Dosimetric and temporal beam characterization of individual pulses in FLASH radiotherapy using Timepix3 pixelated detector placed out-of-field
C. Oancea, K. Sykorova, J. Jakubek, J. Pivec, F. Riemer, S. Worm, A. Bourgouin
Jazyk angličtina Země Itálie
Typ dokumentu časopisecké články
- MeSH
- časové faktory MeSH
- částice - urychlovače * MeSH
- celková dávka radioterapie * MeSH
- fantomy radiodiagnostické MeSH
- radiometrie * přístrojové vybavení metody MeSH
- radioterapie přístrojové vybavení MeSH
- Publikační typ
- časopisecké články MeSH
BACKGROUND: FLASH radiotherapy necessitates the development of advanced Quality Assurance methods and detectors for accurate monitoring of the radiation field. This study introduces enhanced time-resolution detection systems and methods used to measure the delivered number of pulses, investigate temporal structure of individual pulses and dose-per-pulse (DPP) based on secondary radiation particles produced in the experimental room. METHODS: A 20 MeV electron beam generated from a linear accelerator (LINAC) was delivered to a water phantom. Ultra-high dose-per-pulse electron beams were used with a dose-per-pulse ranging from ̴ 1 Gy to over 7 Gy. The pulse lengths ranged from 1.18 μs to 2.88 μs at a pulse rate frequency of 5 Hz. A semiconductor pixel detector Timepix3 was used to track single secondary particles. Measurements were performed in the air, while the detector was positioned out-of-field at a lateral distance of 200 cm parallel with the LINAC exit window. The dose deposited was measured along with the pulse length and the nanostructure of the pulse. RESULTS: The time of arrival (ToA) of single particles was measured with a resolution of 1.56 ns, while the deposited energy was measured with a resolution of several keV based on the Time over Threshold (ToT) value. The pulse count measured by the Timepix3 detector corresponded with the delivered values, which were measured using an in-flange integrating current transformer (ICT). A linear response (R2 = 0.999) was established between the delivered beam current and the measured dose at the detector position (orders of nGy). The difference between the average measured and delivered pulse length was ∼0.003(30) μs. CONCLUSION: This simple non-invasive method exhibits no limitations on the delivered DPP within the range used during this investigation.
ADVACAM U Pergamenky 12 170 00 Prague 7 Czech Republic
Deutsches Elektronen Synchrotron DESY Platanenallee 6 15738 Zeuthen Germany
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc25010442
- 003
- CZ-PrNML
- 005
- 20250429135144.0
- 007
- ta
- 008
- 250415e20241211it f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.ejmp.2024.104872 $2 doi
- 035 __
- $a (PubMed)39667142
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a it
- 100 1_
- $a Oancea, Cristina $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic. Electronic address: cristina.oancea@advacam.cz
- 245 10
- $a Dosimetric and temporal beam characterization of individual pulses in FLASH radiotherapy using Timepix3 pixelated detector placed out-of-field / $c C. Oancea, K. Sykorova, J. Jakubek, J. Pivec, F. Riemer, S. Worm, A. Bourgouin
- 520 9_
- $a BACKGROUND: FLASH radiotherapy necessitates the development of advanced Quality Assurance methods and detectors for accurate monitoring of the radiation field. This study introduces enhanced time-resolution detection systems and methods used to measure the delivered number of pulses, investigate temporal structure of individual pulses and dose-per-pulse (DPP) based on secondary radiation particles produced in the experimental room. METHODS: A 20 MeV electron beam generated from a linear accelerator (LINAC) was delivered to a water phantom. Ultra-high dose-per-pulse electron beams were used with a dose-per-pulse ranging from ̴ 1 Gy to over 7 Gy. The pulse lengths ranged from 1.18 μs to 2.88 μs at a pulse rate frequency of 5 Hz. A semiconductor pixel detector Timepix3 was used to track single secondary particles. Measurements were performed in the air, while the detector was positioned out-of-field at a lateral distance of 200 cm parallel with the LINAC exit window. The dose deposited was measured along with the pulse length and the nanostructure of the pulse. RESULTS: The time of arrival (ToA) of single particles was measured with a resolution of 1.56 ns, while the deposited energy was measured with a resolution of several keV based on the Time over Threshold (ToT) value. The pulse count measured by the Timepix3 detector corresponded with the delivered values, which were measured using an in-flange integrating current transformer (ICT). A linear response (R2 = 0.999) was established between the delivered beam current and the measured dose at the detector position (orders of nGy). The difference between the average measured and delivered pulse length was ∼0.003(30) μs. CONCLUSION: This simple non-invasive method exhibits no limitations on the delivered DPP within the range used during this investigation.
- 650 12
- $a radiometrie $x přístrojové vybavení $x metody $7 D011874
- 650 12
- $a částice - urychlovače $7 D010315
- 650 12
- $a celková dávka radioterapie $7 D011879
- 650 _2
- $a časové faktory $7 D013997
- 650 _2
- $a fantomy radiodiagnostické $7 D019047
- 650 _2
- $a radioterapie $x přístrojové vybavení $7 D011878
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Sykorova, Katerina $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic
- 700 1_
- $a Jakubek, Jan $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic
- 700 1_
- $a Pivec, Jiri $u ADVACAM, U Pergamenky 12, 170 00 Prague 7, Czech Republic
- 700 1_
- $a Riemer, Felix $u Deutsches Elektronen-Synchrotron DESY, Platanenallee 6, 15738 Zeuthen, Germany
- 700 1_
- $a Worm, Steven $u Deutsches Elektronen-Synchrotron DESY, Platanenallee 6, 15738 Zeuthen, Germany
- 700 1_
- $a Bourgouin, Alexandra $u Dosimetry for Radiation Therapy and Diagnostic Radiology, Physikalisch-Technische Bundesanstalt (PTB), Braunschweig 38116, Germany; Metrology Research Center, National Research Council of Canada, 1200 Montreal Road, Ottawa, ON, K1A0R6, Canada
- 773 0_
- $w MED00167391 $t Physica medica $x 1724-191X $g Roč. 129 (20241211), s. 104872
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/39667142 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20250415 $b ABA008
- 991 __
- $a 20250429135139 $b ABA008
- 999 __
- $a ok $b bmc $g 2311663 $s 1247523
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2025 $b 129 $c - $d 104872 $e 20241211 $i 1724-191X $m Physica medica $n Phys Med $x MED00167391
- LZP __
- $a Pubmed-20250415