In vivo thermoluminescence dosimetry for total body irradiation
Language English Country Great Britain, England Media print
Document type Journal Article
- MeSH
- Whole-Body Irradiation * MeSH
- Photons MeSH
- Calibration MeSH
- Humans MeSH
- Thermoluminescent Dosimetry methods MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
An improvement in the clinical results obtained using total body irradiation (TBI) with photon beams requires precise TBI treatment planning, reproducible irradiation, precise in vivo dosimetry, accurate documentation and careful evaluation. In vivo dosimetry using LiF Harshaw TLD-100 chips was used during the TBI treatments performed in our department. The results of in vivo thermoluminescence dosimetry (TLD) show that using TLD measurements and interactive adjustment of some treatment parameters based on these measurements, like monitor unit calculations, lung shielding thickness and patient positioning, it is possible to achieve high precision in absorbed dose delivery (less than 0.5%) as well as in homogeneity of irradiation (less than 6%).