Structural and mechanical implications of PMMA implant shape and interface geometry in cranioplasty--A finite element study

A. Ridwan-Pramana, P. Marcián, L. Borák, N. Narra, T. Forouzanfar, J. Wolff,

. 2016 ; 44 (1) : 34-44. [pub] 20151023

Language English Country Great Britain

Document type Journal Article, Research Support, Non-U.S. Gov't

This computational study investigates the effect of shape (defect contour curvature) and bone-implant interface (osteotomy angle) on the stress distribution within PMMA skull implants. Using finite element methodology, 15 configurations--combinations of simplified synthetic geometric shapes (circular, square, triangular, irregular) and interface angulations--were simulated under 50N static loads. Furthermore, the implant fixation devices were modelled and analysed in detail. Negative osteotomy configurations demonstrated the largest stresses in the implant (275 MPa), fixation devices (1258 MPa) and bone strains (0.04). The circular implant with zero and positive osteotomy performed well with maximum observed magnitudes of--implant stress (1.2 MPa and 1.2 MPa), fixation device stress (11.2 MPa and 2.2 MPa), bone strain (0.218e-3 and 0.750e-4). The results suggest that the preparation of defect sites is a critical procedure. Of the greatest importance is the angle at which the edges of the defect are sawed. If under an external load, the implant has no support from the interface and the stresses are transferred to the fixation devices. This can endanger their material integrity and lead to unphysiological strains in the adjacent bone, potentially compromising the bone morphology required for anchoring. These factors can ultimately weaken the stability of the entire implant assembly.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc18017379
003      
CZ-PrNML
005      
20180521140855.0
007      
ta
008      
180515s2016 xxk f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.jcms.2015.10.014 $2 doi
035    __
$a (PubMed)26646634
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxk
100    1_
$a Ridwan-Pramana, Angela $u Department of Oral and Maxillofacial Surgery/Oral Pathology, VU University Medical Center, Amsterdam, The Netherlands; Department of Maxillofacial Prosthodontics, Center for Special Care in Dentistry, Amsterdam, The Netherlands.
245    10
$a Structural and mechanical implications of PMMA implant shape and interface geometry in cranioplasty--A finite element study / $c A. Ridwan-Pramana, P. Marcián, L. Borák, N. Narra, T. Forouzanfar, J. Wolff,
520    9_
$a This computational study investigates the effect of shape (defect contour curvature) and bone-implant interface (osteotomy angle) on the stress distribution within PMMA skull implants. Using finite element methodology, 15 configurations--combinations of simplified synthetic geometric shapes (circular, square, triangular, irregular) and interface angulations--were simulated under 50N static loads. Furthermore, the implant fixation devices were modelled and analysed in detail. Negative osteotomy configurations demonstrated the largest stresses in the implant (275 MPa), fixation devices (1258 MPa) and bone strains (0.04). The circular implant with zero and positive osteotomy performed well with maximum observed magnitudes of--implant stress (1.2 MPa and 1.2 MPa), fixation device stress (11.2 MPa and 2.2 MPa), bone strain (0.218e-3 and 0.750e-4). The results suggest that the preparation of defect sites is a critical procedure. Of the greatest importance is the angle at which the edges of the defect are sawed. If under an external load, the implant has no support from the interface and the stresses are transferred to the fixation devices. This can endanger their material integrity and lead to unphysiological strains in the adjacent bone, potentially compromising the bone morphology required for anchoring. These factors can ultimately weaken the stability of the entire implant assembly.
650    12
$a rozhraní kost/implantát $7 D000069343
650    12
$a zubní implantáty $7 D015921
650    _2
$a analýza zatížení zubů $7 D003799
650    _2
$a analýza metodou konečných prvků $7 D020342
650    _2
$a lidé $7 D006801
650    _2
$a polymethylmethakrylát $7 D019904
650    12
$a protézy a implantáty $7 D019736
650    12
$a lebka $7 D012886
650    _2
$a mechanický stres $7 D013314
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Marcián, Petr $u Institute of Solid Mechanics, Mechatronics and Biomechanics, Brno University of Technology, Czech Republic.
700    1_
$a Borák, Libor $u Institute of Solid Mechanics, Mechatronics and Biomechanics, Brno University of Technology, Czech Republic.
700    1_
$a Narra, Nathaniel $u Department of Electronics and Communications Engineering, Tampere University of Technology, BioMediTech, Finland. Electronic address: nathaniel.narragirish@tut.fi.
700    1_
$a Forouzanfar, Tim $u Department of Oral and Maxillofacial Surgery/Oral Pathology, VU University Medical Center, Amsterdam, The Netherlands.
700    1_
$a Wolff, Jan $u Department of Oral and Maxillofacial Surgery/Oral Pathology, VU University Medical Center, Amsterdam, The Netherlands.
773    0_
$w MED00002623 $t Journal of cranio-maxillo-facial surgery official publication of the European Association for Cranio-Maxillo-Facial Surgery $x 1878-4119 $g Roč. 44, č. 1 (2016), s. 34-44
856    41
$u https://pubmed.ncbi.nlm.nih.gov/26646634 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20180515 $b ABA008
991    __
$a 20180521141037 $b ABA008
999    __
$a ok $b bmc $g 1301003 $s 1014219
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2016 $b 44 $c 1 $d 34-44 $e 20151023 $i 1878-4119 $m Journal of cranio-maxillofacial surgery $n J Craniomaxillofac Surg $x MED00002623
LZP    __
$a Pubmed-20180515

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...