High-throughput salting-out assisted liquid-liquid extraction using a 3D printed device and its application in the quantification of ibrutinib and its metabolite PCI-45227 in human serum

D. Écsiová, M. Šimkovič, M. Uher, M. Hroch

. 2022 ; 219 (-) : 114923. [pub] 20220702

Jazyk angličtina Země Anglie, Velká Británie

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc22024373

Liquid-liquid extraction methods are widely used for sample treatment in bioanalysis, although their implementation poses a common speed-limiting step in the analytical process when fast separation and detection methods such as UHPLC-MS are used. This study aimed to develop high-throughput salting-out assisted liquid-liquid extraction on a 96-well plate in combination with fast LC-MS analysis of ibrutinib and its active metabolite PCI-45227 (dihydrodiol ibrutinib) in human serum. A specially designed 3D printed extraction device developed in our laboratory allowed for the precise and rapid collection of the organic phase from the 96-well plate using a multichannel pipette, without the risk of aspiration of the bottom aqueous layer. The application of this device significantly accelerated sample preparation and allowed the processing of up to 96 samples in 1 h. The method was successfully validated according to EMA guidelines in the concentration range of 0.1-200 ng/mL for both analytes, providing lower limit of quantification at 100 pg/mL. The intra-day accuracy for IBT was in the range of - 1.67-5.67 %, while the inter-day accuracy was in the range of 0.20-6.90 %. The intra-day precision for IBT was in the range of 3.20-4.37 % and 3.13-4.73 % for the inter-day measurement. PCI-45227 showed intra-day accuracy in the range of - 11.2-3.71 % and the inter-day accuracy in the range of - 5.76-0.92 %. The intra-day precision for PCI was in the range of 3.49-7.64 % and 3.63-8.61 % for the measurement between days. In addition to increasing the speed of sample preparation, this method also offers low consumption of the sample and extraction solvent and can be utilized for other similar small-volume in-well plate extractions where organic solvents of lower density than water are used. The method was successfully applied to the analysis of serum samples (n = 5) of patients with chronic lymphoblastic leukaemia at the trough level (ibrutinib concentration range: 1.63-3.78 ng/mL, PCI-45227 concentration range: 1.84-14.02 ng/mL) and 2 h postdose (ibrutinib concentration range: 7.34-89.0 ng/mL, PCI-45227 concentration range: 5.64-124 ng/mL).

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc22024373
003      
CZ-PrNML
005      
20221031100245.0
007      
ta
008      
221017s2022 enk f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.jpba.2022.114923 $2 doi
035    __
$a (PubMed)35830774
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Écsiová, Dominika $u 4th Department of Internal Medicine-Hematology, University Hospital, Sokolská 581, Hradec Králové 500 05, Czech Republic; Faculty of Medicine in Hradec Kralové, Charles University, Šimkova 870, Hradec Králové 500 03, Czech Republic
245    10
$a High-throughput salting-out assisted liquid-liquid extraction using a 3D printed device and its application in the quantification of ibrutinib and its metabolite PCI-45227 in human serum / $c D. Écsiová, M. Šimkovič, M. Uher, M. Hroch
520    9_
$a Liquid-liquid extraction methods are widely used for sample treatment in bioanalysis, although their implementation poses a common speed-limiting step in the analytical process when fast separation and detection methods such as UHPLC-MS are used. This study aimed to develop high-throughput salting-out assisted liquid-liquid extraction on a 96-well plate in combination with fast LC-MS analysis of ibrutinib and its active metabolite PCI-45227 (dihydrodiol ibrutinib) in human serum. A specially designed 3D printed extraction device developed in our laboratory allowed for the precise and rapid collection of the organic phase from the 96-well plate using a multichannel pipette, without the risk of aspiration of the bottom aqueous layer. The application of this device significantly accelerated sample preparation and allowed the processing of up to 96 samples in 1 h. The method was successfully validated according to EMA guidelines in the concentration range of 0.1-200 ng/mL for both analytes, providing lower limit of quantification at 100 pg/mL. The intra-day accuracy for IBT was in the range of - 1.67-5.67 %, while the inter-day accuracy was in the range of 0.20-6.90 %. The intra-day precision for IBT was in the range of 3.20-4.37 % and 3.13-4.73 % for the inter-day measurement. PCI-45227 showed intra-day accuracy in the range of - 11.2-3.71 % and the inter-day accuracy in the range of - 5.76-0.92 %. The intra-day precision for PCI was in the range of 3.49-7.64 % and 3.63-8.61 % for the measurement between days. In addition to increasing the speed of sample preparation, this method also offers low consumption of the sample and extraction solvent and can be utilized for other similar small-volume in-well plate extractions where organic solvents of lower density than water are used. The method was successfully applied to the analysis of serum samples (n = 5) of patients with chronic lymphoblastic leukaemia at the trough level (ibrutinib concentration range: 1.63-3.78 ng/mL, PCI-45227 concentration range: 1.84-14.02 ng/mL) and 2 h postdose (ibrutinib concentration range: 7.34-89.0 ng/mL, PCI-45227 concentration range: 5.64-124 ng/mL).
650    _2
$a adenin $x analogy a deriváty $7 D000225
650    _2
$a vysokoúčinná kapalinová chromatografie $x metody $7 D002851
650    _2
$a lidé $7 D006801
650    _2
$a extrakce kapalina-kapalina $x metody $7 D059625
650    12
$a koronární angioplastika $7 D062645
650    _2
$a piperidiny $7 D010880
650    _2
$a 3D tisk $7 D066330
650    _2
$a pyrazoly $7 D011720
650    _2
$a rozpouštědla $7 D012997
650    12
$a tandemová hmotnostní spektrometrie $x metody $7 D053719
655    _2
$a časopisecké články $7 D016428
700    1_
$a Šimkovič, Martin $u 4th Department of Internal Medicine-Hematology, University Hospital, Sokolská 581, Hradec Králové 500 05, Czech Republic; Faculty of Medicine in Hradec Kralové, Charles University, Šimkova 870, Hradec Králové 500 03, Czech Republic
700    1_
$a Uher, Martin $u Department of Medical Biochemistry, Faculty of Medicine in Hradec Králové, Charles University, Šimkova 870, Hradec Králové 500 03, Czech Republic
700    1_
$a Hroch, Miloš $u Department of Medical Biochemistry, Faculty of Medicine in Hradec Králové, Charles University, Šimkova 870, Hradec Králové 500 03, Czech Republic. Electronic address: hrochm@lfhk.cuni.cz
773    0_
$w MED00002894 $t Journal of pharmaceutical and biomedical analysis $x 1873-264X $g Roč. 219, č. - (2022), s. 114923
856    41
$u https://pubmed.ncbi.nlm.nih.gov/35830774 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y p $z 0
990    __
$a 20221017 $b ABA008
991    __
$a 20221031100243 $b ABA008
999    __
$a ok $b bmc $g 1854218 $s 1175663
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2022 $b 219 $c - $d 114923 $e 20220702 $i 1873-264X $m Journal of pharmaceutical and biomedical analysis $n J Pharm Biomed Anal $x MED00002894
LZP    __
$a Pubmed-20221017

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...