Jak urcovat vizus?
[How should visual acuity be measured?]
Language Czech Country Czech Republic Media print
Document type English Abstract, Journal Article
PubMed
8768473
- MeSH
- Adult MeSH
- Middle Aged MeSH
- Humans MeSH
- Adolescent MeSH
- Computers MeSH
- Aged MeSH
- Visual Acuity * MeSH
- Vision Tests methods MeSH
- Check Tag
- Adult MeSH
- Middle Aged MeSH
- Humans MeSH
- Adolescent MeSH
- Male MeSH
- Aged MeSH
- Female MeSH
- Publication type
- English Abstract MeSH
- Journal Article MeSH
1) The ordinary Snellen chart with letter optotypes is not suited for research purposes. Unequal legibility of letters, unequal number of optotypes on rows of the chart and unequal progression of letter sizes limit the usage of such charts for screening purposes only. 2) The results of our computerised method for determining of visual acuity correlate very well with the results gained using the standardised Landolt optotype chart (r = 0.931). Our method brings whole line acuity, threshold visual acuity with its variability and a new parameter of line slope (decrease of correct readings from 95% to 5%), expressed in standardised rows (log MAR = 0.1). The results are printed on a simple printer and stored in a database.