Strukturální změny štěpu z hamstringu způsobené fixačním materiálem a jejich vliv na biomechanické vlastnosti náhrady LCA
[Structural damage to the hamstring graft due to interaction with fixation material and its effect on biomechanical properties of ACL reconstruction]

J. Kautzner, M. Držík, M. Handl, C. Povýšil, P. Kos, T. Trč, V. Havlas

. 2017 ; 84 (2) : 101-105.

Jazyk čeština Země Česko

Typ dokumentu anglický abstrakt, časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc18003832

Digitální knihovna NLK
Zdroj

E-zdroje Online

NLK Free Medical Journals od 2006

Odkazy

PubMed 28809626

PURPOSE OF THE STUDY Hamstring grafts are commonly used for ACL reconstruction. The purpose of our study is to determine the effects of the suspension fixation compared to graft cross-pinning transfixation, and the effect(s) of structural damage during the preparation of the graft on biomechanical properties of the graft. MATERIAL AND METHODS The design of the study is a cadaveric biomechanical laboratory study. 38 fresh-frozen human hamstring specimens from 19 cadaveric donors were used. The grafts were tested for their loading properties. One half of each specimen was suspended over a 3.3mm pin, the other half was cross-pinned by a 3.3mm pin to simulate the graft cross-pinning technique. Single impact testing was performed and the failure force, elongation and acceleration/deceleration of each graft was recorded and the loading force vs. elongation of the graft specimens was calculated. Results for suspended and cross-pinned grafts were analysed using ANOVA method, comparing the grafts from each donor. RESULTS The ultimate strength of a double-strand gracilis graft was 1287 ± 134 N when suspended over a pin, the strength of a cross-pinned graft was 833 ± 111 N. For double-strand semitendinosus grafts the strengths were 1883 ± 198 and 997 ± 234 N, respectively. Thus, the failure load for the cross-pinning method is only 64.7% or 52.9% for the suspension method. DISCUSSION Structural damage to the graft significantly reduces the graft strength. Also, extensive suturing during preparation of the graft reduces its strength. CONCLUSIONS Fixation methods that do not interfere with the graft's structure should be used to reduce the risk of graft failure. Key words: ACL reconstruction, hamstring graft, biomechanical testing.

Structural damage to the hamstring graft due to interaction with fixation material and its effect on biomechanical properties of ACL reconstruction

000      
00000naa a2200000 a 4500
001      
bmc18003832
003      
CZ-PrNML
005      
20180215090048.0
007      
ta
008      
180131s2017 xr ad f 000 0|cze||
009      
AR
024    7_
$2 doi $a 10.55095/achot2017/015
035    __
$a (PubMed)28809626
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a cze $b eng
044    __
$a xr
100    1_
$a Kautzner, Jakub $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha $7 xx0227523
245    10
$a Strukturální změny štěpu z hamstringu způsobené fixačním materiálem a jejich vliv na biomechanické vlastnosti náhrady LCA / $c J. Kautzner, M. Držík, M. Handl, C. Povýšil, P. Kos, T. Trč, V. Havlas
246    31
$a Structural damage to the hamstring graft due to interaction with fixation material and its effect on biomechanical properties of ACL reconstruction
520    9_
$a PURPOSE OF THE STUDY Hamstring grafts are commonly used for ACL reconstruction. The purpose of our study is to determine the effects of the suspension fixation compared to graft cross-pinning transfixation, and the effect(s) of structural damage during the preparation of the graft on biomechanical properties of the graft. MATERIAL AND METHODS The design of the study is a cadaveric biomechanical laboratory study. 38 fresh-frozen human hamstring specimens from 19 cadaveric donors were used. The grafts were tested for their loading properties. One half of each specimen was suspended over a 3.3mm pin, the other half was cross-pinned by a 3.3mm pin to simulate the graft cross-pinning technique. Single impact testing was performed and the failure force, elongation and acceleration/deceleration of each graft was recorded and the loading force vs. elongation of the graft specimens was calculated. Results for suspended and cross-pinned grafts were analysed using ANOVA method, comparing the grafts from each donor. RESULTS The ultimate strength of a double-strand gracilis graft was 1287 ± 134 N when suspended over a pin, the strength of a cross-pinned graft was 833 ± 111 N. For double-strand semitendinosus grafts the strengths were 1883 ± 198 and 997 ± 234 N, respectively. Thus, the failure load for the cross-pinning method is only 64.7% or 52.9% for the suspension method. DISCUSSION Structural damage to the graft significantly reduces the graft strength. Also, extensive suturing during preparation of the graft reduces its strength. CONCLUSIONS Fixation methods that do not interfere with the graft's structure should be used to reduce the risk of graft failure. Key words: ACL reconstruction, hamstring graft, biomechanical testing.
655    _2
$a anglický abstrakt $7 D004740
655    _2
$a časopisecké články $7 D016428
700    1_
$a Držík, Milan $7 _BN002280 $u International Laser Center, Bratislava
700    1_
$a Handl, Milan, $d 1961- $7 xx0064271 $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha
700    1_
$a Povýšil, Ctibor, $d 1943- $7 jn20001103579 $u Ústav patologie 1. lékařské fakulty Univerzity Karlovy a Všeobecné fakultní nemocnice, Praha
700    1_
$a Kos, Petr $7 xx0135447 $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha
700    1_
$a Trč, Tomáš, $d 1955- $7 nlk19990073953 $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha
700    1_
$a Havlas, Vojtěch, $d 1971- $7 xx0065575 $u Klinika dětské a dospělé ortopedie a traumatologie 2. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice Motol, Praha
773    0_
$w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $x 0001-5415 $g Roč. 84, č. 2 (2017), s. 101-105
910    __
$a ABA008 $b A 8 $c 507 $y 4 $z 0
990    __
$a 20180131 $b ABA008
991    __
$a 20180213094830 $b ABA008
999    __
$a ok $b bmc $g 1276021 $s 1000534
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2017 $b 84 $c 2 $d 101-105 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
LZP    __
$b NLK118 $a Pubmed-20180131

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...