Thermal neutron detection and track recognition method in reference and out-of-field radiotherapy FLASH electron fields using Timepix3 detectors
C. Oancea, J. Solc, A. Bourgouin, C. Granja, J. Jakubek, J. Pivec, F. Riemer, Z. Vykydal, S. Worm, L. Marek
Jazyk angličtina Země Anglie, Velká Británie
Typ dokumentu časopisecké články, práce podpořená grantem
PubMed
37607560
DOI
10.1088/1361-6560/acf2e1
Knihovny.cz E-zdroje
- MeSH
- algoritmy * MeSH
- elektrony * MeSH
- kalibrace MeSH
- neutrony MeSH
- záření gama MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
Objective.This work presents a method for enhanced detection, imaging, and measurement of the thermal neutron flux.Approach. Measurements were performed in a water tank, while the detector is positioned out-of-field of a 20 MeV ultra-high pulse dose rate electron beam. A semiconductor pixel detector Timepix3 with a silicon sensor partially covered by a6LiF neutron converter was used to measure the flux, spatial, and time characteristics of the neutron field. To provide absolute measurements of thermal neutron flux, the detection efficiency calibration of the detectors was performed in a reference thermal neutron field. Neutron signals are recognized and discriminated against other particles such as gamma rays and x-rays. This is achieved by the resolving power of the pixel detector using machine learning algorithms and high-resolution pattern recognition analysis of the high-energy tracks created by thermal neutron interactions in the converter.Main results. The resulting thermal neutrons equivalent dose was obtained using conversion factor (2.13(10) pSv·cm2) from thermal neutron fluence to thermal neutron equivalent dose obtained by Monte Carlo simulations. The calibrated detectors were used to characterize scattered radiation created by electron beams. The results at 12.0 cm depth in the beam axis inside of the water for a delivered dose per pulse of 1.85 Gy (pulse length of 2.4μs) at the reference depth, showed a contribution of flux of 4.07(8) × 103particles·cm-2·s-1and equivalent dose of 1.73(3) nSv per pulse, which is lower by ∼9 orders of magnitude than the delivered dose.Significance. The presented methodology for in-water measurements and identification of characteristic thermal neutrons tracks serves for the selective quantification of equivalent dose made by thermal neutrons in out-of-field particle therapy.
ADVACAM U Pergamenky 12 170 00 Prague 7 Czech Republic
Czech Metrology Institute Okruzni 31 638 00 Brno Czech Republic
Deutsches Elektronen Synchrotron DESY Platanenallee 6 15738 Zeuthen Germany
Faculty of Mathematics and Physics Charles University Prague Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc23016237
- 003
- CZ-PrNML
- 005
- 20231026110119.0
- 007
- ta
- 008
- 231013s2023 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1088/1361-6560/acf2e1 $2 doi
- 035 __
- $a (PubMed)37607560
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Oancea, Cristina $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic $u University of Bucharest, Bucharest, Romania
- 245 10
- $a Thermal neutron detection and track recognition method in reference and out-of-field radiotherapy FLASH electron fields using Timepix3 detectors / $c C. Oancea, J. Solc, A. Bourgouin, C. Granja, J. Jakubek, J. Pivec, F. Riemer, Z. Vykydal, S. Worm, L. Marek
- 520 9_
- $a Objective.This work presents a method for enhanced detection, imaging, and measurement of the thermal neutron flux.Approach. Measurements were performed in a water tank, while the detector is positioned out-of-field of a 20 MeV ultra-high pulse dose rate electron beam. A semiconductor pixel detector Timepix3 with a silicon sensor partially covered by a6LiF neutron converter was used to measure the flux, spatial, and time characteristics of the neutron field. To provide absolute measurements of thermal neutron flux, the detection efficiency calibration of the detectors was performed in a reference thermal neutron field. Neutron signals are recognized and discriminated against other particles such as gamma rays and x-rays. This is achieved by the resolving power of the pixel detector using machine learning algorithms and high-resolution pattern recognition analysis of the high-energy tracks created by thermal neutron interactions in the converter.Main results. The resulting thermal neutrons equivalent dose was obtained using conversion factor (2.13(10) pSv·cm2) from thermal neutron fluence to thermal neutron equivalent dose obtained by Monte Carlo simulations. The calibrated detectors were used to characterize scattered radiation created by electron beams. The results at 12.0 cm depth in the beam axis inside of the water for a delivered dose per pulse of 1.85 Gy (pulse length of 2.4μs) at the reference depth, showed a contribution of flux of 4.07(8) × 103particles·cm-2·s-1and equivalent dose of 1.73(3) nSv per pulse, which is lower by ∼9 orders of magnitude than the delivered dose.Significance. The presented methodology for in-water measurements and identification of characteristic thermal neutrons tracks serves for the selective quantification of equivalent dose made by thermal neutrons in out-of-field particle therapy.
- 650 12
- $a elektrony $7 D004583
- 650 12
- $a algoritmy $7 D000465
- 650 _2
- $a kalibrace $7 D002138
- 650 _2
- $a záření gama $7 D005720
- 650 _2
- $a neutrony $7 D009502
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Solc, Jaroslav $u Czech Metrology Institute, Okruzni 31, 638 00 Brno, Czech Republic
- 700 1_
- $a Bourgouin, Alexandra $u Dosimetry for Radiation Therapy and Diagnostic Radiology, Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, 38116, Germany $1 https://orcid.org/0000000164294122
- 700 1_
- $a Granja, Carlos $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic $1 https://orcid.org/0000000243981553
- 700 1_
- $a Jakubek, Jan $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic
- 700 1_
- $a Pivec, Jiri $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic
- 700 1_
- $a Riemer, Felix $u Deutsches Elektronen-Synchrotron DESY, Platanenallee 6, 15738 Zeuthen, Germany
- 700 1_
- $a Vykydal, Zdenek $u Czech Metrology Institute, Okruzni 31, 638 00 Brno, Czech Republic
- 700 1_
- $a Worm, Steven $u Deutsches Elektronen-Synchrotron DESY, Platanenallee 6, 15738 Zeuthen, Germany
- 700 1_
- $a Marek, Lukas $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic $u Faculty of Mathematics and Physics, Charles University, Prague, Czech Republic
- 773 0_
- $w MED00003821 $t Physics in medicine and biology $x 1361-6560 $g Roč. 68, č. 18 (2023)
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/37607560 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20231013 $b ABA008
- 991 __
- $a 20231026110113 $b ABA008
- 999 __
- $a ok $b bmc $g 2000014 $s 1202599
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2023 $b 68 $c 18 $e 20230914 $i 1361-6560 $m Physics in medicine and biology $n Phys Med Biol $x MED00003821
- LZP __
- $a Pubmed-20231013