Interlocked Pins Increase Strength by a Lateral Spread of Load in Femoral Neck Fixation: a Cadaver Study [Zajištěné hřeby zvyšují tuhost fixace zlomenin krčku femuru vlivem laterálního rozložení zatížení: studie na kadáveru]
J. E. Brattgjerd, S. Niratisairak, H. Steen, K. Strømsøe
Jazyk angličtina Země Česko
Typ dokumentu časopisecké články
- MeSH
- biomechanika MeSH
- kostní destičky MeSH
- kostní hřeby MeSH
- kostní šrouby * MeSH
- krček femuru * chirurgie MeSH
- lidé MeSH
- mrtvola MeSH
- vnitřní fixace fraktury MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
PURPOSE OF THE STUDY To improve the important torsional, bending and compressive stability in femoral neck fixation, locking plates have been the latest contribution. However, increased strength by restricted fracture motion may come at expense of an altered load distribution and failure patterns. Within locking plate technology, the important intermediate fracture compression may principally be achieved by multiple sliding screws passing through a sideplate fixed to the femur or connected to an interlocking plate not fixed to the femur laterally, sliding "en bloc" with the plate. While biomechanical studies may deliver the short-time patient safety requirements in implant development, no adequate failure evaluation has been performed with interlocking devices ex vivo in this setting. In the present biomechanical study, we analysed if a novel femoral neck interlocking plate with pins could improve fixation performance by changing the parameters involved in the failure mechanism in terms of fixation strength, fracture motion, load distribution and failure pattern. MATERIAL AND METHODS Sixteen pairs of human femurs with stable subcapital osteotomies were fixated by 2 pins or 3 pins interlocked in a plate using a paired design. Femurs were loaded non-destructively to 10° torsion around the neck axis, 200 N anteroposterior bending and 500 N vertical compression in 7° adduction with 1 Hz in 20 000 cycles, and were subsequently subjected to destructive compression to evaluate failure patterns. Bending stiffness, compressive stiffness and displacement from compressive testing reflected fracture motion. Torque and compression to failure replicated known failure mechanisms and defined strength. To evaluate load distribution, associations between biomechanical parameters and measured local bone mineral measurements by quantitative CT were analysed. RESULTS Interlocked pins increased mean strength 73% in torsion and 39% in compression (p = 0.038). Strength was related to all 4 regional mineral masses from the femoral head to subtrochanterically with interlocking (r = 0.64-0.83, p = 0.034), while only to mineral masses in the femoral head in compression and to the head, neck and trochanterically in torsion with individual pins (r = 0.67-0.78, p = 0.024). No difference was detected in fracture motion or failure pattern. DISCUSSION Within the last decade, angular stable implants have expanded our therapeutic arsenal of femoral neck fractures. Increased stability at the expense of altered devastating failure patterns was not retrieved in our study. The broadened understanding of the effect of interlocking pins by an isolated plate as in the current study involved the feature to gain fixation strength. By permitting fracture compression, and through a significant change of correlations between mechanical parameters and local bone mineral factors, a lateral redistribution of load with interlocked pins from the fragile bone medially to the more solid lateral bone was demonstrated. Regarding the long-term patient safety of interlocked pins and healing complications of non-union and segmental collapse of the femoral head, a definite conclusion may be premature. However, the improved biomechanics of an interlocking plate must be considered a favourable development of the pin concept. CONCLUSIONS Interlocked pins may improve fixation performance by a better load distribution, not by restricting fracture motion with corresponding altered failure patterns. This is encouraging and a challenge to complete further studies of the interlocking plate technology in the struggle to find the optimal treatment of the femoral neck fracture. Key words: femoral neck fracture, biomechanics, cadaver bone, bone mineral, internal fixation, locking plate, interlocked pins.
Biomechanics Lab Division of Orthopaedic Surgery Oslo University Hospital Norway
Institute of Clinical Medicine Faculty of Medicine University of Oslo Norway
Zajištěné hřeby zvyšují tuhost fixace zlomenin krčku femuru vlivem laterálního rozložení zatížení: studie na kadáveru
Literatura
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21015614
- 003
- CZ-PrNML
- 005
- 20210617150909.0
- 007
- ta
- 008
- 210604s2021 xr f 000 0|eng||
- 009
- AR
- 024 7_
- $2 doi $a 10.55095/achot2021/024
- 035 __
- $a (PubMed)33960928
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Brattgjerd, J. E. $u Biomechanics Lab, Division of Orthopaedic Surgery, Oslo University Hospital, Norway; Institute of Clinical Medicine, Faculty of Medicine, University of Oslo, Norway
- 245 10
- $a Interlocked Pins Increase Strength by a Lateral Spread of Load in Femoral Neck Fixation: a Cadaver Study / $c J. E. Brattgjerd, S. Niratisairak, H. Steen, K. Strømsøe
- 246 31
- $a Zajištěné hřeby zvyšují tuhost fixace zlomenin krčku femuru vlivem laterálního rozložení zatížení: studie na kadáveru
- 504 __
- $a Literatura
- 520 9_
- $a PURPOSE OF THE STUDY To improve the important torsional, bending and compressive stability in femoral neck fixation, locking plates have been the latest contribution. However, increased strength by restricted fracture motion may come at expense of an altered load distribution and failure patterns. Within locking plate technology, the important intermediate fracture compression may principally be achieved by multiple sliding screws passing through a sideplate fixed to the femur or connected to an interlocking plate not fixed to the femur laterally, sliding "en bloc" with the plate. While biomechanical studies may deliver the short-time patient safety requirements in implant development, no adequate failure evaluation has been performed with interlocking devices ex vivo in this setting. In the present biomechanical study, we analysed if a novel femoral neck interlocking plate with pins could improve fixation performance by changing the parameters involved in the failure mechanism in terms of fixation strength, fracture motion, load distribution and failure pattern. MATERIAL AND METHODS Sixteen pairs of human femurs with stable subcapital osteotomies were fixated by 2 pins or 3 pins interlocked in a plate using a paired design. Femurs were loaded non-destructively to 10° torsion around the neck axis, 200 N anteroposterior bending and 500 N vertical compression in 7° adduction with 1 Hz in 20 000 cycles, and were subsequently subjected to destructive compression to evaluate failure patterns. Bending stiffness, compressive stiffness and displacement from compressive testing reflected fracture motion. Torque and compression to failure replicated known failure mechanisms and defined strength. To evaluate load distribution, associations between biomechanical parameters and measured local bone mineral measurements by quantitative CT were analysed. RESULTS Interlocked pins increased mean strength 73% in torsion and 39% in compression (p = 0.038). Strength was related to all 4 regional mineral masses from the femoral head to subtrochanterically with interlocking (r = 0.64-0.83, p = 0.034), while only to mineral masses in the femoral head in compression and to the head, neck and trochanterically in torsion with individual pins (r = 0.67-0.78, p = 0.024). No difference was detected in fracture motion or failure pattern. DISCUSSION Within the last decade, angular stable implants have expanded our therapeutic arsenal of femoral neck fractures. Increased stability at the expense of altered devastating failure patterns was not retrieved in our study. The broadened understanding of the effect of interlocking pins by an isolated plate as in the current study involved the feature to gain fixation strength. By permitting fracture compression, and through a significant change of correlations between mechanical parameters and local bone mineral factors, a lateral redistribution of load with interlocked pins from the fragile bone medially to the more solid lateral bone was demonstrated. Regarding the long-term patient safety of interlocked pins and healing complications of non-union and segmental collapse of the femoral head, a definite conclusion may be premature. However, the improved biomechanics of an interlocking plate must be considered a favourable development of the pin concept. CONCLUSIONS Interlocked pins may improve fixation performance by a better load distribution, not by restricting fracture motion with corresponding altered failure patterns. This is encouraging and a challenge to complete further studies of the interlocking plate technology in the struggle to find the optimal treatment of the femoral neck fracture. Key words: femoral neck fracture, biomechanics, cadaver bone, bone mineral, internal fixation, locking plate, interlocked pins.
- 650 _2
- $a biomechanika $7 D001696
- 650 _2
- $a kostní hřeby $7 D001858
- 650 _2
- $a kostní destičky $7 D001860
- 650 12
- $a kostní šrouby $7 D001863
- 650 _2
- $a mrtvola $7 D002102
- 650 12
- $a krček femuru $x chirurgie $7 D005272
- 650 _2
- $a vnitřní fixace fraktury $7 D005593
- 650 _2
- $a lidé $7 D006801
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Niratisairak, S. $u Biomechanics Lab, Division of Orthopaedic Surgery, Oslo University Hospital, Norway; Institute of Clinical Medicine, Faculty of Medicine, University of Oslo, Norway
- 700 1_
- $a Steen, H. $u Biomechanics Lab, Division of Orthopaedic Surgery, Oslo University Hospital, Norway
- 700 1_
- $a Strømsøe, K. $u Institute of Clinical Medicine, Faculty of Medicine, University of Oslo, Norway
- 773 0_
- $w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Cechoslovaca $x 0001-5415 $g Roč. 88, č. 2 (2021), s. 144-152
- 856 41
- $u https://achot.cz/pdfs/ach/2021/02/10.pdf $y plný text volně přístupný
- 910 __
- $a ABA008 $b A 8 $c 507 $y p $z 0
- 990 __
- $a 20210604 $b ABA008
- 991 __
- $a 20210611151053 $b ABA008
- 999 __
- $a ok $b bmc $g 1663637 $s 1136022
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 88 $c 2 $d 144-152 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
- LZP __
- $b NLK118 $a Pubmed-20210604