Detail
Článek
Článek online
FT
Medvik - BMČ
  • Je něco špatně v tomto záznamu ?

Clinical outcomes and survival comparison between NexGen all-poly and its metal-backed equivalent in total knee arthroplasty

V. Apostolopoulos, L. Nachtnebl, M. Mahdal, L. Pazourek, P. Boháč, P. Janíček, T. Tomáš

. 2023 ; 47 (9) : 2207-2213. [pub] 20230418

Jazyk angličtina Země Německo

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/bmc23016446

PURPOSE: This study aims to compare total knee replacement (TKA) with NexGen All-Poly (APT) and NexGen Metal-Backed (MBT) in terms of implant survivorship, reasons leading to implant failure and functional results of defined age categories. METHODS: A single-centre, retrospective evaluation of 812 patients who underwent knee replacement with NexGen CR between 2005 and 2021, comparing a modern congruent APT component to a modular MBT equivalent component using a similar surgical technique at a notable mean follow-up duration. Implant survival, functional outcomes using the Knee Society Score and range of motion were evaluated and compared in different age categories. RESULTS: Of the 812 NexGen CR TKAs performed at our institution, 410 (50.4%) used APT components and 402 (49.6%) MBT components. The survival rate of NexGen APT was 97.1% and that of NexGen MBT was 93.2% (p = 0.36). Removal of the implant occurred overall in 15 cases, for MBT in ten cases, and for APT in four cases. The FS was proved to be significantly higher when APT components were implanted in younger patients than for MBT (p = 0.005). A similar range of motion between the components was recorded (p = 0.1926). CONCLUSION: Under defined conditions, we measured the clinical results of implants from a single manufacturer implanted in a single department using a similar surgical technique. Considering the limitations, we suggest that all-polyethylene tibial components are equal or even superior to metal-backed ones across the examined age categories.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc23016446
003      
CZ-PrNML
005      
20250304093927.0
007      
ta
008      
231013s2023 gw f 000 0|eng||
009      
AR
024    7_
$a 10.1007/s00264-023-05772-3 $2 doi
035    __
$a (PubMed)37069277
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a gw
100    1_
$a Apostolopoulos, Vasileios $u First Department of Orthopaedic Surgery, St. Anne's University Hospital and Faculty of Medicine, Masaryk University, FN u Sv Anny Brne, Pekarska 53, Brno, 60200, Czechia $1 https://orcid.org/0000000197194769
245    10
$a Clinical outcomes and survival comparison between NexGen all-poly and its metal-backed equivalent in total knee arthroplasty / $c V. Apostolopoulos, L. Nachtnebl, M. Mahdal, L. Pazourek, P. Boháč, P. Janíček, T. Tomáš
520    9_
$a PURPOSE: This study aims to compare total knee replacement (TKA) with NexGen All-Poly (APT) and NexGen Metal-Backed (MBT) in terms of implant survivorship, reasons leading to implant failure and functional results of defined age categories. METHODS: A single-centre, retrospective evaluation of 812 patients who underwent knee replacement with NexGen CR between 2005 and 2021, comparing a modern congruent APT component to a modular MBT equivalent component using a similar surgical technique at a notable mean follow-up duration. Implant survival, functional outcomes using the Knee Society Score and range of motion were evaluated and compared in different age categories. RESULTS: Of the 812 NexGen CR TKAs performed at our institution, 410 (50.4%) used APT components and 402 (49.6%) MBT components. The survival rate of NexGen APT was 97.1% and that of NexGen MBT was 93.2% (p = 0.36). Removal of the implant occurred overall in 15 cases, for MBT in ten cases, and for APT in four cases. The FS was proved to be significantly higher when APT components were implanted in younger patients than for MBT (p = 0.005). A similar range of motion between the components was recorded (p = 0.1926). CONCLUSION: Under defined conditions, we measured the clinical results of implants from a single manufacturer implanted in a single department using a similar surgical technique. Considering the limitations, we suggest that all-polyethylene tibial components are equal or even superior to metal-backed ones across the examined age categories.
650    _2
$a lidé $7 D006801
650    12
$a totální endoprotéza kolene $x škodlivé účinky $7 D019645
650    12
$a protézy kolene $7 D007720
650    _2
$a retrospektivní studie $7 D012189
650    _2
$a protézy - design $7 D011474
650    _2
$a polyethylen $7 D020959
650    _2
$a kovy $7 D008670
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Nachtnebl, Luboš $u First Department of Orthopaedic Surgery, St. Anne's University Hospital and Faculty of Medicine, Masaryk University, FN u Sv Anny Brne, Pekarska 53, Brno, 60200, Czechia
700    1_
$a Mahdal, Michal $u First Department of Orthopaedic Surgery, St. Anne's University Hospital and Faculty of Medicine, Masaryk University, FN u Sv Anny Brne, Pekarska 53, Brno, 60200, Czechia
700    1_
$a Pazourek, Lukáš $u First Department of Orthopaedic Surgery, St. Anne's University Hospital and Faculty of Medicine, Masaryk University, FN u Sv Anny Brne, Pekarska 53, Brno, 60200, Czechia
700    1_
$a Boháč, Petr $u Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, University of Technology, Brno, Czechia
700    1_
$a Janíček, Pavel, $d 1952-2025 $u First Department of Orthopaedic Surgery, St. Anne's University Hospital and Faculty of Medicine, Masaryk University, FN u Sv Anny Brne, Pekarska 53, Brno, 60200, Czechia $7 mzk2004148320
700    1_
$a Tomáš, Tomáš $u First Department of Orthopaedic Surgery, St. Anne's University Hospital and Faculty of Medicine, Masaryk University, FN u Sv Anny Brne, Pekarska 53, Brno, 60200, Czechia. tomas.tomas@fnusa.cz $1 https://orcid.org/0000000255012143
773    0_
$w MED00002384 $t International orthopaedics $x 1432-5195 $g Roč. 47, č. 9 (2023), s. 2207-2213
856    41
$u https://pubmed.ncbi.nlm.nih.gov/37069277 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y - $z 0
990    __
$a 20231013 $b ABA008
991    __
$a 20250304093939 $b ABA008
999    __
$a ok $b bmc $g 2000142 $s 1202808
BAS    __
$a 3
BAS    __
$a PreBMC-MEDLINE
BMC    __
$a 2023 $b 47 $c 9 $d 2207-2213 $e 20230418 $i 1432-5195 $m International orthopaedics $n Int Orthop $x MED00002384