Thermal neutron detection and track recognition method in reference and out-of-field radiotherapy FLASH electron fields using Timepix3 detectors

C. Oancea, J. Solc, A. Bourgouin, C. Granja, J. Jakubek, J. Pivec, F. Riemer, Z. Vykydal, S. Worm, L. Marek

. 2023 ; 68 (18) : . [pub] 20230914

Jazyk angličtina Země Anglie, Velká Británie

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/bmc23016237

Objective.This work presents a method for enhanced detection, imaging, and measurement of the thermal neutron flux.Approach. Measurements were performed in a water tank, while the detector is positioned out-of-field of a 20 MeV ultra-high pulse dose rate electron beam. A semiconductor pixel detector Timepix3 with a silicon sensor partially covered by a6LiF neutron converter was used to measure the flux, spatial, and time characteristics of the neutron field. To provide absolute measurements of thermal neutron flux, the detection efficiency calibration of the detectors was performed in a reference thermal neutron field. Neutron signals are recognized and discriminated against other particles such as gamma rays and x-rays. This is achieved by the resolving power of the pixel detector using machine learning algorithms and high-resolution pattern recognition analysis of the high-energy tracks created by thermal neutron interactions in the converter.Main results. The resulting thermal neutrons equivalent dose was obtained using conversion factor (2.13(10) pSv·cm2) from thermal neutron fluence to thermal neutron equivalent dose obtained by Monte Carlo simulations. The calibrated detectors were used to characterize scattered radiation created by electron beams. The results at 12.0 cm depth in the beam axis inside of the water for a delivered dose per pulse of 1.85 Gy (pulse length of 2.4μs) at the reference depth, showed a contribution of flux of 4.07(8) × 103particles·cm-2·s-1and equivalent dose of 1.73(3) nSv per pulse, which is lower by ∼9 orders of magnitude than the delivered dose.Significance. The presented methodology for in-water measurements and identification of characteristic thermal neutrons tracks serves for the selective quantification of equivalent dose made by thermal neutrons in out-of-field particle therapy.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc23016237
003      
CZ-PrNML
005      
20231026110119.0
007      
ta
008      
231013s2023 enk f 000 0|eng||
009      
AR
024    7_
$a 10.1088/1361-6560/acf2e1 $2 doi
035    __
$a (PubMed)37607560
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Oancea, Cristina $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic $u University of Bucharest, Bucharest, Romania
245    10
$a Thermal neutron detection and track recognition method in reference and out-of-field radiotherapy FLASH electron fields using Timepix3 detectors / $c C. Oancea, J. Solc, A. Bourgouin, C. Granja, J. Jakubek, J. Pivec, F. Riemer, Z. Vykydal, S. Worm, L. Marek
520    9_
$a Objective.This work presents a method for enhanced detection, imaging, and measurement of the thermal neutron flux.Approach. Measurements were performed in a water tank, while the detector is positioned out-of-field of a 20 MeV ultra-high pulse dose rate electron beam. A semiconductor pixel detector Timepix3 with a silicon sensor partially covered by a6LiF neutron converter was used to measure the flux, spatial, and time characteristics of the neutron field. To provide absolute measurements of thermal neutron flux, the detection efficiency calibration of the detectors was performed in a reference thermal neutron field. Neutron signals are recognized and discriminated against other particles such as gamma rays and x-rays. This is achieved by the resolving power of the pixel detector using machine learning algorithms and high-resolution pattern recognition analysis of the high-energy tracks created by thermal neutron interactions in the converter.Main results. The resulting thermal neutrons equivalent dose was obtained using conversion factor (2.13(10) pSv·cm2) from thermal neutron fluence to thermal neutron equivalent dose obtained by Monte Carlo simulations. The calibrated detectors were used to characterize scattered radiation created by electron beams. The results at 12.0 cm depth in the beam axis inside of the water for a delivered dose per pulse of 1.85 Gy (pulse length of 2.4μs) at the reference depth, showed a contribution of flux of 4.07(8) × 103particles·cm-2·s-1and equivalent dose of 1.73(3) nSv per pulse, which is lower by ∼9 orders of magnitude than the delivered dose.Significance. The presented methodology for in-water measurements and identification of characteristic thermal neutrons tracks serves for the selective quantification of equivalent dose made by thermal neutrons in out-of-field particle therapy.
650    12
$a elektrony $7 D004583
650    12
$a algoritmy $7 D000465
650    _2
$a kalibrace $7 D002138
650    _2
$a záření gama $7 D005720
650    _2
$a neutrony $7 D009502
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Solc, Jaroslav $u Czech Metrology Institute, Okruzni 31, 638 00 Brno, Czech Republic
700    1_
$a Bourgouin, Alexandra $u Dosimetry for Radiation Therapy and Diagnostic Radiology, Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, 38116, Germany $1 https://orcid.org/0000000164294122
700    1_
$a Granja, Carlos $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic $1 https://orcid.org/0000000243981553
700    1_
$a Jakubek, Jan $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic
700    1_
$a Pivec, Jiri $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic
700    1_
$a Riemer, Felix $u Deutsches Elektronen-Synchrotron DESY, Platanenallee 6, 15738 Zeuthen, Germany
700    1_
$a Vykydal, Zdenek $u Czech Metrology Institute, Okruzni 31, 638 00 Brno, Czech Republic
700    1_
$a Worm, Steven $u Deutsches Elektronen-Synchrotron DESY, Platanenallee 6, 15738 Zeuthen, Germany
700    1_
$a Marek, Lukas $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic $u Faculty of Mathematics and Physics, Charles University, Prague, Czech Republic
773    0_
$w MED00003821 $t Physics in medicine and biology $x 1361-6560 $g Roč. 68, č. 18 (2023)
856    41
$u https://pubmed.ncbi.nlm.nih.gov/37607560 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y - $z 0
990    __
$a 20231013 $b ABA008
991    __
$a 20231026110113 $b ABA008
999    __
$a ok $b bmc $g 2000014 $s 1202599
BAS    __
$a 3
BAS    __
$a PreBMC-MEDLINE
BMC    __
$a 2023 $b 68 $c 18 $e 20230914 $i 1361-6560 $m Physics in medicine and biology $n Phys Med Biol $x MED00003821
LZP    __
$a Pubmed-20231013

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...