Methods to determine limit of detection and limit of quantification in quantitative real-time PCR (qPCR)

A. Forootan, R. Sjöback, J. Björkman, B. Sjögreen, L. Linz, M. Kubista,

. 2017 ; 12 (-) : 1-6. [pub] 20170429

Jazyk angličtina Země Německo

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc17032055

Quantitative Real-Time Polymerase Chain Reaction, better known as qPCR, is the most sensitive and specific technique we have for the detection of nucleic acids. Even though it has been around for more than 30 years and is preferred in research applications, it has yet to win broad acceptance in routine practice. This requires a means to unambiguously assess the performance of specific qPCR analyses. Here we present methods to determine the limit of detection (LoD) and the limit of quantification (LoQ) as applicable to qPCR. These are based on standard statistical methods as recommended by regulatory bodies adapted to qPCR and complemented with a novel approach to estimate the precision of LoD.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc17032055
003      
CZ-PrNML
005      
20171101100211.0
007      
ta
008      
171025s2017 gw f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.bdq.2017.04.001 $2 doi
035    __
$a (PubMed)28702366
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a gw
100    1_
$a Forootan, Amin $u Sahlgrenska Cancer Center, Department of Pathology, Sahlgrenska Academy at University of Gothenburg, Gothenburg, Sweden. MultiD Analysis AB, Gothenburg, Sweden.
245    10
$a Methods to determine limit of detection and limit of quantification in quantitative real-time PCR (qPCR) / $c A. Forootan, R. Sjöback, J. Björkman, B. Sjögreen, L. Linz, M. Kubista,
520    9_
$a Quantitative Real-Time Polymerase Chain Reaction, better known as qPCR, is the most sensitive and specific technique we have for the detection of nucleic acids. Even though it has been around for more than 30 years and is preferred in research applications, it has yet to win broad acceptance in routine practice. This requires a means to unambiguously assess the performance of specific qPCR analyses. Here we present methods to determine the limit of detection (LoD) and the limit of quantification (LoQ) as applicable to qPCR. These are based on standard statistical methods as recommended by regulatory bodies adapted to qPCR and complemented with a novel approach to estimate the precision of LoD.
655    _2
$a časopisecké články $7 D016428
700    1_
$a Sjöback, Robert $u TATAA Biocenter AB, Gothenburg, Sweden.
700    1_
$a Björkman, Jens $u TATAA Biocenter AB, Gothenburg, Sweden.
700    1_
$a Sjögreen, Björn $u MultiD Analysis AB, Gothenburg, Sweden.
700    1_
$a Linz, Lucas $u LGC Douglas Scientific, USA.
700    1_
$a Kubista, Mikael $u TATAA Biocenter AB, Gothenburg, Sweden. Institute of Biotechnology, CAS, BIOCEV, Vestec, Czech Republic.
773    0_
$w MED00189513 $t Biomolecular detection and quantification $x 2214-7535 $g Roč. 12, č. - (2017), s. 1-6
856    41
$u https://pubmed.ncbi.nlm.nih.gov/28702366 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20171025 $b ABA008
991    __
$a 20171101100302 $b ABA008
999    __
$a ind $b bmc $g 1255648 $s 993082
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2017 $b 12 $c - $d 1-6 $e 20170429 $i 2214-7535 $m Biomolecular detection and quantification $n Biomol Detect Quantif $x MED00189513
LZP    __
$a Pubmed-20171025

Najít záznam

Citační ukazatele

Pouze přihlášení uživatelé

Možnosti archivace

Nahrávání dat ...