Evaluation of FUS-2000 urine analyzer: analytical properties and particle recognition

M. Beňovská, O. Wiewiorka, J. Pinkavová,

. 2018 ; 78 (1-2) : 143-148. [pub] 20180108

Jazyk angličtina Země Anglie, Velká Británie

Typ dokumentu hodnotící studie, časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc18033404

This study evaluates the performance of microscopic part of a hybrid analyzer FUS-2000 (Dirui Industrial Co., Changchun, China), its analytical properties and particle recognition. The evaluation of trueness, repeatability, detection limit, carry-over, linearity range and analytical stability was performed according to Dirui protocol guidelines designed by Dirui Company to guarantee the quality of the instrument. Trueness for low, medium and high-value concentrations was calculated with bias of 15.5, 4.7 and -6.6%, respectively. Detection limit of 5 Ery/μl was confirmed. Coefficient of variation of 11.0, 5.2 and 3.8% was measured for within-run repeatability of low, medium and high concentration. Between-run repeatability for daily quality control had coefficient of variation of 3.0%. Carry-over did not exceed 0.05%. Linearity was confirmed for range of 0-16,000 particles/μl (R2 = 0.9997). The analytical stability had coefficient of variation of 4.3%. Out of 1258 analyzed urine samples, 362 positive were subjected to light microscopy urine sediment analysis and compared to the analyzer results. Cohen's kappa coefficients were calculated to express the concordance. Squared kappa coefficient was 0.927 (red blood cells), 0.888 (white blood cells), 0.908 (squamous epithelia), 0.634 (transitional epithelia), 0.628 (hyaline casts), 0.843 (granular casts) and 0.623 (bacteria). Single kappa coefficients were 0.885 (yeasts) and 0.756 (crystals), respectively. Aforementioned results show good analytical performance of the analyzer and tight agreement with light microscopy of urine sediment.

000      
00000naa a2200000 a 4500
001      
bmc18033404
003      
CZ-PrNML
005      
20210608140900.0
007      
ta
008      
181008s2018 enk f 000 0|eng||
009      
AR
024    7_
$a 10.1080/00365513.2017.1423108 $2 doi
035    __
$a (PubMed)29310458
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Beňovská, Miroslava $u a Department of Clinical Biochemistry , University Hospital Brno , Brno , Czech Republic. b Department of Laboratory Methods, Faculty of Medicine , Masaryk University , Brno , Czech Republic.
245    10
$a Evaluation of FUS-2000 urine analyzer: analytical properties and particle recognition / $c M. Beňovská, O. Wiewiorka, J. Pinkavová,
520    9_
$a This study evaluates the performance of microscopic part of a hybrid analyzer FUS-2000 (Dirui Industrial Co., Changchun, China), its analytical properties and particle recognition. The evaluation of trueness, repeatability, detection limit, carry-over, linearity range and analytical stability was performed according to Dirui protocol guidelines designed by Dirui Company to guarantee the quality of the instrument. Trueness for low, medium and high-value concentrations was calculated with bias of 15.5, 4.7 and -6.6%, respectively. Detection limit of 5 Ery/μl was confirmed. Coefficient of variation of 11.0, 5.2 and 3.8% was measured for within-run repeatability of low, medium and high concentration. Between-run repeatability for daily quality control had coefficient of variation of 3.0%. Carry-over did not exceed 0.05%. Linearity was confirmed for range of 0-16,000 particles/μl (R2 = 0.9997). The analytical stability had coefficient of variation of 4.3%. Out of 1258 analyzed urine samples, 362 positive were subjected to light microscopy urine sediment analysis and compared to the analyzer results. Cohen's kappa coefficients were calculated to express the concordance. Squared kappa coefficient was 0.927 (red blood cells), 0.888 (white blood cells), 0.908 (squamous epithelia), 0.634 (transitional epithelia), 0.628 (hyaline casts), 0.843 (granular casts) and 0.623 (bacteria). Single kappa coefficients were 0.885 (yeasts) and 0.756 (crystals), respectively. Aforementioned results show good analytical performance of the analyzer and tight agreement with light microscopy of urine sediment.
650    _2
$a směrnice jako téma $7 D017408
650    _2
$a lidé $7 D006801
650    _2
$a referenční hodnoty $7 D012016
650    _2
$a analýza moči $x přístrojové vybavení $7 D016482
655    _2
$a hodnotící studie $7 D023362
655    _2
$a časopisecké články $7 D016428
700    1_
$a Wiewiorka, Ondřej $u a Department of Clinical Biochemistry , University Hospital Brno , Brno , Czech Republic. b Department of Laboratory Methods, Faculty of Medicine , Masaryk University , Brno , Czech Republic. c Department of Biochemistry, Faculty of Medicine , Masaryk University , Brno , Czech Republic. $7 xx0260774
700    1_
$a Pinkavová, Jana $u a Department of Clinical Biochemistry , University Hospital Brno , Brno , Czech Republic.
773    0_
$w MED00010597 $t Scandinavian journal of clinical and laboratory investigation $x 1502-7686 $g Roč. 78, č. 1-2 (2018), s. 143-148
856    41
$u https://pubmed.ncbi.nlm.nih.gov/29310458 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20181008 $b ABA008
991    __
$a 20210608140858 $b ABA008
999    __
$a ok $b bmc $g 1340138 $s 1030398
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2018 $b 78 $c 1-2 $d 143-148 $e 20180108 $i 1502-7686 $m Scandinavian journal of clinical & laboratory investigation $n Scand J Clin Lab Invest $x MED00010597
LZP    __
$a Pubmed-20181008

Najít záznam

Citační ukazatele

Nahrávání dat...

Možnosti archivace

Nahrávání dat...