Something wrong with this record ?
Enhancing Accuracy in Breast Density Assessment Using Deep Learning: A Multicentric, Multi-Reader Study
M. Biroš, D. Kvak, J. Dandár, R. Hrubý, E. Janů, A. Atakhanova, MA. Al-Antari
Status not-indexed Language English Country Switzerland
Document type Journal Article
Grant support
X001
Carebot, Ltd.
NLK
Directory of Open Access Journals
from 2011
PubMed Central
from 2011
Europe PubMed Central
from 2011
ProQuest Central
from 2011-01-01
Open Access Digital Library
from 2011-01-01
Open Access Digital Library
from 2011-01-01
ROAD: Directory of Open Access Scholarly Resources
from 2011
- Publication type
- Journal Article MeSH
The evaluation of mammographic breast density, a critical indicator of breast cancer risk, is traditionally performed by radiologists via visual inspection of mammography images, utilizing the Breast Imaging-Reporting and Data System (BI-RADS) breast density categories. However, this method is subject to substantial interobserver variability, leading to inconsistencies and potential inaccuracies in density assessment and subsequent risk estimations. To address this, we present a deep learning-based automatic detection algorithm (DLAD) designed for the automated evaluation of breast density. Our multicentric, multi-reader study leverages a diverse dataset of 122 full-field digital mammography studies (488 images in CC and MLO projections) sourced from three institutions. We invited two experienced radiologists to conduct a retrospective analysis, establishing a ground truth for 72 mammography studies (BI-RADS class A: 18, BI-RADS class B: 43, BI-RADS class C: 7, BI-RADS class D: 4). The efficacy of the DLAD was then compared to the performance of five independent radiologists with varying levels of experience. The DLAD showed robust performance, achieving an accuracy of 0.819 (95% CI: 0.736-0.903), along with an F1 score of 0.798 (0.594-0.905), precision of 0.806 (0.596-0.896), recall of 0.830 (0.650-0.946), and a Cohen's Kappa (κ) of 0.708 (0.562-0.841). The algorithm achieved robust performance that matches and in four cases exceeds that of individual radiologists. The statistical analysis did not reveal a significant difference in accuracy between DLAD and the radiologists, underscoring the model's competitive diagnostic alignment with professional radiologist assessments. These results demonstrate that the deep learning-based automatic detection algorithm can enhance the accuracy and consistency of breast density assessments, offering a reliable tool for improving breast cancer screening outcomes.
Carebot Ltd 128 00 Prague Czech Republic
Department of Radiology Masaryk Memorial Cancer Institute 602 00 Brno Czech Republic
Department of Simulation Medicine Faculty of Medicine Masaryk University 625 00 Brno Czech Republic
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc24012690
- 003
- CZ-PrNML
- 005
- 20240726151343.0
- 007
- ta
- 008
- 240723s2024 sz f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.3390/diagnostics14111117 $2 doi
- 035 __
- $a (PubMed)38893643
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a sz
- 100 1_
- $a Biroš, Marek $u Carebot, Ltd., 128 00 Prague, Czech Republic $1 https://orcid.org/000000032781623X
- 245 10
- $a Enhancing Accuracy in Breast Density Assessment Using Deep Learning: A Multicentric, Multi-Reader Study / $c M. Biroš, D. Kvak, J. Dandár, R. Hrubý, E. Janů, A. Atakhanova, MA. Al-Antari
- 520 9_
- $a The evaluation of mammographic breast density, a critical indicator of breast cancer risk, is traditionally performed by radiologists via visual inspection of mammography images, utilizing the Breast Imaging-Reporting and Data System (BI-RADS) breast density categories. However, this method is subject to substantial interobserver variability, leading to inconsistencies and potential inaccuracies in density assessment and subsequent risk estimations. To address this, we present a deep learning-based automatic detection algorithm (DLAD) designed for the automated evaluation of breast density. Our multicentric, multi-reader study leverages a diverse dataset of 122 full-field digital mammography studies (488 images in CC and MLO projections) sourced from three institutions. We invited two experienced radiologists to conduct a retrospective analysis, establishing a ground truth for 72 mammography studies (BI-RADS class A: 18, BI-RADS class B: 43, BI-RADS class C: 7, BI-RADS class D: 4). The efficacy of the DLAD was then compared to the performance of five independent radiologists with varying levels of experience. The DLAD showed robust performance, achieving an accuracy of 0.819 (95% CI: 0.736-0.903), along with an F1 score of 0.798 (0.594-0.905), precision of 0.806 (0.596-0.896), recall of 0.830 (0.650-0.946), and a Cohen's Kappa (κ) of 0.708 (0.562-0.841). The algorithm achieved robust performance that matches and in four cases exceeds that of individual radiologists. The statistical analysis did not reveal a significant difference in accuracy between DLAD and the radiologists, underscoring the model's competitive diagnostic alignment with professional radiologist assessments. These results demonstrate that the deep learning-based automatic detection algorithm can enhance the accuracy and consistency of breast density assessments, offering a reliable tool for improving breast cancer screening outcomes.
- 590 __
- $a NEINDEXOVÁNO
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Kvak, Daniel $u Carebot, Ltd., 128 00 Prague, Czech Republic $u Department of Simulation Medicine, Faculty of Medicine, Masaryk University, 625 00 Brno, Czech Republic $1 https://orcid.org/0000000178087773
- 700 1_
- $a Dandár, Jakub $u Carebot, Ltd., 128 00 Prague, Czech Republic
- 700 1_
- $a Hrubý, Robert $u Carebot, Ltd., 128 00 Prague, Czech Republic $1 https://orcid.org/0000000299858060
- 700 1_
- $a Janů, Eva $u Department of Radiology, Masaryk Memorial Cancer Institute, 602 00 Brno, Czech Republic $1 https://orcid.org/000000020147211X
- 700 1_
- $a Atakhanova, Anora $u Carebot, Ltd., 128 00 Prague, Czech Republic
- 700 1_
- $a Al-Antari, Mugahed A $u Department of Artificial Intelligence and Data Science, Daeyang AI Center, Sejong University, Seoul 05006, Republic of Korea $1 https://orcid.org/0000000244574407
- 773 0_
- $w MED00195450 $t Diagnostics $x 2075-4418 $g Roč. 14, č. 11 (2024)
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/38893643 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20240723 $b ABA008
- 991 __
- $a 20240726151335 $b ABA008
- 999 __
- $a ok $b bmc $g 2125447 $s 1224553
- BAS __
- $a 3
- BAS __
- $a PreBMC-PubMed-not-MEDLINE
- BMC __
- $a 2024 $b 14 $c 11 $e 20240528 $i 2075-4418 $m Diagnostics $n Diagnostics (Basel) $x MED00195450
- GRA __
- $a X001 $p Carebot, Ltd.
- LZP __
- $a Pubmed-20240723