Je něco špatně v tomto záznamu ?
Reverse Total Shoulder Arthroplasty: 14-Year Clinical Experience with 496 Performed Arthroplasties [Reverzní totální náhrada ramenního kloubu: 14 let klinických zkušeností s 496 provedenými artroplastikami]
D. Pokorný, P. Fulín, J. Heřt, I. Landor, J. Štefan, A. Sosna
Jazyk angličtina Země Česko
Typ dokumentu časopisecké články
- MeSH
- artroplastika ramenního kloubu * škodlivé účinky MeSH
- lidé MeSH
- následné studie MeSH
- prospektivní studie MeSH
- ramenní kloub * chirurgie MeSH
- retrospektivní studie MeSH
- rozsah kloubních pohybů MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
PURPOSE OF THE STUDY Reverse total shoulder arthroplasty (RSA) can be considered a proven method for the treatment of the most severe shoulder joint affections. In this study, we present 14-year experiences of the authors with the LimaCorporate SMR® RSA system in 496 cases. MATERIAL AND METHODS Included in the study are 496 RSAs performed between 2007 and 2020. We successfully followed up 368 shoulders in 358 patients. This was a prospective study with function being evaluated preoperatively and at the last follow-up in 2020. We evaluated the range of motion in active elevation and the classic Constant Score (CS). According to this score, the pain level was also evaluated. Standard statistical methods were used with a paired t-test used for comparisons of values. RESULTS The mean follow-up in our group was 5.5 years (0.7-13.6, SD 3.22, median 4.96). Indications were: primary osteoarthritis (84), acute trauma (69), posttraumatic sequelae (79), cuff tear arthropathy (37), RA (29), chronic dislocations (18), final treatment of infectious complications (7), avascular necrosis (6), tumours (4) and TSA revisions (9). The mean post-operative CS of all patients was 71.9 (2-94, SD 11.26, median 73). The mean post-operative active elevation was 127.35° (30°-180°, SD 28.36, median 130°). The mean pain level at final follow-up was 0.65 (0-3, SD 0.65, median 1). There was a statistically significant improvement In the CS (26.9 to 71) and the final achieved elevation (48.5° to 127.35°) in all groups except acute traumas for obvious reasons. A significant decrease in pain (2.8 to 0.65) was observed in all groups. We saw no implant failures or UHMWPE component wear. DISCUSSION We compared our results with those published by other authors. When considering the functional outcomes, our results are comparable with those published previously. Compared to other studies, an interesting result is the low incidence of scapular notching. In our cohort of patients, it was only present in 10 (2.7%) cases. We attribute this to the design of the SMR® implant which meets the main criterium of a modern shoulder arthroplasty system: modularity of both glenoid components. The glenospheres are available in 36, 40 and 44mm sizes with standard and distalised options. Furthermore, the system also contains the Axioma® revision glenoid component. Humeral stems are available in a range of sizes and lengths including revision stems. The humeral body is also available in two sizes. The short variant is optimal for revision with hemiarthroplasty to reverse conversion. CONCLUSIONS During a period of 14 years, we performed 496 implantations of RSAs using the SMR® system. We evaluated 368 cases with an average follow-up of 5.5 years. During long-term follow-up we experienced no implant failures or complications indicative of any constructional insufficiencies. We saw no signs of UHMWPE liner wear. The SMR® system allows for treatment of even the most complex shoulder affections due to its high modularity. Key words: shoulder joint replacement, reverse shoulder arthroplasty, SMR®, Axioma®, outcomes, ROM, scapular notching, complication, polyethylene wear, UHMWPE.
Reverzní totální náhrada ramenního kloubu: 14 let klinických zkušeností s 496 provedenými artroplastikami
Citace poskytuje Crossref.org
Literatura
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22005929
- 003
- CZ-PrNML
- 005
- 20220224144217.0
- 007
- ta
- 008
- 220207s2021 xr d f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.55095/achot2021/060 $2 doi
- 035 __
- $a (PubMed)34998442
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Pokorný, David, $d 1968- $7 jn20001103568 $u 1st Department of Orthopaedics, First Faculty of Medicine, Charles University, University Hospital Motol, Prague, Czech Republic
- 245 10
- $a Reverse Total Shoulder Arthroplasty: 14-Year Clinical Experience with 496 Performed Arthroplasties / $c D. Pokorný, P. Fulín, J. Heřt, I. Landor, J. Štefan, A. Sosna
- 246 31
- $a Reverzní totální náhrada ramenního kloubu: 14 let klinických zkušeností s 496 provedenými artroplastikami
- 504 __
- $a Literatura
- 520 9_
- $a PURPOSE OF THE STUDY Reverse total shoulder arthroplasty (RSA) can be considered a proven method for the treatment of the most severe shoulder joint affections. In this study, we present 14-year experiences of the authors with the LimaCorporate SMR® RSA system in 496 cases. MATERIAL AND METHODS Included in the study are 496 RSAs performed between 2007 and 2020. We successfully followed up 368 shoulders in 358 patients. This was a prospective study with function being evaluated preoperatively and at the last follow-up in 2020. We evaluated the range of motion in active elevation and the classic Constant Score (CS). According to this score, the pain level was also evaluated. Standard statistical methods were used with a paired t-test used for comparisons of values. RESULTS The mean follow-up in our group was 5.5 years (0.7-13.6, SD 3.22, median 4.96). Indications were: primary osteoarthritis (84), acute trauma (69), posttraumatic sequelae (79), cuff tear arthropathy (37), RA (29), chronic dislocations (18), final treatment of infectious complications (7), avascular necrosis (6), tumours (4) and TSA revisions (9). The mean post-operative CS of all patients was 71.9 (2-94, SD 11.26, median 73). The mean post-operative active elevation was 127.35° (30°-180°, SD 28.36, median 130°). The mean pain level at final follow-up was 0.65 (0-3, SD 0.65, median 1). There was a statistically significant improvement In the CS (26.9 to 71) and the final achieved elevation (48.5° to 127.35°) in all groups except acute traumas for obvious reasons. A significant decrease in pain (2.8 to 0.65) was observed in all groups. We saw no implant failures or UHMWPE component wear. DISCUSSION We compared our results with those published by other authors. When considering the functional outcomes, our results are comparable with those published previously. Compared to other studies, an interesting result is the low incidence of scapular notching. In our cohort of patients, it was only present in 10 (2.7%) cases. We attribute this to the design of the SMR® implant which meets the main criterium of a modern shoulder arthroplasty system: modularity of both glenoid components. The glenospheres are available in 36, 40 and 44mm sizes with standard and distalised options. Furthermore, the system also contains the Axioma® revision glenoid component. Humeral stems are available in a range of sizes and lengths including revision stems. The humeral body is also available in two sizes. The short variant is optimal for revision with hemiarthroplasty to reverse conversion. CONCLUSIONS During a period of 14 years, we performed 496 implantations of RSAs using the SMR® system. We evaluated 368 cases with an average follow-up of 5.5 years. During long-term follow-up we experienced no implant failures or complications indicative of any constructional insufficiencies. We saw no signs of UHMWPE liner wear. The SMR® system allows for treatment of even the most complex shoulder affections due to its high modularity. Key words: shoulder joint replacement, reverse shoulder arthroplasty, SMR®, Axioma®, outcomes, ROM, scapular notching, complication, polyethylene wear, UHMWPE.
- 650 12
- $a artroplastika ramenního kloubu $x škodlivé účinky $7 D000072228
- 650 _2
- $a následné studie $7 D005500
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a prospektivní studie $7 D011446
- 650 _2
- $a rozsah kloubních pohybů $7 D016059
- 650 _2
- $a retrospektivní studie $7 D012189
- 650 12
- $a ramenní kloub $x chirurgie $7 D012785
- 650 _2
- $a výsledek terapie $7 D016896
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Fulín, Petr $7 xx0231618 $u 1st Department of Orthopaedics, First Faculty of Medicine, Charles University, University Hospital Motol, Prague, Czech Republic
- 700 1_
- $a Heřt, Jan, $d 1989- $7 xx0256941 $u 1st Department of Orthopaedics, First Faculty of Medicine, Charles University, University Hospital Motol, Prague, Czech Republic
- 700 1_
- $a Landor, Ivan, $d 1956- $7 xx0062301 $u 1st Department of Orthopaedics, First Faculty of Medicine, Charles University, University Hospital Motol, Prague, Czech Republic
- 700 1_
- $a Štefan, Jan $7 xx0246287 $u 1st Department of Orthopaedics, First Faculty of Medicine, Charles University, University Hospital Motol, Prague, Czech Republic
- 700 1_
- $a Sosna, Antonín, $d 1943- $7 jn20000710566 $u 1st Department of Orthopaedics, First Faculty of Medicine, Charles University, University Hospital Motol, Prague, Czech Republic
- 773 0_
- $w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Cechoslovaca $x 0001-5415 $g Roč. 88, č. 6 (2021), s. 401-411
- 856 41
- $u https://achot.cz/pdfs/ach/2020/02/02.pdf $y plný text volně přístupný
- 910 __
- $a ABA008 $b A 8 $c 507 $y p $z 0
- 990 __
- $a 20220207 $b ABA008
- 991 __
- $a 20220224144205 $b ABA008
- 999 __
- $a ok $b bmc $g 1763611 $s 1157081
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 88 $c 6 $d 401-411 $e 20211220 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
- LZP __
- $b NLK118 $a Pubmed-20220207