Inter-comparison of quantitative imaging of lutetium-177 (177Lu) in European hospitals
J. Wevrett, A. Fenwick, J. Scuffham, L. Johansson, J. Gear, S. Schlögl, M. Segbers, K. Sjögreen-Gleisner, P. Solný, M. Lassmann, J. Tipping, A. Nisbet,
Language English Country Germany
Document type Journal Article
NLK
Directory of Open Access Journals
from 2014
Free Medical Journals
from 2014
PubMed Central
from 2014
Europe PubMed Central
from 2014
ProQuest Central
from 2014-05-01
Open Access Digital Library
from 2014-01-01
Open Access Digital Library
from 2014-01-01
ROAD: Directory of Open Access Scholarly Resources
from 2014
Springer Journals Complete - Open Access
from 2014-12-01
Springer Nature OA/Free Journals
from 2014-12-01
- Publication type
- Journal Article MeSH
BACKGROUND: This inter-comparison exercise was performed to demonstrate the variability of quantitative SPECT/CT imaging for lutetium-177 (177Lu) in current clinical practice. Our aim was to assess the feasibility of using international inter-comparison exercises as a means to ensure consistency between clinical sites whilst enabling the sites to use their own choice of quantitative imaging protocols, specific to their systems. Dual-compartment concentric spherical sources of accurately known activity concentrations were prepared and sent to seven European clinical sites. The site staff were not aware of the true volumes or activity within the sources-they performed SPECT/CT imaging of the source, positioned within a water-filled phantom, using their own choice of parameters and reported their estimate of the activities within the source. RESULTS: The volumes reported by the participants for the inner section of the source were all within 29% of the true value and within 60% of the true value for the outer section. The activities reported by the participants for the inner section of the source were all within 20% of the true value, whilst those reported for the outer section were up to 83% different to the true value. CONCLUSIONS: A variety of calibration and segmentation methods were used by the participants for this exercise which demonstrated the variability of quantitative imaging across clinical sites. This paper presents a method to assess consistency between sites using different calibration and segmentation methods.
Department of Medical Radiation Physics Clinical Sciences Lund Lund University Lund Sweden
Erasmus University Medical Centre Rotterdam Netherlands
Motol University Hospital Prague Czech Republic
National Physical Laboratory Teddington UK
The Christie NHS Foundation Trust Manchester UK
The Royal Marsden NHS Foundation Trust Sutton UK
University Hospital of Wϋrzburg Wϋrzburg Germany
University of Surrey Guildford UK Royal Surrey County Hospital NHS Foundation Trust Guildford UK
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc18034332
- 003
- CZ-PrNML
- 005
- 20181026114447.0
- 007
- ta
- 008
- 181008s2018 gw f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1186/s40658-018-0213-z $2 doi
- 035 __
- $a (PubMed)30069805
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a gw
- 100 1_
- $a Wevrett, Jill $u University of Surrey, Guildford, UK. Jill.Wevrett@nhs.net. National Physical Laboratory, Teddington, UK. Jill.Wevrett@nhs.net. Royal Surrey County Hospital NHS Foundation Trust, Guildford, UK. Jill.Wevrett@nhs.net.
- 245 10
- $a Inter-comparison of quantitative imaging of lutetium-177 (177Lu) in European hospitals / $c J. Wevrett, A. Fenwick, J. Scuffham, L. Johansson, J. Gear, S. Schlögl, M. Segbers, K. Sjögreen-Gleisner, P. Solný, M. Lassmann, J. Tipping, A. Nisbet,
- 520 9_
- $a BACKGROUND: This inter-comparison exercise was performed to demonstrate the variability of quantitative SPECT/CT imaging for lutetium-177 (177Lu) in current clinical practice. Our aim was to assess the feasibility of using international inter-comparison exercises as a means to ensure consistency between clinical sites whilst enabling the sites to use their own choice of quantitative imaging protocols, specific to their systems. Dual-compartment concentric spherical sources of accurately known activity concentrations were prepared and sent to seven European clinical sites. The site staff were not aware of the true volumes or activity within the sources-they performed SPECT/CT imaging of the source, positioned within a water-filled phantom, using their own choice of parameters and reported their estimate of the activities within the source. RESULTS: The volumes reported by the participants for the inner section of the source were all within 29% of the true value and within 60% of the true value for the outer section. The activities reported by the participants for the inner section of the source were all within 20% of the true value, whilst those reported for the outer section were up to 83% different to the true value. CONCLUSIONS: A variety of calibration and segmentation methods were used by the participants for this exercise which demonstrated the variability of quantitative imaging across clinical sites. This paper presents a method to assess consistency between sites using different calibration and segmentation methods.
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Fenwick, Andrew $u National Physical Laboratory, Teddington, UK.
- 700 1_
- $a Scuffham, James $u University of Surrey, Guildford, UK. Royal Surrey County Hospital NHS Foundation Trust, Guildford, UK.
- 700 1_
- $a Johansson, Lena $u National Physical Laboratory, Teddington, UK.
- 700 1_
- $a Gear, Jonathan $u The Royal Marsden NHS Foundation Trust, Sutton, UK.
- 700 1_
- $a Schlögl, Susanne $u University Hospital of Wϋrzburg, Wϋrzburg, Germany.
- 700 1_
- $a Segbers, Marcel $u Erasmus University Medical Centre, Rotterdam, Netherlands.
- 700 1_
- $a Sjögreen-Gleisner, Katarina $u Department of Medical Radiation Physics, Clinical Sciences Lund, Lund University, Lund, Sweden.
- 700 1_
- $a Solný, Pavel $u Motol University Hospital, Prague, Czech Republic.
- 700 1_
- $a Lassmann, Michael $u University Hospital of Wϋrzburg, Wϋrzburg, Germany.
- 700 1_
- $a Tipping, Jill $u The Christie NHS Foundation Trust, Manchester, UK.
- 700 1_
- $a Nisbet, Andrew $u University of Surrey, Guildford, UK. Royal Surrey County Hospital NHS Foundation Trust, Guildford, UK.
- 773 0_
- $w MED00194294 $t EJNMMI physics $x 2197-7364 $g Roč. 5, č. 1 (2018), s. 17
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/30069805 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20181008 $b ABA008
- 991 __
- $a 20181026114958 $b ABA008
- 999 __
- $a ind $b bmc $g 1340515 $s 1031326
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2018 $b 5 $c 1 $d 17 $e 20180802 $i 2197-7364 $m EJNMMI physics $n EJNMMI Phys $x MED00194294
- LZP __
- $a Pubmed-20181008