PURPOSE: Use of novel medical technologies, such as leadless pacemaker (LP) therapy, may be subjected to a learning curve effect. The objective of the current study was to assess the impact of operators' experience on the occurrence of serious adverse device effects (SADE) and procedural efficiency. METHODS: Patients implanted with a Nanostim LP (Abbott, USA) within two prospective studies (i.e., LEADLESS ll IDE and Leadless Observational Study) were assessed. Patients were categorized into quartiles based on operator experience. Learning curve analysis included the comparison of SADE rates at 30 days post-implant per quartile and between patients in quartile 4 (> 10 implants) and patients in quartiles 1 through 3 (1-10 implants). Procedural efficiency was assessed based on procedure duration and repositioning attempts. RESULTS: Nanostim LP implant was performed in 1439 patients by 171 implanters at 60 centers in 10 countries. A total of 91 (6.4%) patients experienced a SADE in the first 30 days. SADE rates dropped from 7.4 to 4.5% (p = 0.038) after more than 10 implants per operator. Total procedure duration decreased from 30.9 ± 19.1 min in quartile 1 to 21.6 ± 13.2 min (p < 0.001) in quartile 4. The need for multiple repositionings during the LP procedure reduced in quartile 4 (14.8%), compared to quartiles 1 (26.8%; p < 0.001), 2 (26.6%; p < 0.001), and 3 (20.4%; p = 0.03). CONCLUSIONS: Learning curves exist for Nanostim LP implantation. Procedure efficiency improved with increased operator experience, according to a decrease in the incidence of SADE, procedure duration, and repositioning attempts.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc19012465
003      
CZ-PrNML
005      
20190415133852.0
007      
ta
008      
190405s2018 ne f 000 0|eng||
009      
AR
024    7_
$a 10.1007/s10840-018-0438-8 $2 doi
035    __
$a (PubMed)30105428
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a ne
100    1_
$a Tjong, Fleur V Y $u Department of Clinical and Experimental Cardiology, Amsterdam UMC, location AMC, University of Amsterdam, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands. f.v.tjong@amc.nl.
245    14
$a The learning curve associated with the implantation of the Nanostim leadless pacemaker / $c FVY. Tjong, NEG. Beurskens, P. Neuzil, P. Defaye, PP. Delnoy, J. Ip, JJG. Guerrero, M. Rashtian, R. Banker, V. Reddy, D. Exner, J. Sperzel, RE. Knops, . , . ,
520    9_
$a PURPOSE: Use of novel medical technologies, such as leadless pacemaker (LP) therapy, may be subjected to a learning curve effect. The objective of the current study was to assess the impact of operators' experience on the occurrence of serious adverse device effects (SADE) and procedural efficiency. METHODS: Patients implanted with a Nanostim LP (Abbott, USA) within two prospective studies (i.e., LEADLESS ll IDE and Leadless Observational Study) were assessed. Patients were categorized into quartiles based on operator experience. Learning curve analysis included the comparison of SADE rates at 30 days post-implant per quartile and between patients in quartile 4 (> 10 implants) and patients in quartiles 1 through 3 (1-10 implants). Procedural efficiency was assessed based on procedure duration and repositioning attempts. RESULTS: Nanostim LP implant was performed in 1439 patients by 171 implanters at 60 centers in 10 countries. A total of 91 (6.4%) patients experienced a SADE in the first 30 days. SADE rates dropped from 7.4 to 4.5% (p = 0.038) after more than 10 implants per operator. Total procedure duration decreased from 30.9 ± 19.1 min in quartile 1 to 21.6 ± 13.2 min (p < 0.001) in quartile 4. The need for multiple repositionings during the LP procedure reduced in quartile 4 (14.8%), compared to quartiles 1 (26.8%; p < 0.001), 2 (26.6%; p < 0.001), and 3 (20.4%; p = 0.03). CONCLUSIONS: Learning curves exist for Nanostim LP implantation. Procedure efficiency improved with increased operator experience, according to a decrease in the incidence of SADE, procedure duration, and repositioning attempts.
650    _2
$a senioři $7 D000368
650    _2
$a senioři nad 80 let $7 D000369
650    _2
$a srdeční arytmie $x diagnostické zobrazování $x terapie $7 D001145
650    _2
$a srdeční elektrofyziologie $x výchova $7 D054849
650    _2
$a kardiostimulace umělá $x škodlivé účinky $x metody $7 D002304
650    _2
$a implantované elektrody $7 D004567
650    12
$a design vybavení $7 D004867
650    _2
$a ženské pohlaví $7 D005260
650    _2
$a lidé $7 D006801
650    _2
$a internacionalita $7 D038622
650    12
$a křivka učení $7 D059032
650    _2
$a logistické modely $7 D016015
650    _2
$a mužské pohlaví $7 D008297
650    _2
$a monitorování fyziologických funkcí $x metody $7 D008991
650    _2
$a multivariační analýza $7 D015999
650    12
$a kardiostimulátor $7 D010138
650    _2
$a prognóza $7 D011379
650    _2
$a prospektivní studie $7 D011446
650    _2
$a plnění a analýza úkolů $7 D013647
650    _2
$a časové faktory $7 D013997
655    _2
$a srovnávací studie $7 D003160
655    _2
$a časopisecké články $7 D016428
655    _2
$a multicentrická studie $7 D016448
655    _2
$a pozorovací studie $7 D064888
700    1_
$a Beurskens, Niek E G $u Department of Clinical and Experimental Cardiology, Amsterdam UMC, location AMC, University of Amsterdam, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands.
700    1_
$a Neuzil, Petr $u Na Homolce Hospital, Praha, Czech Republic.
700    1_
$a Defaye, Pascal $u CHRU Albert Michallon, Grenoble, France.
700    1_
$a Delnoy, Peter-Paul $u Isala, Zwolle, The Netherlands.
700    1_
$a Ip, John $u Sparrow Research Institute, Lansing, MI, USA.
700    1_
$a Guerrero, Juan Jose Garcia $u Hospital Universitario Infanta Cristina, Badajoz, Spain.
700    1_
$a Rashtian, Mayer $u Huntington Memorial Hospital, Pasadena, CA, USA.
700    1_
$a Banker, Rajesh $u Premier Cardiology, Inc., Newport Beach, CA, USA.
700    1_
$a Reddy, Vivek $u Mount Sinai Hospital, New York, NY, USA.
700    1_
$a Exner, Derek $u Foothills Medical Centre, Calgary, AB, USA.
700    1_
$a Sperzel, Johannes $u Kerckhoff-Klinik gGmbH, Bad Nauheim, Germany.
700    1_
$a Knops, Reinoud E $u Department of Clinical and Experimental Cardiology, Amsterdam UMC, location AMC, University of Amsterdam, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands.
700    1_
$a ,
700    1_
$a ,
773    0_
$w MED00008134 $t Journal of interventional cardiac electrophysiology an international journal of arrhythmias and pacing $x 1572-8595 $g Roč. 53, č. 2 (2018), s. 239-247
856    41
$u https://pubmed.ncbi.nlm.nih.gov/30105428 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20190405 $b ABA008
991    __
$a 20190415133915 $b ABA008
999    __
$a ok $b bmc $g 1391775 $s 1050770
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2018 $b 53 $c 2 $d 239-247 $e 20180813 $i 1572-8595 $m Journal of interventional cardiac electrophysiology $n J Interv Card Electrophysiol $x MED00008134
LZP    __
$a Pubmed-20190405

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...