Real-time measurement of ICD lead motion during stereotactic body radiotherapy of ventricular tachycardia
L. Knybel, J. Cvek, R. Neuwirth, O. Jiravsky, J. Hecko, M. Penhaker, M. Sramko, J. Kautzner
Jazyk angličtina Země Polsko
Typ dokumentu časopisecké články
NLK
Directory of Open Access Journals
od 2001
Free Medical Journals
od 2010
PubMed Central
od 2010
Europe PubMed Central
od 2010 do Před 1 rokem
ROAD: Directory of Open Access Scholarly Resources
od 1998
PubMed
34046223
DOI
10.5603/rpor.a2021.0020
Knihovny.cz E-zdroje
- Publikační typ
- časopisecké články MeSH
Background: Here we aimed to evaluate the respiratory and cardiac-induced motion of a ICD lead used as surrogate in the heart during stereotactic body radiotherapy (SBRT) of ventricular tachycardia (VT). Data provides insight regarding motion and motion variations during treatment. Materials and methods: We analyzed the log files of surrogate motion during SBRT of ventricular tachycardia performed in 20 patients. Evaluated parameters included the ICD lead motion amplitudes; intrafraction amplitude variability; correlation error between the ICD lead and external markers; and margin expansion in the superior-inferior (SI), latero-lateral (LL), and anterior-posterior (AP) directions to cover 90% or 95% of all amplitudes. Results: In the SI, LL, and AP directions, respectively, the mean motion amplitudes were 5.0 ± 2.6, 3.4. ± 1.9, and 3.1 ± 1.6 mm. The mean intrafraction amplitude variability was 2.6 ± 0.9, 1.9 ± 1.3, and 1.6 ± 0.8 mm in the SI, LL, and AP directions, respectively. The margins required to cover 95% of ICD lead motion amplitudes were 9.5, 6.7, and 5.5 mm in the SI, LL, and AP directions, respectively. The mean correlation error was 2.2 ± 0.9 mm. Conclusions: Data from online tracking indicated motion irregularities and correlation errors, necessitating an increased CTV-PTV margin of 3 mm. In 35% of cases, the motion variability exceeded 3 mm in one or more directions. We recommend verifying the correlation between CTV and surrogate individually for every patient, especially for targets with posterobasal localization where we observed the highest difference between the lead and CTV motion.
Department of Cardiology Podlesi Hospital Trinec Czech Republic
Department of Oncology University Hospital Ostrava Ostrava Czech Republic
Institute for Clinical and Experimental Medicine Prague Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21018246
- 003
- CZ-PrNML
- 005
- 20210729103945.0
- 007
- ta
- 008
- 210726s2021 pl f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.5603/RPOR.a2021.0020 $2 doi
- 035 __
- $a (PubMed)34046223
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a pl
- 100 1_
- $a Knybel, Lukas $u Department of Oncology, University Hospital Ostrava, Ostrava, Czech Republic
- 245 10
- $a Real-time measurement of ICD lead motion during stereotactic body radiotherapy of ventricular tachycardia / $c L. Knybel, J. Cvek, R. Neuwirth, O. Jiravsky, J. Hecko, M. Penhaker, M. Sramko, J. Kautzner
- 520 9_
- $a Background: Here we aimed to evaluate the respiratory and cardiac-induced motion of a ICD lead used as surrogate in the heart during stereotactic body radiotherapy (SBRT) of ventricular tachycardia (VT). Data provides insight regarding motion and motion variations during treatment. Materials and methods: We analyzed the log files of surrogate motion during SBRT of ventricular tachycardia performed in 20 patients. Evaluated parameters included the ICD lead motion amplitudes; intrafraction amplitude variability; correlation error between the ICD lead and external markers; and margin expansion in the superior-inferior (SI), latero-lateral (LL), and anterior-posterior (AP) directions to cover 90% or 95% of all amplitudes. Results: In the SI, LL, and AP directions, respectively, the mean motion amplitudes were 5.0 ± 2.6, 3.4. ± 1.9, and 3.1 ± 1.6 mm. The mean intrafraction amplitude variability was 2.6 ± 0.9, 1.9 ± 1.3, and 1.6 ± 0.8 mm in the SI, LL, and AP directions, respectively. The margins required to cover 95% of ICD lead motion amplitudes were 9.5, 6.7, and 5.5 mm in the SI, LL, and AP directions, respectively. The mean correlation error was 2.2 ± 0.9 mm. Conclusions: Data from online tracking indicated motion irregularities and correlation errors, necessitating an increased CTV-PTV margin of 3 mm. In 35% of cases, the motion variability exceeded 3 mm in one or more directions. We recommend verifying the correlation between CTV and surrogate individually for every patient, especially for targets with posterobasal localization where we observed the highest difference between the lead and CTV motion.
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Cvek, Jakub $u Department of Oncology, University Hospital Ostrava, Ostrava, Czech Republic
- 700 1_
- $a Neuwirth, Radek $u MUNI-Masaryk University, Brno, Czech Republic
- 700 1_
- $a Jiravsky, Otakar $u Department of Cardiology, Podlesi Hospital, Trinec, Czech Republic
- 700 1_
- $a Hecko, Jan $u VŠB-Technical University of Ostrava, Ostrava, Czech Republic
- 700 1_
- $a Penhaker, Marek $u VŠB-Technical University of Ostrava, Ostrava, Czech Republic
- 700 1_
- $a Sramko, Marek $u Institute for Clinical and Experimental Medicine, Prague, Czech Republic
- 700 1_
- $a Kautzner, Josef $u Institute for Clinical and Experimental Medicine, Prague, Czech Republic
- 773 0_
- $w MED00154188 $t Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology $x 1507-1367 $g Roč. 26, č. 1 (2021), s. 128-137
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/34046223 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20210726 $b ABA008
- 991 __
- $a 20210729103944 $b ABA008
- 999 __
- $a ind $b bmc $g 1676590 $s 1138690
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 26 $c 1 $d 128-137 $e 20210225 $i 1507-1367 $m Reports of Practical Oncology and Radiotherapy $n Rep Pract Oncol Radiother $x MED00154188
- LZP __
- $a Pubmed-20210726