Localization of the ventricular pacing site from BSPM and standard 12-lead ECG: a comparison study
KA. Sedova, PM. van Dam, M. Blahova, L. Necasova, J. Kautzner
Language English Country England, Great Britain
Document type Journal Article, Research Support, Non-U.S. Gov't
NLK
Directory of Open Access Journals
from 2011
Free Medical Journals
from 2011
Nature Open Access
from 2011-12-01
PubMed Central
from 2011
Europe PubMed Central
from 2011
ProQuest Central
from 2011-01-01
Open Access Digital Library
from 2011-01-01
Open Access Digital Library
from 2011-01-01
Health & Medicine (ProQuest)
from 2011-01-01
ROAD: Directory of Open Access Scholarly Resources
from 2011
Springer Nature OA/Free Journals
from 2011-12-01
- MeSH
- Cardiomyopathy, Dilated * MeSH
- Electrodes MeSH
- Electrocardiography MeSH
- Humans MeSH
- Cardiac Resynchronization Therapy Devices MeSH
- Cardiac Resynchronization Therapy * MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
Inverse ECG imaging methods typically require 32-250 leads to create body surface potential maps (BSPM), limiting their routine clinical use. This study evaluated the accuracy of PaceView inverse ECG method to localize the left or right ventricular (LV and RV, respectively) pacing leads using either a 99-lead BSPM or the 12-lead ECG. A 99-lead BSPM was recorded in patients with cardiac resynchronization therapy (CRT) during sinus rhythm and sequential LV/RV pacing. The non-contrast CT was performed to localize precisely both ECG electrodes and CRT leads. From a BSPM, nine signals were selected to obtain the 12-lead ECG. Both BSPM and 12-lead ECG were used to localize the RV and LV lead, and the localization error was calculated. Consecutive patients with dilated cardiomyopathy, previously implanted with a CRT device, were enrolled (n = 19). The localization error for the RV/LV lead was 9.0 [IQR 4.8-13.6] / 7.7 [IQR 0.0-10.3] mm using the 12-lead ECG and 9.1 [IQR 5.4-15.7] / 9.8 [IQR 8.6-13.1] mm for the BSPM. Thus, the noninvasive lead localization using the 12-lead ECG was accurate enough and comparable to 99-lead BSPM, potentially increasing the capability of 12-lead ECG for the optimization of the LV/RV pacing sites during CRT implant or for the most favorable programming.
Department of Cardiology Institute for Clinical and Experimental Medicine Prague Czech Republic
Department of Cardiology University Medical Center Utrecht Utrecht The Netherlands
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc23011052
- 003
- CZ-PrNML
- 005
- 20230801132800.0
- 007
- ta
- 008
- 230718s2023 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1038/s41598-023-36768-z $2 doi
- 035 __
- $a (PubMed)37316547
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Sedova, Ksenia A $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Sitna Sq. 3105, 27201, Kladno, Czech Republic. ksenia.sedova@fbmi.cvut.cz
- 245 10
- $a Localization of the ventricular pacing site from BSPM and standard 12-lead ECG: a comparison study / $c KA. Sedova, PM. van Dam, M. Blahova, L. Necasova, J. Kautzner
- 520 9_
- $a Inverse ECG imaging methods typically require 32-250 leads to create body surface potential maps (BSPM), limiting their routine clinical use. This study evaluated the accuracy of PaceView inverse ECG method to localize the left or right ventricular (LV and RV, respectively) pacing leads using either a 99-lead BSPM or the 12-lead ECG. A 99-lead BSPM was recorded in patients with cardiac resynchronization therapy (CRT) during sinus rhythm and sequential LV/RV pacing. The non-contrast CT was performed to localize precisely both ECG electrodes and CRT leads. From a BSPM, nine signals were selected to obtain the 12-lead ECG. Both BSPM and 12-lead ECG were used to localize the RV and LV lead, and the localization error was calculated. Consecutive patients with dilated cardiomyopathy, previously implanted with a CRT device, were enrolled (n = 19). The localization error for the RV/LV lead was 9.0 [IQR 4.8-13.6] / 7.7 [IQR 0.0-10.3] mm using the 12-lead ECG and 9.1 [IQR 5.4-15.7] / 9.8 [IQR 8.6-13.1] mm for the BSPM. Thus, the noninvasive lead localization using the 12-lead ECG was accurate enough and comparable to 99-lead BSPM, potentially increasing the capability of 12-lead ECG for the optimization of the LV/RV pacing sites during CRT implant or for the most favorable programming.
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a srdeční resynchronizační terapie $7 D058406
- 650 _2
- $a prostředky srdeční resynchronizační terapie $7 D058409
- 650 12
- $a dilatační kardiomyopatie $7 D002311
- 650 _2
- $a elektrokardiografie $7 D004562
- 650 _2
- $a elektrody $7 D004566
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a van Dam, Peter M $u Department of Cardiology, University Medical Center Utrecht, Utrecht, The Netherlands
- 700 1_
- $a Blahova, Marie $u Department of Cardiology, Institute for Clinical and Experimental Medicine, Prague, Czech Republic
- 700 1_
- $a Necasova, Lucie $u Department of Cardiology, Institute for Clinical and Experimental Medicine, Prague, Czech Republic
- 700 1_
- $a Kautzner, Josef $u Department of Cardiology, Institute for Clinical and Experimental Medicine, Prague, Czech Republic
- 773 0_
- $w MED00182195 $t Scientific reports $x 2045-2322 $g Roč. 13, č. 1 (2023), s. 9618
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/37316547 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20230718 $b ABA008
- 991 __
- $a 20230801132757 $b ABA008
- 999 __
- $a ok $b bmc $g 1963454 $s 1197317
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2023 $b 13 $c 1 $d 9618 $e 20230614 $i 2045-2322 $m Scientific reports $n Sci Rep $x MED00182195
- LZP __
- $a Pubmed-20230718