Atmospheric pressure chemical ionization mass spectrometry at low flow rates: Importance of ion source housing
T. Strmeň, V. Vrkoslav, Z. Bosáková, J. Cvačka,
Jazyk angličtina Země Velká Británie
Typ dokumentu časopisecké články
Grantová podpora
Project SVV260440
Charles University in Prague
16-01639S
Czech Science Foundation
PubMed
31912928
DOI
10.1002/rcm.8722
Knihovny.cz E-zdroje
- Publikační typ
- časopisecké články MeSH
RATIONALE: Hyphenation of atmospheric pressure chemical ionization (APCI) mass spectrometry with capillary and micro high-performance liquid chromatography (HPLC) is attractive for many applications, but reliable ion sources dedicated to these conditions are still missing. There are a number of aspects to consider when designing such an ion source, including the susceptibility of the ionization processes to ambient conditions. Here we discuss the importance of ion source housing for APCI at low flow rates. METHODS: Selected compounds dissolved in various solvents were used to study ionization reactions at 10 μL/min flow rate. APCI spectra were generated using the Ion Max-S source (Thermo Fisher Scientific) operated with or without the ion source housing. RESULTS: The APCI spectra of most compounds measured in the open and enclosed ion sources were markedly different. The differences were explained by water and oxygen molecules that entered the plasma region of the open ion source. Water tended to suppress charge transfer processes while oxygen diminished electron capture reactions and prevented the formation of acetonitrile-related radical cations useful for localizing double bonds in lipids. The effects associated with the ion source housing were significantly less important for compounds that are easy to protonate or deprotonate. CONCLUSIONS: The use of ion source housing prevented alternative ionization channels leading to unwanted or unexpected ions. Compared with the conventional flow rate mode (1 mL/min), the effects of ambient air components were significantly higher at 10 μL/min, emphasizing the need for ion source housing in APCI sources dedicated to low flow rates.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc20005409
- 003
- CZ-PrNML
- 005
- 20200521085818.0
- 007
- ta
- 008
- 200511s2020 xxk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1002/rcm.8722 $2 doi
- 035 __
- $a (PubMed)31912928
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxk
- 100 1_
- $a Strmeň, Timotej $u Institute of Organic Chemistry and Biochemistry of the Czech Academy of Sciences, Flemingovo nám. 2, CZ-166 10, Prague 6, Czech Republic. Department of Analytical Chemistry, Faculty of Science, Charles University in Prague, Hlavova 2030/8, CZ-128 43, Prague 2, Czech Republic.
- 245 10
- $a Atmospheric pressure chemical ionization mass spectrometry at low flow rates: Importance of ion source housing / $c T. Strmeň, V. Vrkoslav, Z. Bosáková, J. Cvačka,
- 520 9_
- $a RATIONALE: Hyphenation of atmospheric pressure chemical ionization (APCI) mass spectrometry with capillary and micro high-performance liquid chromatography (HPLC) is attractive for many applications, but reliable ion sources dedicated to these conditions are still missing. There are a number of aspects to consider when designing such an ion source, including the susceptibility of the ionization processes to ambient conditions. Here we discuss the importance of ion source housing for APCI at low flow rates. METHODS: Selected compounds dissolved in various solvents were used to study ionization reactions at 10 μL/min flow rate. APCI spectra were generated using the Ion Max-S source (Thermo Fisher Scientific) operated with or without the ion source housing. RESULTS: The APCI spectra of most compounds measured in the open and enclosed ion sources were markedly different. The differences were explained by water and oxygen molecules that entered the plasma region of the open ion source. Water tended to suppress charge transfer processes while oxygen diminished electron capture reactions and prevented the formation of acetonitrile-related radical cations useful for localizing double bonds in lipids. The effects associated with the ion source housing were significantly less important for compounds that are easy to protonate or deprotonate. CONCLUSIONS: The use of ion source housing prevented alternative ionization channels leading to unwanted or unexpected ions. Compared with the conventional flow rate mode (1 mL/min), the effects of ambient air components were significantly higher at 10 μL/min, emphasizing the need for ion source housing in APCI sources dedicated to low flow rates.
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Vrkoslav, Vladimír $u Institute of Organic Chemistry and Biochemistry of the Czech Academy of Sciences, Flemingovo nám. 2, CZ-166 10, Prague 6, Czech Republic.
- 700 1_
- $a Bosáková, Zuzana $u Department of Analytical Chemistry, Faculty of Science, Charles University in Prague, Hlavova 2030/8, CZ-128 43, Prague 2, Czech Republic.
- 700 1_
- $a Cvačka, Josef $u Institute of Organic Chemistry and Biochemistry of the Czech Academy of Sciences, Flemingovo nám. 2, CZ-166 10, Prague 6, Czech Republic. Department of Analytical Chemistry, Faculty of Science, Charles University in Prague, Hlavova 2030/8, CZ-128 43, Prague 2, Czech Republic.
- 773 0_
- $w MED00004050 $t Rapid communications in mass spectrometry $x 1097-0231 $g Roč. 34, č. 10 (2020), s. e8722
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/31912928 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20200511 $b ABA008
- 991 __
- $a 20200521085815 $b ABA008
- 999 __
- $a ind $b bmc $g 1524339 $s 1095464
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2020 $b 34 $c 10 $d e8722 $e 2020May30 $i 1097-0231 $m Rapid communications in mass spectrometry $n Rapid Commun Mass Spectrom $x MED00004050
- GRA __
- $a Project SVV260440 $p Charles University in Prague
- GRA __
- $a 16-01639S $p Czech Science Foundation
- LZP __
- $a Pubmed-20200511