameter (> 7 Nm). DISCUSSION: Although our study confirmed that the bearing couples produced by different manufacturers varied to some extent in the parameters studied, in our opinion, this variability was not significant because it was not within an order of magnitude in any of the tests. CONCLUSIONS: The study showed that both the friction coefficient and the friction moment are affected more by the combination of materials than by the diameter of a femoral head. The best results were achieved in ceramic-on-ceramic implants.
650    _2
$a náhrada kyčelního kloubu $x přístrojové vybavení $7 D019644
650    _2
$a keramika $7 D002516
650    _2
$a tření $7 D017276
650    12
$a kyčelní protézy $7 D006622
650    _2
$a lidé $7 D006801
650    _2
$a testování materiálů $x metody $7 D008422
650    _2
$a protézy - design $7 D011474
650    _2
$a selhání protézy $7 D011475
650    _2
$a povrchové vlastnosti $7 D013499
655    _2
$a anglický abstrakt $7 D004740
655    _2
$a časopisecké články $7 D016428
700    1_
$a Nečas, David $7 xx0209867 $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno
700    1_
$a Bartošík, J. $7 _AN088222 $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno
700    1_
$a Hartl, Martin, $7 mzk2002157052 $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno $d 1967-
700    1_
$a Křupka, Ivan, $7 mzk2002157050 $u Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně, Brno $d 1967-
700    1_
$a Galandáková, Adéla, $7 stk2008428623 $u Ústav lékařské chemie a biochemie LF UP v Olomouci, Olomouc $d 1979-
700    1_
$a Gallo, Jiří $7 xx0019005 $u Ortopedická klinika LF UP a FN Olomouc, Olomouc
773    0_
$w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $x 0001-5415 $g Roč. 82, č. 5 (2015), s. 341-347
910    __
$a ABA008 $b A 8 $c 507 $y 4 $z 0
990    __
$a 20160811 $b ABA008
991    __
$a 20190708091206 $b ABA008
999    __
$a ok $b bmc $g 1158903 $s 947121
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2015 $b 82 $c 5 $d 341-347 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
GRA    __
$a NT14267 $p MZ0
LZP    __
$b NLK118 $a Pubmed-20160811

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...