Separation of structurally related primary aliphatic amines using hydrophilic interaction chromatography with fluorescence detection after postcolumn derivatization with o-phthaldialdehyde/mercaptoethanol
M. Douša, J. Doubský,
Language English Country Germany
Document type Journal Article
- MeSH
- Amines isolation & purification MeSH
- Cyclohexylamines isolation & purification MeSH
- Cyclopropanes isolation & purification MeSH
- Hydrophobic and Hydrophilic Interactions MeSH
- Pharmaceutical Preparations chemistry MeSH
- Mercaptoethanol * MeSH
- o-Phthalaldehyde * MeSH
- Propylamines isolation & purification MeSH
- Chromatography, High Pressure Liquid MeSH
- Publication type
- Journal Article MeSH
The retention behavior of primary aliphatic amines (homologous series of aliphatic alkyl amines and cycloalkyl amines) and positional isomers of alkylamines in the hydrophilic interaction chromatography mode was studied. The study was carried out on a TSKgel Amide-80 column followed by postcolumn derivatization with fluorescence detection to describe the retention mechanism of tested compounds. The effect of chromatographic conditions including column temperature, acetonitrile content in the mobile phase, mobile phase pH (ranging from 3.5 to 6.8), and salt concentration in the mobile phase was investigated. The final mobile phase consisted of acetonitrile and solution of 20 mM potassium formate pH 3.5 in ratio 80:20 v/v. The analyses were carried out at mobile phase flow rate of 1.0 mL/min and the column temperature of 20°C. The developed method was fully validated in terms of linearity, sensitivity (limit of detection and limit of quantification), accuracy, and precision according to International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use guidelines. The proposed new methods were proved to be highly sensitive, simple, and rapid, and were successfully applied to the determinations of isopropylamine, cyclohexylamine, and cyclopropylamine in relevant active pharmaceutical ingredients.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc18024541
- 003
- CZ-PrNML
- 005
- 20180710095137.0
- 007
- ta
- 008
- 180709s2017 gw f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1002/jssc.201701005 $2 doi
- 035 __
- $a (PubMed)29057598
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a gw
- 100 1_
- $a Douša, Michal $u Zentiva, k.s. Praha, a Sanofi Company, Prague, Czech Republic.
- 245 10
- $a Separation of structurally related primary aliphatic amines using hydrophilic interaction chromatography with fluorescence detection after postcolumn derivatization with o-phthaldialdehyde/mercaptoethanol / $c M. Douša, J. Doubský,
- 520 9_
- $a The retention behavior of primary aliphatic amines (homologous series of aliphatic alkyl amines and cycloalkyl amines) and positional isomers of alkylamines in the hydrophilic interaction chromatography mode was studied. The study was carried out on a TSKgel Amide-80 column followed by postcolumn derivatization with fluorescence detection to describe the retention mechanism of tested compounds. The effect of chromatographic conditions including column temperature, acetonitrile content in the mobile phase, mobile phase pH (ranging from 3.5 to 6.8), and salt concentration in the mobile phase was investigated. The final mobile phase consisted of acetonitrile and solution of 20 mM potassium formate pH 3.5 in ratio 80:20 v/v. The analyses were carried out at mobile phase flow rate of 1.0 mL/min and the column temperature of 20°C. The developed method was fully validated in terms of linearity, sensitivity (limit of detection and limit of quantification), accuracy, and precision according to International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use guidelines. The proposed new methods were proved to be highly sensitive, simple, and rapid, and were successfully applied to the determinations of isopropylamine, cyclohexylamine, and cyclopropylamine in relevant active pharmaceutical ingredients.
- 650 _2
- $a aminy $x izolace a purifikace $7 D000588
- 650 _2
- $a vysokoúčinná kapalinová chromatografie $7 D002851
- 650 _2
- $a cyklohexylaminy $x izolace a purifikace $7 D003514
- 650 _2
- $a cyklopropany $x izolace a purifikace $7 D003521
- 650 _2
- $a hydrofobní a hydrofilní interakce $7 D057927
- 650 12
- $a merkaptoethanol $7 D008623
- 650 _2
- $a léčivé přípravky $x chemie $7 D004364
- 650 _2
- $a propylaminy $x izolace a purifikace $7 D011437
- 650 12
- $a o-ftalaldehyd $7 D009764
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Doubský, Jan $u Zentiva, k.s. Praha, a Sanofi Company, Prague, Czech Republic.
- 773 0_
- $w MED00006463 $t Journal of separation science $x 1615-9314 $g Roč. 40, č. 24 (2017), s. 4689-4699
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/29057598 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20180709 $b ABA008
- 991 __
- $a 20180710095427 $b ABA008
- 999 __
- $a ok $b bmc $g 1316672 $s 1021462
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2017 $b 40 $c 24 $d 4689-4699 $i 1615-9314 $m Journal of separation science $n J Sep Sci $x MED00006463
- LZP __
- $a Pubmed-20180709