The effect of albumin and γ-globulin on synovial fluid lubrication: Implication for knee joint replacements
D. Nečas, K. Sadecká, M. Vrbka, A. Galandáková, MA. Wimmer, J. Gallo, M. Hartl
Jazyk angličtina Země Nizozemsko
Typ dokumentu časopisecké články, práce podpořená grantem
- MeSH
- albuminy MeSH
- gama-globuliny MeSH
- kolenní kloub MeSH
- lidé MeSH
- lubrikace MeSH
- synoviální tekutina * MeSH
- totální endoprotéza kolene * MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
Total knee arthroplasty has become a routine procedure for patients suffering from joint diseases. Although the number of operations continuously increases, a limited service-life of implants represents a persisting challenge for scientists. Understanding of lubrication may help to suitably explain tribological processes on the way to replacements that become durable well into the third decade of service. The aim of the present study is to assess the formation of protein lubricating film in the knee implant. A developed knee simulator was used to observe the contact of real femoral and transparent polymer tibial component using fluorescent microscopy. The contact was lubricated by various protein solutions with attention to the behaviour of albumin and γ-globulin. In order to suitably mimic a human synovial fluid, hyaluronic acid and phospholipids were subsequently added to the solutions. Further, the change in shape and the migration of the contact zone were studied. The results showed considerable appearance differences of the contact over the swing phase of the simplified gait cycle. Regarding film formation, a strong interaction of the various molecules of synovial fluid was observed. It was found that the thickness of the lubricating layer stabilizes within around 50 s. Throughout the contact zone, protein agglomerations were present and could be clearly visualised using the applied optical technique.
Dept of Orthopedic Surgery Rush University Medical Center Chicago IL USA
Dept of Tribology Faculty of Mechanical Engineering Brno University of Technology Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21019560
- 003
- CZ-PrNML
- 005
- 20210830101144.0
- 007
- ta
- 008
- 210728s2021 ne f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.jmbbm.2020.104117 $2 doi
- 035 __
- $a (PubMed)33065468
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Nečas, D $u Dept of Tribology, Faculty of Mechanical Engineering, Brno University of Technology, Czech Republic. Electronic address: david.necas@vut.cz
- 245 14
- $a The effect of albumin and γ-globulin on synovial fluid lubrication: Implication for knee joint replacements / $c D. Nečas, K. Sadecká, M. Vrbka, A. Galandáková, MA. Wimmer, J. Gallo, M. Hartl
- 520 9_
- $a Total knee arthroplasty has become a routine procedure for patients suffering from joint diseases. Although the number of operations continuously increases, a limited service-life of implants represents a persisting challenge for scientists. Understanding of lubrication may help to suitably explain tribological processes on the way to replacements that become durable well into the third decade of service. The aim of the present study is to assess the formation of protein lubricating film in the knee implant. A developed knee simulator was used to observe the contact of real femoral and transparent polymer tibial component using fluorescent microscopy. The contact was lubricated by various protein solutions with attention to the behaviour of albumin and γ-globulin. In order to suitably mimic a human synovial fluid, hyaluronic acid and phospholipids were subsequently added to the solutions. Further, the change in shape and the migration of the contact zone were studied. The results showed considerable appearance differences of the contact over the swing phase of the simplified gait cycle. Regarding film formation, a strong interaction of the various molecules of synovial fluid was observed. It was found that the thickness of the lubricating layer stabilizes within around 50 s. Throughout the contact zone, protein agglomerations were present and could be clearly visualised using the applied optical technique.
- 650 _2
- $a albuminy $7 D000418
- 650 12
- $a totální endoprotéza kolene $7 D019645
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a kolenní kloub $7 D007719
- 650 _2
- $a lubrikace $7 D008153
- 650 12
- $a synoviální tekutina $7 D013582
- 650 _2
- $a gama-globuliny $7 D005719
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Sadecká, K $u Dept of Tribology, Faculty of Mechanical Engineering, Brno University of Technology, Czech Republic
- 700 1_
- $a Vrbka, M $u Dept of Tribology, Faculty of Mechanical Engineering, Brno University of Technology, Czech Republic
- 700 1_
- $a Galandáková, A $u Dept of Medical Chemistry and Biochemistry, Faculty of Medicine and Dentistry, Palacky University Olomouc, Czech Republic
- 700 1_
- $a Wimmer, M A $u Dept of Orthopedic Surgery, Rush University Medical Center, Chicago, IL, USA
- 700 1_
- $a Gallo, J $u Dept of Orthopaedics, Faculty of Medicine and Dentistry, Palacky University Olomouc, University Hospital Olomouc, Czech Republic
- 700 1_
- $a Hartl, M $u Dept of Tribology, Faculty of Mechanical Engineering, Brno University of Technology, Czech Republic
- 773 0_
- $w MED00166961 $t Journal of the mechanical behavior of biomedical materials $x 1878-0180 $g Roč. 113, č. - (2021), s. 104117
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/33065468 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20210728 $b ABA008
- 991 __
- $a 20210830101145 $b ABA008
- 999 __
- $a ok $b bmc $g 1690401 $s 1140006
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 113 $c - $d 104117 $e 20200928 $i 1878-0180 $m Journal of the mechanical behavior of biomedical materials $n J Mech Behav Biomed Mater $x MED00166961
- LZP __
- $a Pubmed-20210728