Strukturální změny štěpu z hamstringu způsobené fixačním materiálem a jejich vliv na biomechanické vlastnosti náhrady LCA
[Structural damage to the hamstring graft due to interaction with fixation material and its effect on biomechanical properties of ACL reconstruction]
J. Kautzner, M. Držík, M. Handl, C. Povýšil, P. Kos, T. Trč, V. Havlas
Jazyk čeština Země Česko
Typ dokumentu anglický abstrakt, časopisecké články
PubMed
28809626
- Publikační typ
- anglický abstrakt MeSH
- časopisecké články MeSH
PURPOSE OF THE STUDY Hamstring grafts are commonly used for ACL reconstruction. The purpose of our study is to determine the effects of the suspension fixation compared to graft cross-pinning transfixation, and the effect(s) of structural damage during the preparation of the graft on biomechanical properties of the graft. MATERIAL AND METHODS The design of the study is a cadaveric biomechanical laboratory study. 38 fresh-frozen human hamstring specimens from 19 cadaveric donors were used. The grafts were tested for their loading properties. One half of each specimen was suspended over a 3.3mm pin, the other half was cross-pinned by a 3.3mm pin to simulate the graft cross-pinning technique. Single impact testing was performed and the failure force, elongation and acceleration/deceleration of each graft was recorded and the loading force vs. elongation of the graft specimens was calculated. Results for suspended and cross-pinned grafts were analysed using ANOVA method, comparing the grafts from each donor. RESULTS The ultimate strength of a double-strand gracilis graft was 1287 ± 134 N when suspended over a pin, the strength of a cross-pinned graft was 833 ± 111 N. For double-strand semitendinosus grafts the strengths were 1883 ± 198 and 997 ± 234 N, respectively. Thus, the failure load for the cross-pinning method is only 64.7% or 52.9% for the suspension method. DISCUSSION Structural damage to the graft significantly reduces the graft strength. Also, extensive suturing during preparation of the graft reduces its strength. CONCLUSIONS Fixation methods that do not interfere with the graft's structure should be used to reduce the risk of graft failure. Key words: ACL reconstruction, hamstring graft, biomechanical testing.
International Laser Center Bratislava
Ústav patologie 1 lékařské fakulty Univerzity Karlovy a Všeobecné fakultní nemocnice Praha
Structural damage to the hamstring graft due to interaction with fixation material and its effect on biomechanical properties of ACL reconstruction
- 000
- 00000naa a2200000 a 4500
- 001
- bmc18003832
- 003
- CZ-PrNML
- 005
- 20180215090048.0
- 007
- ta
- 008
- 180131s2017 xr ad f 000 0|cze||
- 009
- AR
- 024 7_
- $2 doi $a 10.55095/achot2017/015
- 035 __
- $a (PubMed)28809626
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a cze $b eng
- 044 __
- $a xr
- 100 1_
- $a Kautzner, Jakub $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha $7 xx0227523
- 245 10
- $a Strukturální změny štěpu z hamstringu způsobené fixačním materiálem a jejich vliv na biomechanické vlastnosti náhrady LCA / $c J. Kautzner, M. Držík, M. Handl, C. Povýšil, P. Kos, T. Trč, V. Havlas
- 246 31
- $a Structural damage to the hamstring graft due to interaction with fixation material and its effect on biomechanical properties of ACL reconstruction
- 520 9_
- $a PURPOSE OF THE STUDY Hamstring grafts are commonly used for ACL reconstruction. The purpose of our study is to determine the effects of the suspension fixation compared to graft cross-pinning transfixation, and the effect(s) of structural damage during the preparation of the graft on biomechanical properties of the graft. MATERIAL AND METHODS The design of the study is a cadaveric biomechanical laboratory study. 38 fresh-frozen human hamstring specimens from 19 cadaveric donors were used. The grafts were tested for their loading properties. One half of each specimen was suspended over a 3.3mm pin, the other half was cross-pinned by a 3.3mm pin to simulate the graft cross-pinning technique. Single impact testing was performed and the failure force, elongation and acceleration/deceleration of each graft was recorded and the loading force vs. elongation of the graft specimens was calculated. Results for suspended and cross-pinned grafts were analysed using ANOVA method, comparing the grafts from each donor. RESULTS The ultimate strength of a double-strand gracilis graft was 1287 ± 134 N when suspended over a pin, the strength of a cross-pinned graft was 833 ± 111 N. For double-strand semitendinosus grafts the strengths were 1883 ± 198 and 997 ± 234 N, respectively. Thus, the failure load for the cross-pinning method is only 64.7% or 52.9% for the suspension method. DISCUSSION Structural damage to the graft significantly reduces the graft strength. Also, extensive suturing during preparation of the graft reduces its strength. CONCLUSIONS Fixation methods that do not interfere with the graft's structure should be used to reduce the risk of graft failure. Key words: ACL reconstruction, hamstring graft, biomechanical testing.
- 655 _2
- $a anglický abstrakt $7 D004740
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Držík, Milan $7 _BN002280 $u International Laser Center, Bratislava
- 700 1_
- $a Handl, Milan, $d 1961- $7 xx0064271 $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha
- 700 1_
- $a Povýšil, Ctibor, $d 1943- $7 jn20001103579 $u Ústav patologie 1. lékařské fakulty Univerzity Karlovy a Všeobecné fakultní nemocnice, Praha
- 700 1_
- $a Kos, Petr $7 xx0135447 $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha
- 700 1_
- $a Trč, Tomáš, $d 1955- $7 nlk19990073953 $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha
- 700 1_
- $a Havlas, Vojtěch, $d 1971- $7 xx0065575 $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha
- 773 0_
- $w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $x 0001-5415 $g Roč. 84, č. 2 (2017), s. 101-105
- 910 __
- $a ABA008 $b A 8 $c 507 $y 4 $z 0
- 990 __
- $a 20180131 $b ABA008
- 991 __
- $a 20180213094830 $b ABA008
- 999 __
- $a ok $b bmc $g 1276021 $s 1000534
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2017 $b 84 $c 2 $d 101-105 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
- LZP __
- $b NLK118 $a Pubmed-20180131